Class 1 Device Recall Faaborg Person Lift, PL Series

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall Faaborg Person Lift, PL Seriessee related information
Date Initiated by FirmJanuary 21, 2004
Date PostedJuly 20, 2004
Recall Status1 Terminated 3 on October 22, 2007
Recall NumberZ-0551-04
Recall Event ID 28199
Product Classification Lift, Patient, Non-Ac-Powered - Product Code FSA
ProductFaaborg patient lift/hoist, battery operated lifts. The PL series lifts have electric parallel base widening, with a weight capacity of 365 lbs. to 550 lbs. depending on the base model. Please note: the Faaborg Person Lift arm assembly can be interchanged on the various lift series bases. The arm assembly is labeled ''Faaborgliften Made in Denmark'', while the base plate reads in part ''Faaborg Rehab Technic Lift Model: VL-***** Serial No: **** Capacity: *** KG. Manufactured: **-**** 5600 Faaborg Phone+4570205553 Fax+4570205554'' A sticker is also present on the base reading ''Moving Solutions 800-228-7980 Fax 630-963-9588''. The Users Guide refers to the lift as a ''Faaborghoist''.
Code Information all PL series models, all serial numbers
FEI Number 1000305914
Recalling Firm/
Manufacturer
Moving Solutions, Inc.
2550 Wisconsin Ave
Downers Grove IL 60515
For Additional Information ContactMr. Andrew M. Ringo
800-228-7980
Manufacturer Reason
for Recall
Excessive wear of the hanger bar bolt, that connects the lift arm to the sling spreader bar, may cause the bolt to snap allowing the patient to fall.
FDA Determined
Cause 2
Other
ActionLetters were sent to all Faaborg lift customers on 11/29/01 and 01/24/04 advising the users that all bolts in connection with the hanger bar must be inspected, greased and tightened, and all defective or worn down bolts and washers must be changed. The 1/24/04 letters included an upgraded maintenance schedule and an Important Service/Safety Reminder, as well as a supply of hanger bar bolt washers. Any questions were directed to Moving Solutions' Service Department at 800-228-7980. Follow-up letters were sent to the accounts on 3/11/04, referencing the 1/21/04 letter and instructing the accounts not to use the lifts until the bolts have been inspected and retrofitted with the washers. Copies of the revised maintenance schedule and operating instructions were included.
Quantity in Commerce604 lifts
DistributionNationwide to nursing homes, residential schools and home patients, and internationally to Canada.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
-
-