| | Class 2 Device Recall CryoValve |  |
| Date Initiated by Firm | June 17, 2004 |
| Date Posted | July 20, 2004 |
| Recall Status1 |
Terminated 3 on March 07, 2005 |
| Recall Number | Z-1057-04 |
| Recall Event ID |
29451 |
| Product Classification |
Heart-Valve, Allograft - Product Code MIE
|
| Product | CryoValve Pulmonary Valve and Conduit |
| Code Information |
Donor Number: 64208, Serial Number: 7887057, Model Number: SGPV00, Donor Number: 64222, Serial Number: 7894333, Model Number: SGPV00 and Donor Number: 64224, Serial Number: 7894298, Model Number: SGPV00 |
| FEI Number |
3001451326
|
Recalling Firm/ Manufacturer |
Cryolife Inc
1655 Roberts Blvd Nw
Kennesaw
GA
30144-3632
|
| For Additional Information Contact | Mr. P. Tyler Cochran, CQA, CTBS 770-419-3355 |
Manufacturer Reason for Recall | The Amphotericin B Intermediate (G10030), one of the components of the antimicrobial solution, had reached expiration at the time it was utilized to manufacture the Combo Antibiotic Stock Solution (G10070). |
FDA Determined Cause 2 | Other |
| Action | Consignees were notified by letter on 06/17/2004 and 06/18/2004. |
| Quantity in Commerce | 3 unit |
| Distribution | CA, OH, OR, PA. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
|
|
|