Date Initiated by Firm |
September 21, 2004 |
Date Posted |
November 02, 2004 |
Recall Status1 |
Terminated 3 on February 07, 2005 |
Recall Number |
Z-0062-05 |
Recall Event ID |
30168 |
510(K)Number |
K030739 K992044
|
Product Classification |
Pump, Infusion - Product Code FRN
|
Product |
Infusion Dynamics Power Infuser pump. Model number M100B-3A |
Code Information |
Serial numbers 13-05863 through 13-05929 and 13-6005 through 13-06036 |
Recalling Firm/ Manufacturer |
Infusion Dynamics, A Div. of Zoll Med Corpinfusion Dynamics 5209 Militia Hill Rd Plymouth Meeting PA 19462-1216
|
For Additional Information Contact |
Andrew Fleischacker 610-941-0136 Ext. 114
|
Manufacturer Reason for Recall |
Front Panel may be cracked on some units. If cracked fluid can enter the infuser and short the start/stop button. Shorting would cause the infuser to stay in the open position.
|
FDA Determined Cause 2 |
Other |
Action |
The recalling firm issued a recall letter dated 9/20/04 to their customers informing them of the problem and the need to return the units. |
Quantity in Commerce |
255 units |
Distribution |
The products were shipped to dealers in TX, MO, and VT; medical facilities in CA, NV, and NY. The product was also shipped to government accounts in PA and Germany. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = FRN and Original Applicant = INFUSION DYNAMICS, INC.
|