Class 2 Device Recall FlexTrayTM Endopath Bariatric Tray |
|
Date Initiated by Firm |
December 15, 2004 |
Date Posted |
April 27, 2005 |
Recall Status1 |
Terminated 3 on May 30, 2012 |
Recall Number |
Z-0735-05 |
Recall Event ID |
30727 |
Product Classification |
Instrument, Manual, Surgical, General Use - Product Code MDM
|
Product |
FlexTrayTM Endopath¿ Bariatric Tray, Kit code KBW71S. |
Code Information |
V42A83, V42R81, V42Z57, V4330Z |
Recalling Firm/ Manufacturer |
Ethicon Endo-Surgery 4545 Creek Rd Cincinnati OH 45242-2803
|
For Additional Information Contact |
Carol Montandon 513-337-3419
|
Manufacturer Reason for Recall |
Due to tolerance issues with the device, pins can work free from their respective holes. The result during use could be separation of the handles that could lead to compromised ratchet button functionality. Loss of ratchet button functionality may lead to the end effector being locked in the open or closed position.
|
FDA Determined Cause 2 |
Other |
Action |
The recalling firm sent letters to hospitals and distributors, dated 12/14/04. |
Quantity in Commerce |
150 devices. |
Distribution |
The devices were distributed to hospitals and distributors located throughout the United States and to foreign consignees located in Saudi Arabia, Austria, Egypt, United Kingdom, Germany, France, Lebanon, The Netherlands, Spain, Pakistan, South Africa, Italy, United Arab Emirate, Czech Republic, Sweden, Switzerland, Greece, India, Portugal, Slovenia, Hungary, Palestine, and Kuwait. |