| Class 2 Device Recall Medline Custom Sterile Surgical Packs | |
Date Initiated by Firm | March 04, 2005 |
Date Posted | March 11, 2005 |
Recall Status1 |
Terminated 3 on May 25, 2006 |
Recall Number | Z-0554-05 |
Recall Event ID |
31315 |
Product Classification |
General Surgery Tray - Product Code LRO
|
Product | Medline Custom Sterile Surgical Packs which contain a Tyvek Breather Pouch, part number 32901, lot numbers 44134-01 and 43589-01; the breather pouch measures 25.25 inches x 30.75 inches, 5 mil thick; Medline Industries, Inc., Mundelein, IL 60060; |
Code Information |
all lots distributed between 6/15/04 and 3/1/05 of the following reorder numbers: DYNJ02170A, DYNJ02176A, DYNJ02179B, DYNJ02182A, DYNJ04314A, DYNJ04316, DYNJ04321, DYNJ04325, DYNJ04329, DYNJ04336B, DYNJ04338B, DYNJ04371, DYNJ04383, DYNJ04388, DYNJ04407A, DYNJ04408A, DYNJ04419A, DYNJ04427, DYNJ04455, DYNJ04468B, DYNJ04472A, DYNJ04476, DYNJ04478A, DYNJ04481B, DYNJ04482, DYNJ04483, DYNJ04485, DYNJ04494, DYNJ04497, DYNJ04502, DYNJ04504, DYNJ04505, DYNJ04509, DYNJ04542, DYNJ04545, DYNJ04548A, DYNJ04556A, DYNJ04557B, DYNJ04558A, DYNJ04572, DYNJ04579A, DYNJ04597, DYNJ04609, DYNJ04611A, DYNJ04620A, DYNJ04629B, DYNJ04676A, DYNJ04677, DYNJ04683, DYNJ04691, DYNJ04709A, DYNJ04710, DYNJ04711, DYNJ04711A, DYNJ04712, DYNJ04713A, DYNJ04720A, DYNJ04723, DYNJ04729, DYNJ04744, DYNJ04745A, DYNJ04754A, DYNJ04805A, DYNJ04807A, DYNJ04811, DYNJ04822A, DYNJ04825A, DYNJ04832, DYNJ04834, DYNJ04840, DYNJ04841, DYNJ04878, DYNJ04880, DYNJ04888, DYNJ05787K, DYNJ05798K, DYNJ06075G, DYNJ06324L, DYNJ06682B, DYNJ06739B, DYNJ06743A, DYNJ06744B, DYNJ06782A, DYNJ06796J, DYNJ06839A, DYNJ06873A, DYNJ06876D, DYNJ06876F, DYNJ06908B, DYNJ06927A, DYNJ06930F, DYNJ06952F, DYNJ06976, DYNJ07536, DYNJ07537C, DYNJ07626A, DYNJ07626A, DYNJ07626A, DYNJ07670C, DYNJ07833F, DYNJ07881C, DYNJ07910A, DYNJ07942, DYNJ07942, DYNJ07951A, DYNJ07954A, DYNJ07954A, DYNJ07982A, DYNJ07995, DYNJ08351K, DYNJ08387F, DYNJ08601A, DYNJ08671I, DYNJ08713C, DYNJ08796D, DYNJ08802, DYNJ08814F, DYNJ08815D, DYNJ08816C, DYNJ08840J, DYNJ08844I, DYNJ09066C, DYNJ09156F, DYNJ09174D, DYNJ09188A, DYNJ09194F, DYNJ09258R, DYNJ09259I, DYNJ09261J, DYNJ09294B, DYNJ09322D, DYNJ09343F, DYNJ09346, DYNJ09397A, DYNJ09398, DYNJ09449C, DYNJ09515C, DYNJ09528D, DYNJ09531D, DYNJ09532F, DYNJ09533F, DYNJ09590D, DYNJ09617B, DYNJ09619D, DYNJ09624A, DYNJ09677A, DYNJ09677C, DYNJ09681C, DYNJ09682B, DYNJ09684B, DYNJ09694A, DYNJ09703B, DYNJ09787B, DYNJ09807K, DYNJ09857B, DYNJ09858A, DYNJ09874C, DYNJ09875G, DYNJ09882A, DYNJ09921, DYNJ09923, DYNJ09932B, DYNJ09933, DYNJ09941C, DYNJ09981A, DYNJ14000A, DYNJ14002, DYNJ14005A, DYNJ14006, DYNJ14008A, DYNJ14014, DYNJ14017A, DYNJ14018A, DYNJ14024, DYNJ14038, DYNJ14039, DYNJ14069A, DYNJ14078A, DYNJ14079, DYNJ14095A, DYNJ14096B, DYNJ14097, DYNJ14107, DYNJ14107, DYNJ14110, DYNJ14112A, DYNJ14114, DYNJ14119, DYNJ14127, DYNJ14130, DYNJ14134, DYNJ14137, DYNJ14147A, DYNJ14148A, DYNJ14149A, DYNJ14150A, DYNJ14156B, DYNJ14157B, DYNJ14170, DYNJ14187, DYNJ14218A, DYNJ14225A, DYNJ15062, DYNJ15332A, DYNJ15392D, DYNJ15393G, DYNJ15428F, DYNJ15455P, DYNJ15494A, DYNJ15504D, DYNJ15519A, DYNJ15615D, DYNJ15643L, DYNJ15660C, DYNJ15660D, DYNJ15668D, DYNJ15703A, DYNJ15751C, DYNJ15752, DYNJ15814I, DYNJ15903C, DYNJ15956B, DYNJ15980A, DYNJ16018F, DYNJ16025B, DYNJ16026B, DYNJ16065A, DYNJ16072D, DYNJ16076I, DYNJ16079B, DYNJ16082B, DYNJ16114F, DYNJ16127C, DYNJ16131B, DYNJ16206A, DYNJ16250C, DYNJ16263C, DYNJ16274B, DYNJ16280B, DYNJ16284, DYNJ16289B, DYNJ16327C, DYNJ16327D, DYNJ16342, DYNJ16346A, DYNJ16350C, DYNJ16376C, DYNJ16379B, DYNJ16381C, DYNJ16384, DYNJ16435, DYNJ16461C, DYNJ16510D, DYNJ16514B, DYNJ16518A, DYNJ16535A, DYNJ16536A, DYNJ16541B, DYNJ16567C, DYNJ16575, DYNJ16577, DYNJ16589, DYNJ16591, DYNJ16604, DYNJ16604A, DYNJ16623, DYNJ16625, DYNJ16644, DYNJ17003A, DYNJ17008, DYNJ17016A, DYNJ17018, DYNJ17020, DYNJ17042, DYNJ17044, DYNJ17045A, DYNJ17053, DYNJ17065, DYNJ17068, DYNJ17082, DYNJ19022D, DYNJ19023C, DYNJ19029A, DYNJ19051B, DYNJ19061D, DYNJ19069B, DYNJ19073B, DYNJ19081D, DYNJ19109A, DYNJ19121, DYNJ19138C, DYNJ19139A, DYNJ19146C, DYNJ19146D, DYNJ19154B, DYNJ19170B, DYNJ19199A, DYNJ19230D, DYNJ19233B, DYNJ19234B, DYNJ19235A, DYNJ19249B, DYNJ19251A, DYNJ19256B, DYNJ19257A, DYNJ19296C, DYNJ19329C, DYNJ19371, DYNJ19390B, DYNJ19395B, DYNJ19416B, DYNJ19418D, DYNJ19419A, DYNJ19419B, DYNJ19441D, DYNJ19451I, DYNJ19461B, DYNJ19468A, DYNJ19469, DYNJ19479A, DYNJ19483, DYNJ19500A, DYNJ19501B, DYNJ19509C, DYNJ19515, DYNJ19527D, DYNJ19529A, DYNJ19550A, DYNJ19551B, DYNJ19564A, DYNJ19589A, DYNJ19590A, DYNJ19603B, DYNJ19611B, DYNJ19612B, DYNJ19620A, DYNJ19637G, DYNJ19638F, DYNJ19646D, DYNJ19655A, DYNJ19657, DYNJ19674B, DYNJ19675D, DYNJ19679D, DYNJ19684A, DYNJ19686, DYNJ19694A, DYNJ19696A, DYNJ19697A, DYNJ19699A, DYNJ19709C, DYNJ19710B, DYNJ19724A, DYNJ19726A, DYNJ19729A, DYNJ19730B, DYNJ19738, DYNJ19771B, DYNJ19775A, DYNJ19777A, DYNJ19777A, DYNJ19787B, DYNJ19788A, DYNJ19789A, DYNJ19790B, DYNJ19791D, DYNJ19804B, DYNJ19813, DYNJ19814A, DYNJ19816A, DYNJ19830, DYNJ19833, DYNJ19842A, DYNJ19850B, DYNJ19853, DYNJ19880B, DYNJ19882B, DYNJ19885B, DYNJ19887B, DYNJ19888, DYNJ19900C, DYNJ19901D, DYNJ19907, DYNJ19910, DYNJ19918B, DYNJ19919B, DYNJ19923, DYNJ19929A, DYNJ19929B, DYNJ19933G, DYNJ19935F, DYNJ19938, DYNJ19944, DYNJ19949A, DYNJ19951B, DYNJ19953A, DYNJ19955A, DYNJ19960B, DYNJ19964B, DYNJ19981A, DYNJ19997F, DYNJ20004F, DYNJ20007C, DYNJ20010A, DYNJ20012A, DYNJ20013A, DYNJ20020A, DYNJ20021A, DYNJ20026A, DYNJ20027A, DYNJ20028A, DYNJ20029, DYNJ20030B, DYNJ20031, DYNJ20034, DYNJ20040A, DYNJ20053B, DYNJ20055A, DYNJ20079A, DYNJ20081A, DYNJ20084, DYNJ20104, DYNJ20106A, DYNJ20107, DYNJ20114C, DYNJ20115, DYNJ20117, DYNJ20124A, DYNJ20125, DYNJ20126A, DYNJ20127A, DYNJ20134B, DYNJ20135C, DYNJ20137A, DYNJ20138B, DYNJ20142A, DYNJ20151B, DYNJ20169, DYNJ20170, DYNJ20181, DYNJ20185, DYNJ20186, DYNJ20187A, DYNJ20198A, DYNJ20200, DYNJ20202A, DYNJ20208, DYNJ20211A, DYNJ20213A, DYNJ20214A, DYNJ20215A, DYNJ20216, DYNJ20218A, DYNJ20221A, DYNJ20230, DYNJ20232, DYNJ20248B, DYNJ20252, DYNJ20335A, DYNJ20354A, DYNJ20355B, DYNJ20397A, DYNJ20405, DYNJ20409, DYNJ20422, DYNJ20425B, DYNJ20433, DYNJ20440A, DYNJ20441, DYNJ20442, DYNJ20462B, DYNJ20463, DYNJ20465B, DYNJ20466A, DYNJ20468B, DYNJ20472B, DYNJ20485B, DYNJ20487B, DYNJ20488B, DYNJ20489, DYNJ20489B, DYNJ20490A, DYNJ20491B, DYNJ20495B, DYNJ20496A, DYNJ20498A, DYNJ20498B, DYNJ20502B, DYNJ20503A, DYNJ20507A, DYNJ20508A, DYNJ20509B, DYNJ20518C, DYNJ20521A, DYNJ20522, DYNJ20522A, DYNJ20540B, DYNJ20567B, DYNJ20567B, DYNJ20609A, DYNJ20609B, DYNJ20639A, DYNJ20643A, DYNJ20696A, DYNJ20738, DYNJ20767B, DYNJ20775B, DYNJ20788B, DYNJ20789B, DYNJ20792, DYNJ20792, DYNJ20794, DYNJ20797A, DYNJ20806A, DYNJ20810, DYNJ20829A, DYNJ20834, DYNJ20838A, DYNJ20844A, DYNJ20846A, DYNJ20870, DYNJ20872A, DYNJ20888A, DYNJ20890A, DYNJ20891A, DYNJ20899, DYNJ20910, DYNJ20911A, DYNJ20913, DYNJ20928B, DYNJ20932B, DYNJ20933A, DYNJ20942A, DYNJ20955, DYNJ20994, DYNJ21003, DYNJ21011B, DYNJ21016, DYNJ21018, DYNJ21019A, DYNJ21020B, DYNJ21030, DYNJ21049A, DYNJ21063A, DYNJ21073B, DYNJ21088C, DYNJ21096, DYNJ21097, DYNJ21100A, DYNJ21123, DYNJ21140, DYNJ21149, DYNJ21149A, DYNJ21175A, DYNJ21183, DYNJ21187A, DYNJ21190, DYNJ21191, DYNJ21193, DYNJ21226, DYNJ21249, DYNJ21265, DYNJ21294A, DYNJ21296A, DYNJ21304, DYNJ21307A, DYNJ21307B, DYNJ21320, DYNJ21322, DYNJ21324B, DYNJ21327A, DYNJ21328B, DYNJ21329A, DYNJ21378, DYNJ21399A, DYNJ21400, DYNJ21404B, DYNJ21405A, DYNJ21406A, DYNJ21425, DYNJ21451A, DYNJ21452, DYNJ21475, DYNJ21476, DYNJ21484, DYNJ21485, DYNJ21491, DYNJ21494, DYNJ21497, DYNJ21497A, DYNJ21511, DYNJ21513, DYNJ21526, DYNJ21527, DYNJ21529, DYNJ21530, DYNJ21540, DYNJ21545, DYNJ21557, DYNJ21562, DYNJ21567, DYNJ21569, DYNJ21584, DYNJ21601, DYNJ21633, DYNJ21635, DYNJ21635A, DYNJ21677, DYNJ21718, DYNJ21721, DYNJ21722, DYNJ21726, DYNJ21728, DYNJ21736, DYNJ21737, DYNJ21739, DYNJ21777A, DYNJ21785, DYNJ21792A, DYNJ21797, DYNJ21798, DYNJ21803, DYNJ21804, DYNJ21805, DYNJ21817, DYNJ21818, DYNJ21826, DYNJ21827, DYNJ21828, DYNJ21839, DYNJ21886, DYNJ21965, DYNJ25096D, DYNJ25104D, DYNJ25167B, DYNJ25222B, DYNJ25229C, DYNJ25235C, DYNJ25240A, DYNJ25243A, DYNJ25270C, DYNJ25421A, DYNJ25458, DYNJ25461A, DYNJ25468C, DYNJ25538A, DYNJ25605B, DYNJ25624B, DYNJ25645B, DYNJ25715, DYNJ25757B, DYNJ25792A, DYNJ25795, DYNJ25796A, DYNJ25821A, DYNJ25825B, DYNJ25827A, DYNJ25829A, DYNJ25831B, DYNJ25891D, DYNJ25896B, DYNJ25896B, DYNJ25908, DYNJ25925, DYNJ25929, DYNJ25935A, DYNJ25936A, DYNJ25941A, DYNJ25957, DYNJ25987B, DYNJ26045A, DYNJ26045A, DYNJ26078A, DYNJ26100A, DYNJ26123A, DYNJ26171C, DYNJ26172C, DYNJ26178A, DYNJ26181A, DYNJ26189, DYNJ26190A, DYNJ26191A, DYNJ26192B, DYNJ26193B, DYNJ26196A, DYNJ26197, DYNJ26237A, DYNJ26241B, DYNJ26257B, DYNJ26260A, DYNJ26277, DYNJ26278, DYNJ26289, DYNJ26294, DYNJ26294, DYNJ26325, DYNJ26394A, DYNJ26420A, DYNJ26448, DYNJ26453A, DYNJ26455, DYNJ26459B, DYNJ26566, DYNJ26570, DYNJ26573, DYNJ26608, DYNJ26635A, DYNJ26685, DYNJ26724A, DYNJ26726, DYNJ26730, DYNJ26732, DYNJ26733A, DYNJ26750, DYNJ26753, DYNJ26757, DYNJ26761, DYNJ26764, DYNJ26776, DYNJ26786B, DYNJ26848, DYNJ26848A, DYNJ26849A, DYNJ26850, DYNJ26852, DYNJ26853, DYNJ26873, DYNJ26891, DYNJ26901, DYNJ26901A, DYNJ26922, DYNJ26937, DYNJ26956A, DYNJ26964, DYNJ26964A, DYNJ26988, DYNJ26993, DYNJ27055, DYNJ27073, DYNJ27088, DYNJ27110, DYNJ27154, DYNJ27154, DYNJ27160, DYNJ27165, DYNJ27173, DYNJ27234, DYNJ27273, DYNJ27278, DYNJ27309, DYNJ27313, DYNJ27315, DYNJ27337, DYNJ27412, DYNJ27438, DYNJC1028F, DYNJC1030G, DYNJC1131B, DYNJC1254J, DYNJC1267F, DYNJC1361G, DYNJC1382D, DYNJC1517D, DYNJC1520C, DYNJC1541G, DYNJC1542C, DYNJC1639G, DYNJC1657G, DYNJC1748B, DYNJC1753F, DYNJC1754A, DYNJC1834G, DYNJC1877A, DYNJC1910F, DYNJC1912C, DYNJC1916B, DYNJC1981F, DYNJC1986N, DYNJC2023I, DYNJC2202N, DYNJC2229L, DYNJC2237A, DYNJC2274G, DYNJC2281F, DYNJC2287B, DYNJC2288, DYNJC2338B, DYNJC2373B, DYNJC2375B, DYNJC2393A, DYNJC2399A, DYNJC2413B, DYNJC2414B, DYNJC2418, DYNJC2419G, DYNJC2420D, DYNJC2422B, DYNJC2427B, DYNJC2431G, DYNJC2431I, DYNJC2460A, DYNJC2484C, DYNJC2485, DYNJC2588D, DYNJC2606G, DYNJC2684D, DYNJC2701A, DYNJC2714A, DYNJC2715A, DYNJC2758J
DYNJC2759I , DYNJC2760I, DYNJC2762J, DYNJC2763G, DYNJC2764G, DYNJC2766G, DYNJC2767I, DYNJC2768N, DYNJC2769I, DYNJC2794B, DYNJC2821, DYNJC2848C, DYNJC2915D, DYNJC2956D, DYNJC2985A, DYNJC2994D, DYNJC3091F, DYNJC3110C, DYNJC3117D, DYNJC3118G, DYNJC3137A, DYNJC3143G, DYNJC3184D, DYNJC3186C, DYNJC3187C, DYNJC3188D, DYNJC3191C, DYNJC3225C, DYNJC3256A, DYNJC3259B, DYNJC3276, DYNJC3279C, DYNJC3280B, DYNJC3292C, DYNJC3294A, DYNJC3299A
DYNJC3304A, DYNJC3349C, DYNJC3360A, DYNJC3362A, DYNJC3367C, DYNJC3380A, DYNJC3417, DYNJC3418, DYNJH1054I, DYNJH1060B, DYNJH1158A, DYNJH1223C, DYNJH1227A, DYNJH1235F, DYNJH1236, DYNJHS0116F, DYNJHS0145A, DYNJHS0181C, DYNJHS0570, DYNJHS0597A, DYNJHS0597B, DYNJHS0604C, DYNJHS0605C, DYNJHS0628, DYNJHS0709B, DYNJQ0009A, DYNJQ0020A, DYNJQ0029D, DYNJQ0043, DYNJQ0059F, DYNJQ0060B, DYNJQ0063B, DYNJQ0064F, DYNJQ0097A, DYNJQ0100A, DYNJQ0114, DYNJQ0118B, DYNJQ0119B, DYNJQ0120A, DYNJQ0123A, DYNJQ0124B, DYNJQ0127, DYNJQ0128A, DYNJQ0131A, DYNJQ0140C, DYNJQ0157D, DYNJQ0162B, DYNJQ0171A, DYNJQ0172, DYNJQ0177A, DYNJQ0178C, DYNJQ0180B, DYNJQ0185A, DYNJQ0191F, DYNJQ0207, DYNJQ0221, DYNJQ0226C, DYNJQ0227B, DYNJQ0234A, DYNJQ0235A, DYNJQ0237A, DYNJQ0240A, DYNJQ0242A, DYNJQ0244B, DYNJQ0271, DYNJQ0286D, DYNJQ0299A, DYNJQ0301A, DYNJQ0304A, DYNJQ0306B, DYNJQ0308B, DYNJQ0320A, DYNJQ0338A, DYNJQ0365A, DYNJQ0369A, DYNJQ0370A, DYNJQ0376B, DYNJQ0405A, DYNJQ0411A, DYNJQ0415, DYNJQ0439B, DYNJQ0461A, DYNJQ0486, DYNJQ0497, DYNJQ0508, DYNJQ0526, DYNJQ0529, DYNJQ0541A, DYNJQ0543, DYNJQ0564, DYNJQ0567, DYNJQ0578, DYNJQ0584, DYNJSB0013, DYNJT0017G, DYNJT0018I, DYNJT0019I, DYNJT0055B, DYNJT0056A, DYNJT0087B, DYNJT0162D, DYNJT0225C, DYNJT0258A, DYNJT0483A, DYNJT0505B, DYNJT0507F, DYNJT0616B, DYNJT0617A, DYNJT0627G, DYNJT1076G, DYNJT1077J, DYNJT1081G, DYNJT1083F, DYNJT1084B, DYNJT1085G, DYNJT1107B, DYNJT1185C, DYNJT1187F, DYNJT1259B, DYNJT1273, DYNJT1282B, DYNJT1299B, DYNJT1302B, DYNJT1416F, DYNJT1459, DYNJT1485B, DYNJT1589A, DYNJT1678A, DYNJT1679C, DYNJT1695A, DYNJT1745F, DYNJT1759B, DYNJT1877B, DYNJT1878, DYNJT1901C, DYNJT1926B, DYNJT1927A, DYNJT1936, DYNJT1938, DYNJT1943A, DYNJT1945A, DYNJT1947, DYNJT1972D, DYNJT2003, DYNJT2014C, DYNJT2110B, DYNJT2119A, DYNJT2121B, DYNJT2144A, DYNJT2159C, DYNJT2179, DYNJT2180D, DYNJT2185A, DYNJT2248A, DYNJT2262, DYNJT2388A, DYNJT2432A, DYNJT2444, DYNJT2445, DYNJT2484C, DYNJV0039F, DYNJV0067I, DYNJV0069C, DYNJV0071G, DYNJV0073C, DYNJV0076A, DYNJV0078C, DYNJV0079D, DYNJV0083C, DYNJV0098B, DYNJV0100D, DYNJV0107F, DYNJV0160C, DYNJV0196B, DYNJV0218, DYNJY0001A, DYNJY0003A, DYNJY0010, DYNJY0019B, DYNJY0041B, DYNJY0042A, DYNJY0050B, DYNJY0056B, DYNJY0057C, DYNJY0059B, DYNJY0060C, DYNJY0064D, DYNJY0067C, DYNJY0069, DYNJY0074B, DYNJY0075A, DYNJY0080A, DYNJY0083B, DYNJY0085A, DYNJY0090, DYNJY0097A, DYNJY0098A, DYNJY0102A, DYNJY0110B, DYNJY0112A, DYNJY0118B, DYNJY0119, DYNJY0121, DYNJY0130A, DYNJY0131A, DYNJY0135, DYNJY0141A, DYNJY0144, DYNJY0150A, DYNJY0151A, DYNJY0152A, DYNJY0153, DYNJY0162B, DYNJY0166A, DYNJY0167A, DYNJY0168, DYNJY0169, DYNJY0173B, DYNJY0179B, DYNJY0181B, DYNJY0183B, DYNJY0184B, DYNJY0186A, DYNJY0195A, DYNJY0203C, DYNJY0207B, DYNJY0210, DYNJY0216, DYNJY0217A, DYNJY0218A, DYNJY0220, DYNJY0221, DYNJY0222, DYNJY0225, DYNJY0227, DYNJY0228, DYNJY0236, DYNJY0238A, DYNJY0239, DYNJY0240, DYNJY0241, DYNJY0244B, DYNJY0246, DYNJY0247, DYNJY0254A, DYNJY0276, DYNJY0277 |
Recalling Firm/ Manufacturer |
Medline Industries, Inc. 1170 S Northpoint Blvd Waukegan IL 60085-6757
|
For Additional Information Contact | Ms. Lara M. Simmons 800-879-2791 Ext. 6072 |
Manufacturer Reason for Recall | There is a potential for open side seals post-sterilization of the Tyvek breather pouches used in the custom sterile packs. |
FDA Determined Cause 2 | Other |
Action | Field Correction letters dated 3/4/05 were sent to thedirect account hospitals on the same date to the attention of the manager of surgery, informing them of the potential indadequate side seal of the Tyvek breather pouches in the packs which may cause a compromise in the sterility of the product. The accounts were provided a list of the affected pack sent to the account and were requested to inspect any stocks of the affected product in stock, using the enclosed diagrams as an aid in the inspection for the defective side seals. Any defective product found should be returned to Medline. Any questions were directed to Lara Simmons, QA/RA Director at 1-800-879-2781, ext. 6072. |
Quantity in Commerce | 31,497 cases |
Distribution | Nationwide |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|