| Date Initiated by Firm | June 16, 2005 |
| Date Posted | August 03, 2005 |
| Recall Status1 |
Terminated 3 on November 17, 2005 |
| Recall Number | Z-1065-05 |
| Recall Event ID |
32495 |
| PMA Number | P010012S008 |
| Product Classification |
Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D) - Product Code NIK
|
| Product | CONTAK RENEWAL 3 DS HE, model H177 CRT-D Cardiac Resynchronization Therapy Defibrillator. Contents have been Sterilized with gaseous ethylene oxide. Guidant Corporation Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798, USA. Provides ventricular tachyarrhythmia and cardiac resynchronization therapies. |
| Code Information |
508822, 508827 and 508862 |
Recalling Firm/ Manufacturer |
Guidant Corporation 4100 Hamline Ave N Saint Paul MN 55112-5700
|
Manufacturer Reason for Recall | Devices were manufactured with PEEK insulation material on the feedthrough wires and distributed prior to PMAS approval. |
FDA Determined Cause 2 | Other |
| Action | Sales Representatives notified 06/16/05 to return devices in their area. |
| Quantity in Commerce | 3 |
| Distribution | CA, IL, MI, MO, OR and PA |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| PMA Database | PMAs with Product Code = NIK
|