Class 1 Device Recall Colleague and Colleague CX Volumetric Infusion Pumps |
|
Date Initiated by Firm |
July 20, 2005 |
Date Posted |
August 10, 2005 |
Recall Status1 |
Terminated 3 on June 01, 2011 |
Recall Number |
Z-1059-05 |
Recall Event ID |
32678 |
510(K)Number |
K041191
|
Product Classification |
infusion pump - Product Code FRN
|
Product |
Baxter Colleague Single Channel Volumetric Infusion Pumps; Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015 U.S.A., Made in Singapore; product codes 2M8151, 2M8151R, 2M8161, 2M8161R |
Code Information |
all serial numbers |
Recalling Firm/ Manufacturer |
Baxter Healthcare Corp. Rt. 120 & Wilson Rd Round Lake IL 60073
|
For Additional Information Contact |
Center for One Baxter 800-422-9837
|
Manufacturer Reason for Recall |
A hardware problem can cause internal communications errors which halt therapies, generate alarms and communication failure codes.
|
FDA Determined Cause 2 |
Device Design |
Action |
Baxter sent the 7/20/05 Urgent Product Recall letter to all Colleague Infusion Pump customers, to the attention of the Director of Nursing and the Biomedical Engineer, via overnight delivery to alert them of a patient death which may have been associated with an electronic failure resulting from a design issue with the pump, and listed failure codes 402, 403, 532, 533, 534, 535, 702, 703, 704, 720, 804:22, 804:24, 804:29, 804:34, 804:52, 804:54, 804:58 and 12:303:xxx:0006 that may be related to the electronic failure. The users were advised to have a contingency plan to mitigate any disruptions of infusions of life sustaining drugs if the listed failure codes occur, and to take the pump out of service. They were also requested to review the event history of their pumps and take any pump with a previous history of the listed failures out of service. The accounts were also advised that all pumps processed through Baxter's service operations will be checked for the listed failure codes, and if a pump is found to have any of the codes in its event history, it will not be returned until a corrective action has been implemented. Any questions were directed to Baxter's Medication Delivery Services at 1-800-843-7867. |
Quantity in Commerce |
152,260 pumps |
Distribution |
Nationwide including Puerto Rico, and internationally to Australia, the Bahamas, Bahrain, Belgium, Brazil, Canada, Colombia, El Salvador, Guatemala, Honduras, Israel, Lebanon, Mexico, New Zealand, Panama, the Philippines, Saudi Arabia, South Africa, Taiwan, Turkey, the United Arab Emirates and teh United Kingdom. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = FRN and Original Applicant = BAXTER HEALTHCARE CORP.
|
|
|
|