Date Initiated by Firm | November 18, 2005 |
Date Posted | January 06, 2006 |
Recall Status1 |
Terminated 3 on January 12, 2012 |
Recall Number | Z-0352-06 |
Recall Event ID |
34087 |
510(K)Number | K974597 |
Product Classification |
System, X-Ray, Stationary - Product Code KPR
|
Product | ADC Solo & Initia CR 1000, Path Speed CR SP1001, Centricity CR SP1001 and Centricity CR SP 100, image digitizer, components in Agfa''s computed radiography product line. |
Code Information |
Serial Numbers: ADC Solo: Units with Serial numbers les than 4748, Initia CR1000, Path Speed CR SP1001, Centricity CR SP1001 and Centricity CR SP100: all serial numbers are considered affected unless they have been upgraded to software version SOL_2205 or later. |
Recalling Firm/ Manufacturer |
AGFA Corp. 10 S Academy St Greenville SC 29601-2632
|
For Additional Information Contact | Ray B. Myers Ph.D 864-864-1827 |
Manufacturer Reason for Recall | Users of affected systems may experience a loss of images, the need to repeat the imaging procedure and possibly a short delay in diagnosis. |
FDA Determined Cause 2 | Other |
Action | Consignees were notified by letter on/about December 1, 2005. |
Quantity in Commerce | 1059 units |
Distribution | Nationwide |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = KPR
|