Class 2 Device Recall Baxter Interlink System Extension Set with Standard Bore 4Way Stopcock Manifold Extension Set and M

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Baxter Interlink System Extension Set with Standard Bore 4Way Stopcock Manifold Extension Set and Msee related information
Date Initiated by FirmDecember 15, 2005
Date PostedMarch 24, 2006
Recall Status1 Terminated 3 on August 18, 2008
Recall NumberZ-0646-06
Recall Event ID 34290
510(K)NumberK022895 K811078 K921899 K924721 K962581 
Product Classification Stopcock, I.V. Set - Product Code FMG
ProductBaxter Interlink System Extension Set with Standard Bore 4-Way Stopcock Manifold Extension Set and Minivolume Extension Set, product code 3C0071; An Rx sterile, nonpyrogenic fluid pathway; extension set: 19'', 3.2 mL Vol., with injection site and male Luer Lock adapter, manifold extension set: 46'', 6.2 mL Vol, male luer adapter, and minivolume extension set: 10'', 0.7 mL Vol, injection site and stationary swivel lock T-connector; Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A.
Code Information product code 3C0071, lots GR226944, GR231183, GR224683
Recalling Firm/
Manufacturer
Baxter Healthcare Corp.
Rt. 120 & Wilson Rd
Round Lake IL 60073
For Additional Information ContactCenter for One Baxter
800-422-9837
Manufacturer Reason
for Recall
The extension sets contain microbore female luers that may crack during use, resulting in leakage of medication or incomplete delivery of medication.
FDA Determined
Cause 2
Other
ActionBaxter sent Urgent Product Recall letters dated 8/3/05 to their customers, to the attention of the Director of Materials Management, via first class mail on the same date. The accounts were informed of the microbore female luers that may crack during use, resulting in the possible under delivery or incomplete delivery of medication. The accounts were instructed to examine their inventory for the affected lots of product, remove it from inventory and return it to Baxter for credit. Any questions were directed to the Center for One Baxter at 1-800-422-9837. Baxter extended the recall to include additional lots and product codes via letter dated 12/15/05. The letters referenced the 8/3/05 letter, including a copy of that letter, and listed the affected lots of sets. The instructions for return of the product remained the same as that described in the 8/3/05 letter.
Quantity in Commerce5,856 sets
DistributionNationwide, including Puerto Rico and American Samoa, and internationally to Canada, Australia, Japan, South Korea, Saudi Arabia, India, Mexico, Venezuela, Malaysia, East Timor and New Zealand.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = FMG
510(K)s with Product Code = FMG
510(K)s with Product Code = FMG
-
-