• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Liko

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Likosee related information
Date Initiated by FirmJanuary 09, 2006
Date PostedApril 26, 2006
Recall Status1 Terminated 3 on October 25, 2006
Recall NumberZ-0790-06
Recall Event ID 34463
Product Classification Lift, Patient, Non-Ac-Powered - Product Code FSA
ProductLiko Universal Slingbar 450 (aluminum) used with the Viking M Patient Lift
Code Information Serial Numbers: 7100102 - 7100104;   7100109  7100147 - 7100159  7100179 - 7100182  7100257 - 7100261  EXPANDED RECALL: Viking M:  Serial Numbers: 7100101 through Z-7100381
FEI Number 3001451524
Recalling Firm/
Manufacturer
Liko, Inc.
842 Upper Union Street, Suite 4
Franklin MA 02038
For Additional Information ContactAnn Jordan
508-553-3993
Manufacturer Reason
for Recall
The swivel adapter which is mounted directly above the slingbar's split center bolt could be removed and allow the slingbar to be improperly mounted directly into the aluminum link arm. This could cause an unsafe transfer. Patient may fall.
FDA Determined
Cause 2
Other
ActionLiko notified Distributors of the recall by letter on 1/9/06 via UPS and provided replacement Slingbars Version 2. Expanded Recall: On 6/20/06, Liko sent a 2nd recall letter expanding the recall to additonal serial numbers via Certified Mail.
Quantity in Commerce67 units
DistributionIL, MA, MO, PA, TN,
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-