| Date Initiated by Firm | February 02, 2006 |
| Date Posted | March 18, 2006 |
| Recall Status1 |
Terminated 3 on June 17, 2008 |
| Recall Number | Z-0661-06 |
| Recall Event ID |
34569 |
| 510(K)Number | K042536 |
| Product Classification |
Oximeter - Product Code DQA
|
| Product | Rainbow DC-I-DC Pulse CO-Oximeter Sensor used with the Masimo SET Rad 57 Pulse CoOximeter |
| Code Information |
Serial Numbers: 4793 5108 5405 5792 6096 6261 6421 6576 4794 5109 5506 5797 6118 6263 6436 6578 4818 5115 5513 5800 6136 6265 6441 6596 4821 5123 5530 5811 6140 6266 6444 6604 4833 5139 5538 5816 6146 6268 6451 6608 4836 5169 5571 5828 6148 6274 6463 6628 4852 5176 5579 5838 6149 6277 6473 6638 4854 5182 5588 5839 6153 6283 6478 6645 4860 5186 5589 5841 6158 6287 6481 6664 4866 5187 5592 5845 6159 6288 6482 6666 4870 5191 5597 5848 6164 6302 6483 6668 4875 5194 5599 5852 6166 6304 6484 6671 4880 5196 5601 5856 6168 6309 6487 6673 4881 5197 5608 5868 6169 6310 6488 6678 4886 5199 5610 5870 6172 6311 6489 6680 4887 5201 5613 5872 6173 6313 6492 6683 4895 5219 5616 5877 6174 6318 6493 6686 4898 5220 5619 5884 6178 6320 6495 6690 4912 5221 5620 5934 6179 6321 6498 6696 4914 5224 5621 5939 6180 6324 6503 6697 4915 5226 5625 5949 6183 6325 6509 6743 4933 5233 5635 5960 6189 6326 6511 4934 5235 5649 5961 6196 6329 6514 4939 5236 5652 5966 6198 6333 6515 4968 5237 5656 5971 6199 6337 6517 4992 5238 5657 5979 6200 6338 6518 4994 5249 5662 5986 6201 6341 6520 5002 5252 5663 6017 6205 6343 6521 5013 5269 5664 6026 6207 6346 6525 5017 5274 5668 6029 6208 6353 6527 5022 5275 5679 6030 6209 6354 6529 5029 5278 5681 6034 6215 6363 6538 5031 5279 5682 6039 6217 6367 6539 5043 5281 5683 6040 6218 6368 6541 5049 5287 5690 6049 6220 6369 6542 5051 5288 5692 6051 6223 6374 6547 5052 5291 5693 6053 6224 6380 6550 5053 5317 5697 6056 6225 6381 6551 5054 5319 5699 6061 6230 6382 6552 5058 5321 5700 6065 6231 6385 6554 5061 5340 5704 6067 6232 6388 6555 5063 5361 5710 6070 6233 6392 6557 5069 5370 5724 6071 6235 6393 6559 5070 5371 5729 6074 6237 6394 6560 5075 5374 5744 6078 6239 6395 6566 5077 5376 5746 6079 6246 6397 6570 5091 5378 5750 6080 6247 6411 6571 5093 5380 5757 6084 6248 6413 6573 5096 5391 5758 6089 6256 6415 6574 5097 5398 5764 6094 6257 6419 6575 |
| FEI Number |
3001452206
|
Recalling Firm/ Manufacturer |
Masimo Corporation
40 Parker
Irvine
CA
92618-1604
|
| For Additional Information Contact | James J. Cronin 949-297-7218 |
Manufacturer Reason for Recall | Internal testing revealed that some of the sensors might not meet the accuracy specifications of : 2% for SpO2 and of: 3% for SpCO, due to the tolerance of one of the emitter LED's used in the calculation ofSpO2 and SpCO. The SpO2 reading may have a negative bias of up to 1 % and the SpCO reading may have a negative bias of up to 4%.
|
FDA Determined Cause 2 | Other |
| Action | Masimo notified all customers by letter sent via Federal Express who were shipped Rainbow DC- I-DC sensors to return them to Masimo for repair or replacement.
A Tracking/Verification form will be mailed with the letter to be faxed back to Masimo.
Masimo will repair or replace the returned units and return them back to the customer.
|
| Quantity in Commerce | 321 |
| Distribution | Nationwide and Belgium, Canada, France, Germany, Italy, South Korea, new Zealand, Slovenia, Spain, Sweden, Switzerland, and United Arab Emirates |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|