|
Class 2 Device Recall LIFEPAK 20 |
|
Date Initiated by Firm |
March 22, 2006 |
Date Posted |
April 04, 2006 |
Recall Status1 |
Terminated 3 on September 06, 2011 |
Recall Number |
Z-0711-06 |
Recall Event ID |
34947 |
510(K)Number |
K012274
|
Product Classification |
Dc-Defibrillator, Low-Energy, (Including Paddles) - Product Code LDD
|
Product |
LIFEPAK 20 defibrillator/monitor |
Code Information |
Domestic Serial Numbers: 33884879, 33884872, 33884912, 33884871, 33884867, 33884866, 33829105, 33829090, 33950589, 33911492, 33946302, 33884898, 33951494, 33980272, 33986554, 33818744, 33818718, 33924702, 33881437, 33905482, 33894423, 33951266, 33905466, 33955767, 33900980, 33980275, 33980280, 33980287, 33980281, 33819680, 33829407, 33971829, 33986933, 33967385, 34011822, 34011819, 33894426, 33894572, 33900981, 33936889, 33884924, 33826068, 33981789, 33981799, 33981792, 33981805, 33981794, 33981796, 33981793, 33981795, 33981797, 33959538, 33938363, 34011821, 33894425, 33959550, 33938924, 33938921, 33922299, 33930122, 33933904, 33826017, 33955760, 33897337, 33955747, 33951295, 33951456, 33976455, 33951447, 33951275, 33979473, 33979470, 33979471, 33979472, 33979474, 33979469, 33976420, 33976429, 33976413, 33976414, 33976415, 33976416, 33976411, 33976418, 33976410, 33976424, 33976426, 33976427, 33976417, 33976412, 33976430, 33976406, 33976407, 33976408, 33976428, 33976431, 33976419, 33933878, 33946268, 33946277, 33897329, 33905622, 33905621, 33905609, 33911477, 33911470, 33911450, 33911449, 33911446, 33905590, 33905585, 33905581, 33911571, 33897342, 33897320, 33897322, 33905514, 33911528, 33911627, 33905518, 33955832, 33955765, 33955759, 33951605, 33951484, 33951059, 33951467, 33951473, 33951487, 33955761, 33894580, 33905444, 33951128, 33950588, 33951434, 33951273, 33951287, 33951551, 33826023, 33826022, 33826009, 33826008, 33826005, 33826003, 33826021, 33915964, 33915884, 33922357, 33911587, 33986931, 33997554, 33971953, 33955806, 33955833, 33826029, 33951560, 33955752, 33951589, 33951588, 33951586, 33951585, 33951583, 33951576, 33951595, 33951569, 33951580, 33951604, 33955751, 33955748, 33951592, 33951593, 33951601, 33951600, 33951599, 33951598, 33951597, 33951582, 33951409, 33951403, 33951293, 33951292, 33951468, 33951462, 33951553, 33951552, 33951535, 33951497, 33951518, 33959570, 33946296, 31014367, 33997567, 33904931, 33904929, 33904928, 33904926, 33904923, 33904921, 33904936, 33904917, 33904914, 33904913, 33904910, 33904922, 33905443, 33905441, 33905440, 33905434, 33905432, 33904933, 33905430, 33904932, 33905427, 33904938, 33904937, 33904876, 33905470, 33905488, 33905474, 33905473, 33901782, 33901777, 33901773, 33901771, 33901770, 33905451, 33901787, 33901762, 33901753, 33901761, 33911601, 33915916, 33915917, 33915910, 33915961, 33959578, 33959606, 33897319, 33905516, 33905507, 33922294, 33911635, 33911634, 33911632, 33911631, 33911611, 33911624, 33911614, 33917454, 33917452, 33917448, 33917447, 33917444, 33917443, 33917440, 33894289, 33905521, 33905520, 33917455, 33917453, 33917470, 33921055, 33921053, 33921052, 33921050, 33921049, 33921047, 33921046, 33917468, 33917467, 33917466, 33917465, 33917464, 33917463, 33917461, 33917457, 33917456, 33921045, 33955771, 33955762, 33955750, 33951603, 33986555, 33986346, 33986324, 33979488, 33980264, 33986878, 33986897, 33981778, 33169330, 33904924, 33911468, 33905495, 33905496, 33905459, 33905587, 33905497, 33933906, 33901735, 33894413, 33955802, 33955839, 33955831, 33955829, 33955827, 33955803, 33938926, 33951596, 33951426, 33951405, 33997552, 33986873, 33986877, 33905438, 33924704, 33904878, 33901737, 33950587, 33951298, 33951283, 33976463, 33980266, 33980290, 33980294, 33980278, 33980279, 33980291, 33980285, 33980301, 33980304, 33986548, 33976478, 33976479, 33976485, 33976477, 33976468, 33976470, 33976471, 33976456, 33976446, 33976447, 33819634, 33947249, 33947931, 33947264, 33915931, 33915930, 33915929, 33915928, 33915925, 33915889, 33915887, 33915888, 33915899, 33915898, 33915897, 33915895, 33915894, 33915893, 33915892, 33915890, 33911603, 33911602, 33911597, 33936877, 33936874, 33915914, 33915919, 33915922, 33915912, 33915911, 33915908, 33980268, 33980300, 33986563, 33986561, 33986564, 33986565, 33986562, 33986543, 33986351, 33986330, 33986342, 33986349, 33986348, 33986345, 33986338, 33986344, 33971910, 33971905, 33971901, 33971838, 33897350, 33897344, 33884882, 33884875, 33884911, 33990265, 34015699, 31368552, 33986913, 33911532, 33911535, 33936914, 33971950, 33894290, 33911625, 33986559, 33986549, 33986551, 33986323, 33911595, 33915885, 33933909, 33826019, 33826001, 33825999, 33826000, 33823372, 33823373, 33823375, 33823342, 33823340, 33823348, 33823367, 33823366, 33823358, 33823356, 33829111, 33823382, 33823386, 33823381, 33823378, 33823384, 33951602, 33951449, 33986558, 33986556, 33986553, 33986325, 33986339, 33818733, 33951572, 33951568, 33951439, 33951556, 33980297, 33897327, 33980274, 33915956, 33826048, 33951476, 31388033, 33905592, 33905589, 33990271, 34011832, 33933911, 33905582, 33894597, 33955840, 33915926, 33915939, 33933931, 33884958, 33997573, 33997566, 33946303, 33930103, 33924697, 33924699, 33924696, 33924695, 33924703, 33924705, 33924706, 33924692, 33930102, 33930105, 33897353, 33924701, 33818749, 33946314, 33930107, 33924684, 33924682, 33955755, 33986911, 34011824, 34017741, 33980296, 33986546, 33986334, 33986343, 33986335, 33986337, 33915896, 33819582, 33823323, 33898439, 33898437, 33818727, 33951475, 33951517, 33980298, 33915962, 33901774, 33986894, 33986879, 33986880, 33971926, 33971913, 33818750, 33818774, 33986871, 33911512, 33936879, 33826056, 33819681, 33819669, 33951430, 33955775, 33884945, 33885006, 33884962, 33826069, 33826084, 33818739, 33971918, 33971867, 33986898, 33826097, 33900862, 33900986, 33900985, 33900983, 33901091, 33900859, 33900858, 33900763, 33900762, 33900864, 33900987, 33901090, 33900761, 33898455, 33900759, 33971921, 33986886, 33986901, 33900982, 33900979, 33900856, 33900765, 33901122, 33901120, 33901117, 33901100, 33901099, 33901098, 33901116, 33901127, 33901769, 33901768, 33901725, 33901733, 33901732, 33901731, 33901729, 33901726, 33901722, 33901720, 33901717, 33901128, 33905616, 33905603, 33905600, 33911445, 33905598, 33901785, 33911538, 33911516, 33911539, 31138863, 30559175, 33976453, 33986896, 33986567, 33986881, 33976464, 33976458, 33947259, 33946315, 33936899, 33967359, 33922369, 33967350, 33967360, 33936906, 33922377, 33922356, 33967379, 33955816, 33900984, 33904927, 33901092, 33823337, 33936893, 33884926, 33951280, 33951457, 30827240, 33955764, 33955772, 33955769, 33955768, 33955763, 33955758, 33955757, 33955766, 33955774, 33951297, 33951543, 33901115, 33818783, 33955838, 33829109, 33980283, 33818712, 33951472, 33951288, 33967362, 33947910, 33933925, 33924661, 30498162, 34011820, 33823339, 33818743, 33897340, 33911638, 33897331, 33951285, 33915958, 33915953, 33917442, 33922383, 33915921, 33922372, 33905526, 33905517, 33897343, 33897321, 33897325, 33897328, 33905511, 33905509, 33905513, 33905501, 33905503, 33911641, 33911633, 33911628, 33917450, 33911610, 33917441, 33905522, 33911604, 33905525, 33905519, 33889433, 30569241, 33901106, 33901096, 33951118, 33951489, 33946322, 33946307, 33946319, 33947256, 33933922, 33933920, 33933917, 33933915, 33933916, 33976487, 33979475, 33979479, 33955818, 33955813, 33900855, 33818776, 33905428, 33818721, 33818714, 33818726, 33818729, 33818751, 33818752, 33818753, 33946305, 33946317, 33946311, 33947900, 33936878, 33933905, 33967377, 33971822, 33971814, 33971813, 33967411, 33967410, 33967409, 33967406, 33967396, 33967388, 33967387, 33967384, 33967383, 33967398, 33971823, 33971826, 33971833, 33897336, 33897323, 33897318, 33905512, 33905515, 33905506, 33911626, 33911623, 33911616, 33911615, 33911612, 33911605, 33905523, 33905524, 33986340, 33980293, 33947899, 33933907, 33936849, 33933932, 33905611, 33894410, 33894585, 33951469, 33947884, 33884980, 33889427, 33905580, 33946301, 33884970, 33911596, 33980276, 33986560, 33986895, 33997570, 33986883, 33986350, 33986910, 33986899, 33986908, 33986547, 33986331, 33971906, 33971903, 33971891, 33971877, 33971836, 33971902, 33951474, 33884923, 33826088, 34017732, 33997579, 33946297, 34015705, 34015711, 34011828, 34017743, 34017737, 33997561, 33904916, 33904912, 33904911, 33900978, 33901126, 33900861, 33901124, 33901119, 33901095, 33900758, 33900710, 33901781, 33901780, 33901779, 33901778, 33901125, 33901767, 33901766, 33901765, 33901764, 33901763, 33901791, 33901790, 33901736, 33901734, 33901730, 33901738, 33901739, 33901724, 33901723, 33901721, 33901719, 33901718, 33901727, 33901749, 33901759, 33901758, 33901757, 33901756, 33901755, 33901754, 33901752, 33901750, 33901748, 33901747, 33901746, 33901745, 33901744, 33901743, 33901742, 33901741, 33901751, 33947252, 33933927, 33936854, 33936857, 33933926, 33933928, 33933923, 33933921, 33971923, 33971919, 33955809, 33938360, 33938365, 33951130, 33951471, 33951284, 33905626, 33911509, 33905499, 33905586, 33946299, 33936896, 33898430, 33898438, 33898434, 33951270, 33911483, 33979550, 33979554, 33986791, 33894418, 33894415, 33980271, 33980273, 33955819, 33955852, 33947890, 33959540, 33947255, 33947253, 33946320, 33946312, 33951581, 33951291, 33951465, 33951452, 33951549, 33951504, 33951516, 33951558, 33951499, 33894582, 34011817, 33818722, 33955812, 33826090, 33911562, 33947920, 33986893, 33997580, 33997571, 33990252, 34011830, 33901121, 33959539, 30559192, 33946306, 33946316, 33947248, 33947925, 33947919, 33947917, 33947929, 33947257, 33947894, 33947893, 33947260, 33967342, 33933910, 33933914, 33933919, 33933918, 33967414, 33826053, 33819672, 33997577, 33911435, 33986909, 33986900, 33898504, 33911613, 33911606, 33981770, 33894594, 33825998, 33951267, 33915941, 33894409, 33905614, 33911520, 33951440, 33951121, 33915951, 33819665, 33819611, 33936852, 33933888, 33905625, 33905617, 33905599, 33905615, 33911502, 33911462, 33911448, 33911447, 33905596, 33911518, 33911533, 33911536, 33911555, 33884974, 33951453, 33894592, 33894419, 33894591, 33894583, 33905508, 33898505, 33905505, 33959559, 33826027, 33915909, 30706023, 33951442, 33959556, 33955815, 33955792, 33955841, 33955828, 33955814, 33951420, 33951450, 33976432, 33971900, 34017735, 34015709, 33915944, 33986916, 33997551, 33980295, 33986545, 33986336, 33819636, 33818758, 33911510, 33818762, 33951458, 33823371, 33823370, 33823376, 33823345, 33823352, 33823350, 33823349, 33823354, 33823346, 33823343, 33823362, 33823369, 33823368, 33823353, 33823363, 33823359, 33823357, 33823379, 33905595, 33938922, 33951444, 30540230, 33826046, 30874406, 33951466, 33823338, 33826041, 33915945, 33911530, 33823341, 33904934, 33911452, 33959555, 33959554, 33981791, 33959605, 33959546, 33981798, 33955848, 33938364, 33959567, 33959549, 33823344, 33955753, 33951131, 33951407, 33951464, 33826013, 33826010, 33823365, 33894416, 33955776, 30476106, 30573033, 34015715, 34015713, 34011823, 34011827, 33990272, 33990268, 33990254, 34015707, 34015704, 34015698, 34011833, 34015697, 34015751, 34017736, 34017734, 34017729, 34015720, 34015718, 34015717, 34011812, 34011811, 34011814, 33884953, 33889416, 33829400, 33905445, 33905475, 33905471, 33905491, 33905578, 33894569, 33826040, 33826004, 33826042, 33938923, 33818754, 33818736, 33911526, 33915936, 33915940, 33915955, 33915954, 33915952, 33915947, 33915915, 33897333, 33905510, 33889430, 33826087, 33981771, 33980267, 30587187, 33915937, 33915935, 33915949, 30540981, 30880107, 30523113, 30416301, 30493029, 33915900, 33915913, 33950577, 33826055, 33826054, 33826052, 33819668, 33818760, 33951289, 31005264, 31003144, 33905623, 33884876, 33976423, 33818738, 33884937, 33980282, 33911598, 33951511, 33951491, 33884950, 33829119, 33951092, 33976451, 33884930, 33905583, 33905579, 34015708, 33894424, 33894417, 33976433, 33823360, 30569230, 33947933, 33971920, 33971925, 33971924, 33971949, 33971940, 33971937, 33971917, 33971884, 30559186, 33947916, 33947943, 33947923, 33947908, 33947907, 33947905, 33947903, 33947902, 33947901, 33947898, 33947262, 33947913, 33967346, 33967344, 33967343, 33936894, 33823364, 33980303, 33986557, 33980286, 33971946, 33971904, 33951522, 33951515, 33884959, 33826018, 33826007, 33915950, 33971954, 33884925, 33884965, 33884957, 33889423, 33889419, 33826081, 33829129, 33959564, 33959563, 33959565, 33955797, 33936910, 33894414, 33951546, 33986552, 33986929, 33946308, 33947254, 33967373, 33967366, 33967364, 33967357, 33967353, 33967392, 33915938, 33915918, 33915920, 33911441, 33911442, 33905584, 33905429, 33905439, 33905437, 33905426, 33904939, 33905431, 33905450, 33905612, 33905448, 33911443, 33905607, 33905604, 33905602, 33905601, 33911508, 33911472, 33911451, 33911466, 33905490, 33905489, 33905481, 33905479, 33905469, 33905465, 33905464, 33905594, 33905593, 33905588, 33905597, 33905498, 33901775, 33901789, 33911567, 33911576, 30553853, 33911527, 33911515, 33911513, 33911521, 33911522, 33911523, 33911524, 33911529, 33911531, 33911534, 33911537, 33911525, 33826035, 33829098, 33829107, 33829112, 33829104, 33829103, 33829096, 33951567, 30568542, 33946304, 33947263, 33898454, 33986876, 33946298, 33971837, 33955842, 32878915, 33955807, 33981800, 33981802, 33981803, 33938362, 33951412, 33889429, 33979557, 33932721, 33924698, 33924693, 33924694, 33932704, 33932720, 33932703, 33932702, 33932701, 33932699, 33932698, 33932717, 33932713, 33932705, 33932682, 33932692, 33932685, 33932680, 33932690, 33932676, 33932686, 33932677, 33932675, 33932684, 33932683, 33932697, 33932696, 33932695, 33932694, 33932693, 33929484, 33929498, 33929497, 33929496, 33929495, 33929494, 33929493, 33929492, 33929490, 33929489, 33929487, 33929501, 33929483, 33929481, 33929480, 33924690, 33924689, 33924688, 33924687, 33924686, 33929488, 33929514, 33929531, 33929530, 33929527, 33929526, 33929525, 33929524, 33929522, 33929521, 33929519, 33929499, 33929516, 33929500, 33929513, 33929512, 33929511, 33929510, 33929509, 33929507, 33929505, 33929504, 33929503, 33929517, 33924685, 33894552, 33924681, 33924664, 33924663, 33932738, 33932743, 33932739, 33932737, 33932733, 33932731, 33932729, 33932728, 33932727, 33932726, 33932725, 33932724, 33932740, 33898445, 33929532, 33929533, 33898428, 33898440, 33898447, 33898448, 33930017, 33930030, 33930028, 33930027, 33930026, 33930025, 33930024, 33930021, 33930020, 33930032, 33930018, 33930033, 33930015, 33930014, 33930011, 33930010, 33929540, 33929539, 33929538, 33929536, 33929535, 33930019, 33930031, 33930101, 33930042, 33930041, 33930040, 33930039, 33930038, 33930036, 33930034, 33894420, 33894573, 33894577, 33823351, 33881444, 33826061, 33826059, 33826057, 33819678, 33819663, 33818713, 33818717, 33818724, 33818728, 33818730, 33818734, 33911444, 33955786, 33981788, 33938927, 33950571, 33911630, 33904879, 33951416, 33951281, 33951274, 33911629, 33917451, 33997568, 33915923, 33915924, 33922355, 33915963, 33955849, 33921044, 33955749, 33992550, 33992554, 33959560, 33959561, 33959568, 33959562, 34052683, 34052645, 34052646, 34052647, 34052649, 34052652, 34052656, 34052657, 34052665, 34052671, 34052672, 34052675, 34052676, 34060188, 34054500, 34060186, 34060185, 34054534, 34054532, 34052677, 34054510, 34052679, 34054494, 34054492, 34052695, 34052692, 34052684, 34052641, 34054512, 33992561, 34052643, 33992567, 33992566, 33992565, 33992564, 33992571, 33992562, 33992572, 33992560, 33992559, 33992558, 33992557, 33992556, 33992555, 33992563, 33992553, 34060189, 34052639, 34052638, 34052637, 34052635, 33992568, 34022199, 34052642, 34022197, 34022196, 34022194, 34022193, 34022191, 34022190, 34022200, 34060187, 33992552, 33955782, 33951459, 33955779, 33955780, 33955781, 34060216, 34060193, 34060190, 33955785, 33986865, 33992549, 33992547, 33992545, 33992542, 33955784, 33955789, 33992540, 34060200, 34060196, 34060195, 34060194, 33955787, 34060209, 33955796, 33955790, 33955791, 33955793, 33981809, 34060201, 33955795, 34060205, 34060824, 34060819, 34060215, 34060214, 34060210, 33955788, 33981808, 33959573, 33959574, 33955805, 33959602, 33959603, 33955804, 33955801, 34060820, 33955830, 33959572, 33959609, 33955800, 33955798, 33938928, 33955853, 33955847, 33955846, 33894287, 33938359, 33938357, 33938356, 33955770, 33955756, 33955778, 33951441, 33894574, 33904918, 33904915, 33904877, 33905454, 33900863, 33901123, 33901772, 33901776, 33901783, 33901784, 33901788, 33901786, 33901760, 33901740, 33900860, 33898456, 33898453, 33946288, 33946285, 34052666, 33950581, 33819651, 33894593, 33894584, 33897332, 33951507, 33897352, 33897354, 33951443, 33951286, 33951290, 33980265, 33980307, 33981766, 33981767, 33981769, 33823355, 33997576, 33947868, 33930112, 33936855, 33936883, 33936881, 33936880, 33936856, 33936853, 33936851, 33933930, 33936898, 33936884, 33936885, 33936897, 33936895, 33936891, 33936888, 33936903, 33930118, 33933896, 33933924, 33933901, 33936904, 33986912, 34011818, 33986341, 33976461, 33959569, 33947937, 33951124, 33967369, 33967368, 33967361, 33967358, 33967354, 33967380, 33971927, 33971922, 33971944, 33971942, 33967389, 33967386, 33901118, 33901101, 33901097, 33955845, 33894412, 33911590, 33894590, 33898503, 33894578, 33894589, 33980277, 33951584, 33823377, 33936861, 33936859, 33936886, 33936892, 33936907, 33884885, 33884889, 33829091, 33829095, 33818755, 33955799, 33955835, 33946300, 33946321, 33947896, 33947921, 33936909, 33936912, 33829088, 33922371, 33971945, 33976465, 33976460, 33976457, 33976452, 33976459, 33976448, 33976450, 33976445, 31305464, 30553849, 33955844, 33955837, 33955836, 33955817, 33955851, 33959542, 33922296, 33955773, 30553852, 34015706, 33894401, 31421058, 33947912, 33980289, 33933900. International Serial Numbers: 33947246, 33947247, 33890408, 33892605, 33790557, 33819595, 33819594, 33894427, 33826015, 33823319, 33823313, 33946278, 33933874, 33946270, 33933948, 33933865, 33933882, 33950568, 33933892, 33933893, 33933895, 33933887, 33933912, 33933913, 33933898, 33933889, 33930119, 33933884, 33826079, 33819604, 33905620, 33823314, 33946287, 33889422, 33892602, 33884927, 33884920, 33884935, 33894575, 33894408, 33894407, 33894595, 33894576, 33905610, 33905613, 33911432, 33905605, 33950445, 33915942, 33915932, 33922354, 33915948, 33915934, 33819610, 33823315, 33946291, 33946292, 33946294, 33946295, 33976467, 33946290, 33881436, 33986796, 33986798, 33986801, 33959551, 33959548, 33976462, 33959557, 33979559, 33986803, 33976405, 33976466, 33980269, 33955843, 33986812, 33986813, 33955783, 33980292, 33976473, 33976469, 33986875, 33976438, 33976439, 33986805, 33946309, 33976442, 33986806, 33976444, 33976484, 33980284, 33976449, 33986810, 33890389, 33890402, 33890398, 33890397, 33890396, 33889497, 33890390, 33889504, 33890395, 33881434, 33890417, 33890403, 33890404, 33890409, 33890407, 33890406, 33889418, 33889432, 33889431, 33889426, 33889420, 33889472, 33889417, 33889415, 33889414, 33885001, 33884998, 33884994, 33884989, 33889421, 33889481, 33889475, 34003588, 34003564, 33992548, 33986914, 34011707, 34011704, 34003590, 34003621, 34003619, 34003611, 34003605, 34003594, 34011697, 33892597, 33892606, 33892604, 33892600, 33892598, 33892609, 33892596, 33892592, 33892591, 33892589, 33986347, 33981801, 33892607, 33892608, 33892614, 33892613, 33892612, 33892610, 33884955, 33959604, 33881413, 33881411, 33881410, 33881408, 33881399, 33884986, 33881409, 33881438, 33881418, 33881421, 33881396, 33881383, 33881441, 33884914, 33884921, 33884919, 33884918, 33881439, 33884915, 33884932, 33884913, 33884907, 33884906, 33884905, 33884916, 33884954, 33884951, 33884949, 33884947, 33884943, 33884941, 33884939, 33884933, 33884869, 33884880, 33884956, 33884900, 33884895, 33884877, 33884893, 33884891, 33884883, 33819599, 33829408, 33819612, 33829394, 33829397, 33819585, 33819587, 33794788, 33814241, 33794789, 33894570, 33930104, 33932691, 33894422, 33930111, 33897330, 33897347, 33905504, 33897351, 33894550, 33897348, 33897345, 33897339, 33897338, 33897334, 33897349, 33826095, 33826096, 33823347, 33819584, 33826091, 33829094, 33826038, 33826044, 33826045, 33826047, 33826075, 33826051, 33905606, 33905624, 33911440, 33911519, 33911517, 33911514, 33826032, 33819583, 33818771, 33818767, 33819659, 33819647, 33819653, 33819658, 33819657, 33819630, 33819616, 33819619, 33819621, 33819623, 33819625, 33819626, 33819633, 33819637, 33829113, 33829108, 33829101, 33826024, 33826030, 33819615, 33823312, 33823316, 33823318, 33823320, 33823321, 33823327, 33933940, 33946279, 33933935, 33933936, 33922361, 33898444, 33922360, 33936913, 33936908, 33894549, 33933943, 33946284, 33898433, 33933957, 33932732, 33898443, 33936905, 33924677, 33933879, 33924668, 33929478, 33922379, 33922374, 33933875, 33924670, 33922373, 33922380, 33922381, 33924667, 33933871, 33933934, 33933951, 33946275, 33932735, 33922365, 33922366, 33933933, 33933862, 33933880, 33951127, 33950578, 33951103, 33950576, 33951122, 33955777, 33950456, 33947932, 33950447, 33950452, 33950453, 33950579, 33950455, 33950457, 33950570, 33950572, 33950573, 33950575, 33950454, 33922297, 33951271, 33947892, 33955850, 33959541, 33959543, 33915946, 33915959, 33951402, 33951401, 33971930, 33971931, 33971929, 33933908, 33971915, 33971912, 33936858, 33971958, 33971957, 33971952, 33971951, 33936900, 33930126, 33947881, 33947882, 33947885, 33947887, 33947888, 33936882, 33947878, 33951591, 33951587, 33971960, 33947889, 33947870, 33936901, 33947869, 33947879, 33947871, 33947872, 33947873, 33947874, 33936890, 33986326, 33955834, 33955811, 33976481, 33955810, 33976482, 33976480, 33955808, 33976483, 33889425, 33889434, 33881406, 33881393, 33881362, 33881372, 33881369, 33881366, 33881363, 33881375, 33881361, 33881360, 33881359, 33881358, 33881357, 33881356, 33881355, 33881364, 33881374, 33881380, 33881379, 33881376, 33884909, 33881450, 33881445, 33884886, 33884884, 33884894, 33819598, 33819606, 33819608, 33829152, 33829149, 33829154, 33829153, 33829150, 33829148, 33819592, 33829146, 33829144, 33819596, 33932689, 33894390, 33894389, 33892593, 33894571, 33894386, 33932679, 33881331, 33894388, 33881354, 33881353, 33881352, 33881351, 33881350, 33881333, 33881330, 33881329, 33881327, 33830738, 33881349, 33894385, 33894384, 33894383, 33894381, 33894291, 33904919, 33826060, 33826072, 33818791, 33818742, 33818787, 33819655, 33819643, 33819667, 33819635, 33823336, 33823324, 33823331, 33823317, 33946282, 33933937, 33898446, 33898432, 33933944, 33894554, 33898435, 33898436, 33933942, 33933869, 33932741, 33933870, 33933868, 33933876, 33933872, 33933858, 33932744, 33946269, 33933863, 33933861, 33950583, 33950446, 33950449, 33950574, 33947930, 33921056, 33921048, 33921054, 33933886, 33947883, 33947886, 33936902, 33947875, 33894596, 33980270, 33980299, 33986794, 33976425, 33955823, 33959553, 33959545, 33959544, 33979505, 33991489, 33991484, 33991483, 33991478, 33991494, 33991477, 33991499, 33890401, 33890400, 33890399, 33890405, 33889503, 33889502, 33889501, 33889500, 33889499, 33890424, 33890413, 33890422, 33890421, 33890420, 33890418, 33890416, 33890414, 33890412, 33890411, 33890410, 33889496, 33890415, 33889428, 33889498, 33889486, 33889495, 33889494, 33889493, 33889492, 33889491, 33889490, 33889489, 33889487, 33889484, 33889483, 33889480, 33889479, 33889477, 33892588, 33889488, 34003578, 34003565, 33997578, 33997575, 34003593, 33890423, 34011834, 34022211, 34015723, 34011708, 34011705, 34011703, 34011696, 34003592, 34011829, 34011813, 33892603, 33892601, 33892595, 33892594, 33892590, 33892599, 33981790, 33959608, 33959607, 33959600, 33959566, 33892611, 33881407, 33881416, 33881415, 33881414, 33881412, 33881395, 33881419, 33881405, 33881400, 33881398, 33881397, 33881428, 34022212, 33881435, 33946286, 33881433, 33881432, 33881431, 33881417, 33881429, 33881427, 33881426, 33881425, 33881424, 33881423, 33881420, 33881430, 33881370, 33881368, 33881365, 33881392, 33881391, 33881390, 33881389, 33881388, 33881387, 33881386, 33881373, 33881384, 33881382, 33881381, 33881378, 33881377, 33881385, 33884917, 33884910, 33884908, 33884942, 34022198, 33884948, 33884946, 33884931, 33884938, 33884936, 33884934, 33884902, 33884944, 33884904, 33884874, 33884870, 33881455, 33881454, 33881453, 33884878, 33881451, 33881449, 33881448, 33881447, 33881446, 33881443, 33881442, 33881452, 33884892, 33881440, 33884901, 33884899, 33884897, 33884896, 33884903, 33884890, 33884888, 33884887, 33884881, 33829395, 33819600, 33819601, 33819607, 33819597, 33819613, 33829409, 33829406, 33829405, 33829404, 33829403, 33829401, 33819609, 33829399, 33829398, 33829396, 33829128, 33829402, 33829142, 33819588, 33829132, 33829131, 33829130, 33819586, 33829143, 33829410, 33814240, 33814239, 33829393, 33829392, 33819589, 33819590, 33829147, 33819591, 33819593, 33819614, 33930108, 33930115, 33932678, 33894551, 33830737, 33830736, 33829411, 33924700, 33826093, 33826077, 33826078, 33826080, 33826082, 33826083, 33823374, 33826085, 33826086, 33826089, 33826033, 33826092, 33826073, 33826094, 33829140, 33823361, 33819605, 33819603, 33819602, 33823380, 33819676, 33826039, 33829093, 33826043, 33829089, 33826049, 33826050, 33823383, 33826074, 33826058, 33826062, 33826063, 33826064, 33826065, 33826066, 33826067, 33826070, 33826071, 33817019, 33818757, 33818756, 33818748, 33818745, 33818740, 33818737, 33818725, 33818761, 33818763, 33818790, 33818789, 33818788, 33818786, 33818785, 33818784, 33818782, 33818759, 33818780, 33818775, 33818773, 33818772, 33818770, 33818769, 33818768, 33818766, 33818764, 33818781, 33819646, 33819648, 33819649, 33819650, 33829125, 33819652, 33819654, 33819656, 33826034, 33819660, 33819661, 33829124, 33819664, 33829123, 33819666, 33829122, 33829121, 33819670, 33819671, 33819617, 33819618, 33819620, 33819622, 33819624, 33819627, 33819645, 33819629, 33819631, 33819632, 33829127, 33829126, 33819638, 33819641, 33819674, 33819628, 33829120, 33829115, 33829114, 33826002, 33829110, 33826006, 33829106, 33826011, 33823334, 33826014, 33823333, 33826016, 33829102, 33826020, 33826025, 33826026, 33829097, 33826028, 33826031, 33826012, 33819675, 33819677, 33819679, 33829118, 33829117, 33823335, 33819673, 33823322, 33823325, 33823326, 33823328, 33823329, 33823330, 33823332, 33922359, 33946280, 33946283, 33933954, 33933952, 33933953, 33900686, 33933939, 33933955, 33933938, 33894553, 33933946, 33898429, 33898431, 33898449, 33933956, 33898441, 33898442, 33933941, 33924678, 33929476, 33924669, 33929477, 33922368, 33922367, 33933867, 33922370, 33922376, 33946267, 33933947, 33922382, 33922378, 33929479, 33933860, 33946273, 33922363, 33946276, 33933949, 33933950, 33922364, 33933864, 33933866, 33951115, 33950582, 33950585, 33950586, 33951064, 33951074, 33951116, 33951117, 33951119, 33951120, 33951123, 33951125, 33951084, 33951132, 33924676, 33951126, 33922375, 33924662, 33924665, 33924666, 33924671, 33924672, 33924674, 33922295, 33924680, 33924683, 33924691, 33930141, 33930100, 33930099, 33924673, 33915933, 33932722, 33947926, 33915943, 33915957, 33915960, 33933902, 33933891, 33933897, 33933903, 33930120, 33930123, 33933883, 33933885, 33947895, 33947891, 33947906, 33947911, 33947914, 33947918, 33947928, 33938358, 33938361, 33947880, 33947876, 33894387, 33979498, 33979556, 33979520, 33976422, 33933881, 33946310, 33889471, 33881371, 33884922, 33894568, 33894588, 33894581, 33894579, 33884928, 33826076, 33915902. |
Recalling Firm/ Manufacturer |
Medtronic Emergency Response Systems, Inc. 11811 Willows Rd Ne Redmond WA 98052-2003
|
For Additional Information Contact |
Todd Bandy 425-867-4514
|
Manufacturer Reason for Recall |
Therapy connectors may have connector sockets that do not meet design specs - short drill depth. Interruption of the connections may cause unexpected, intermittent or complete loss of pacing and defibrillation functions.
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
On 3/22/06, the firm began sending out letters dated March 2006, and flagged URGENT MEDICAL DEVICE CORRECTION. The letters advised customers of the issue and that a service representative would visit and inspect the therapy connector sockets, and replace the therapy connector and the mating cable if short drill sockets are found. |
Quantity in Commerce |
2678 devices |
Distribution |
Nationwide, and Canada, Netherlands, Germany, Australia, Guam, Hong Kong, New Zealand, New Caledonia, Argentina, Brazil, Chile, Costa Rica, Mexico, Uruguay, Venezuela. Domestic distribution includes hospitals, clinics, |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = LDD and Original Applicant = MEDTRONIC PHYSIO-CONTROL CORP.
|
|
|
|