| | Class 2 Device Recall CONTAK RENEWAL 3 & 4 |  |
| Date Initiated by Firm | May 15, 2006 |
| Date Posted | July 04, 2006 |
| Recall Status1 |
Terminated 3 on January 06, 2007 |
| Recall Number | Z-1192-06 |
| Recall Event ID |
35537 |
| PMA Number | P010012S008 |
| Product Classification |
Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D) - Product Code NIK
|
| Product | Guidant CONTAK RENEWAL 3 (models H170, H173, H175), CONTAK RENEWAL 3 HE (models H177, H179) CONTAK RENEWAL 4 (models H190, H195), CONTAK RENEWAL 4 HE (models H197, H199). Cardiac Resynchronization Therapy Defibrillator (CRT-D). Sterile EO. Guidant Corporation, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. Cardiac resynchronization therapy defibrillators (CRT-D) provide ventricular tachyarrhythmia and cardiac resynchronization therapies. Ventricular tachyarrhythmia therapy is for the treatment of ventricular tachycardia (VT) and ventricular fibrillation (VF), rhythms that are associated with sudden cardiac death (SCD). Cardiac resynchronization therapy is for the treatment of heart failure (HF) and uses biventricular electrical stimulation to synchronize ventricular contractions. Cardioversion/defibrillation therapies include a range of low- and high-energy shocks using either a biphasic or monophasic waveform. |
| Code Information |
Devices that are implanted subpectorally with the serial number facing the ribs. Implant orientation is not reported to Guidant. |
Recalling Firm/ Manufacturer |
Guidant Corporation 4100 Hamline Ave N Saint Paul MN 55112-5700
|
Manufacturer Reason for Recall | Guidant has received 2 reports of device malfunction associated with subpectoral implantation in an uncommon orientation (serial number facing ribs). Repetitive mechanical stress in this orientation can result in Loss of shock therapy, Loss of pacing therapy (intermittent or permanent), Loss of telemetry, Programmer warning screen upon interrogation, and beeping. |
FDA Determined Cause 2 | Other |
| Action | The recall was initiated on May 15, 2006, with a press release and a letter to physicians. Physicians are asked to review the specific positioning of each device for patients implanted with affected models. This is a review of medical records to determine if the device is subpectoral. Then use of an AP radiograph on patients with a subpectoral implant to determine device orientation and if leads exit in a clockwise direction. In those cases, it''s recommended physicians advise the patient of the potential for failure, and follow-up at least once per quarter per device labeling and consider device replacement for larger muscle/active patients or for patients who regularly utilize therapy.
Physicians are instructed to orient device with the serial number facing away from the ribs for future implants. |
| Quantity in Commerce | 54,971 devices have been distributed however this only affects those that have been implanted subpectorally with the serial number facing the ribs. |
| Distribution | Nationwide, US Virgin Islands, Puerto Rico and worldwide to include: Argentina, Aruba, Australia, Austria, Barbados,
Belgium, Bermuda, Brazil, Canada, Cayman Islands, Chile,
Colombia, Costa Rica, Cyprus, Czech Republic, Denmark,
Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guadeloupe, Guam, Hong Kong, Hungary,
India, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lebanon, Luxembourg, Malaysia, Marshall Islands, Mexico,
Netherlands, New Zealand, Northern Mariana Islands, Norway,
Peru, Polynesia, Portugal, Qatar, Russia, Saint Lucia, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Kingdom, Uruguay, Venezuela, West Indies. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| PMA Database | PMAs with Product Code = NIK
|
|
|
|