Class 2 Device Recall CONTAK RENEWAL 3 & 4 AVT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall CONTAK RENEWAL 3 & 4 AVTsee related information
Date Initiated by FirmMay 15, 2006
Date PostedJuly 04, 2006
Recall Status1 Terminated 3 on January 06, 2007
Recall NumberZ-1194-06
Recall Event ID 35537
PMA NumberP010012S008 
Product Classification Implantable Pulse Generator, Pacemaker (Non-Crt) - Product Code LWP
ProductGuidant CONTAK RENEWAL 3 AVT (models M150, M155), CONTAK RENEWAL 3 AVT HE (models M157, M159), CONTAK RENEWAL 4 AVT (models M170, M175), CONTAK RENEWAL 4 AVT HE (models M177, M179). Cardiac Resynchronization Therapy Defibrillator. Sterilized using gaseous ethylene oxide. Guidant Corporation Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798, USA. CONTAK RENEWAK AVT models provide both atrial and ventricular tachyarrhythmia and cardiac resynchronization therapies.
Code Information Devices that are implanted subpectorally with the serial number facing the ribs. Implant orientation is not reported to Guidant.
Recalling Firm/
Manufacturer
Guidant Corporation
4100 Hamline Ave N
Saint Paul MN 55112-5700
Manufacturer Reason
for Recall
Guidant has received 2 reports of device malfunction associated with subpectoral implantation in an uncommon orientation (serial number facing ribs). Repetitive mechanical stress in this orientation can result in Loss of shock therapy, Loss of pacing therapy (intermittent or permanent), Loss of telemetry, Programmer warning screen upon interrogation, and beeping.
FDA Determined
Cause 2
Other
ActionThe recall was initiated on May 15, 2006, with a press release and a letter to physicians. Physicians are asked to review the specific positioning of each device for patients implanted with affected models. This is a review of medical records to determine if the device is subpectoral. Then use of an AP radiograph on patients with a subpectoral implant to determine device orientation and if leads exit in a clockwise direction. In those cases, it''s recommended physicians advise the patient of the potential for failure, and follow-up at least once per quarter per device labeling and consider device replacement for larger muscle/active patients or for patients who regularly utilize therapy. Physicians are instructed to orient device with the serial number facing away from the ribs for future implants.
Quantity in Commerce54,971 devices have been distributed however this only affects those that have been implanted subpectorally with the serial number facing the ribs.
DistributionNationwide, US Virgin Islands, Puerto Rico and worldwide to include: Argentina, Aruba, Australia, Austria, Barbados, Belgium, Bermuda, Brazil, Canada, Cayman Islands, Chile, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guadeloupe, Guam, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lebanon, Luxembourg, Malaysia, Marshall Islands, Mexico, Netherlands, New Zealand, Northern Mariana Islands, Norway, Peru, Polynesia, Portugal, Qatar, Russia, Saint Lucia, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Kingdom, Uruguay, Venezuela, West Indies.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
PMA DatabasePMAs with Product Code = LWP
-
-