| | Class 3 Device Recall D711, Reference Membranes, REF 942058 |  |
| Date Initiated by Firm | June 17, 2005 |
| Date Posted | August 02, 2006 |
| Recall Status1 |
Terminated 3 on September 14, 2011 |
| Recall Number | Z-1319-06 |
| Recall Event ID |
35665 |
| 510(K)Number | K041874 |
| Product Classification |
in vitro diagnostic - Product Code CHL
|
| Product | D711, Reference Membranes, Part #942-058, used with the ABL835 Blood Gas Analyzers. |
| Code Information |
Lot Code: R0308, Expiration Date: 4/07; Lot Code: R0309, Expiration Date: 5/07; and Lot Code: R0310, Expiration Date: 5/07. |
Recalling Firm/ Manufacturer |
Radiometer America Inc 810 Sharon Dr Westlake OH 44145-1598
|
| For Additional Information Contact | Vince Sigmund 800-736-0600 |
Manufacturer Reason for Recall | The recalled membranes cause the status value of the pH and electrolyte electrodes to be offset causing readings outside of the acceptable range. |
FDA Determined Cause 2 | Other |
| Action | The recalling firm notified their distributor/customers of this recall, via telephone on 6/17/2005, at which time, the customers were asked to destroy any and all suspect reference membranes found on-site and at their subaccount customer locations. |
| Quantity in Commerce | 13 boxes/4 membranes per box |
| Distribution | Nationwide, including TX, OK, MI, CO, NY, and Canada. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = CHL
|
|
|
|