| | Class 2 Device Recall Steris Ultrasonic instrument cleaner |  |
| Date Initiated by Firm | June 22, 2006 |
| Date Posted | August 25, 2006 |
| Recall Status1 |
Terminated 3 on January 08, 2007 |
| Recall Number | Z-1407-06 |
| Recall Event ID |
35737 |
| Product Classification |
Ultrasonic instrument cleaner - Product Code FLG
|
| Product | Steris Amsco Sonic Energy Cleaner, Part Nos.: P080000005, B018905020, B018905JBJ, and B018905JBJSS |
| Code Information |
All units bearing serial numbers 0432000018 to 0403105051. |
Recalling Firm/ Manufacturer |
Steris Corporation 2720 Gunter Park Dr E Montgomery AL 36109-1410
|
| For Additional Information Contact | Kenneth P. Finch 334-274-5183 |
Manufacturer Reason for Recall | Smoke, sparking and fire hazard-Units manufactured between November 2000 and January 2005 are susceptible to water damage at the lid switch cable wiring harness, which may result in smoking, sparking, and fire. |
FDA Determined Cause 2 | Other |
| Action | The firm sent out a certified letter to all affected consignees alerting them to problems with the device on 6/22/06. |
| Quantity in Commerce | 322 units |
| Distribution | Worldwide USA and countries of Japan, Canada, Hong Kong, Korea, Lebanon, Qatar, Poland, Russia, Saudi Arabia, Singapore, Taiwan, Germany, Northern Mariana Islands, and Malaysia. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|