| Class 2 Device Recall Fugo Blade Incising Tip Assembly | |
Date Initiated by Firm | April 28, 2006 |
Date Posted | August 02, 2006 |
Recall Status1 |
Terminated 3 on September 08, 2006 |
Recall Number | Z-1224-06 |
Recall Event ID |
35738 |
510(K)Number | K001498 |
Product Classification |
ophthalmic electrocautery - Product Code NCR
|
Product | Fugo Blade Incising Tip Assembly. Incising tips to be used with the Fugo Blade M1000 Anterior Capsulotomy Unit. For Ophthalmic Use Only. The product is shipped 12 units per carton. |
Code Information |
Lot numbers 01013786, 01017917, 01017918, 01017684 and 01017442 |
Recalling Firm/ Manufacturer |
Medisurg Research & Management Corp 100 W Fornance Street The Fugo Building Norristown PA 19401
|
For Additional Information Contact | Dawn DelCampo 610-277-3937 |
Manufacturer Reason for Recall | No documentation to support shelf life/sterility |
FDA Determined Cause 2 | Other |
Action | On 4/28/06, the recalling firm faxed a notification to the customers informing them not to use the product during surgery and to return the product. |
Quantity in Commerce | 960 tips |
Distribution | Nationwide, including CA, CT, FL, GA, MA, NC, NY, OH, OR, PA, TX, and WI. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = NCR
|
|
|
|