|
Class 2 Device Recall ambIT |
|
Date Initiated by Firm |
October 10, 2006 |
Date Posted |
November 16, 2006 |
Recall Status1 |
Terminated 3 on May 26, 2011 |
Recall Number |
Z-0188-2007 |
Recall Event ID |
36505 |
510(K)Number |
K002434 K033325 K052221
|
Product Classification |
Ambulatory infusion pump cassette. - Product Code FRN
|
Product |
ambIT Ambulatory Infusion Pump Cassette; Filterless, Male, Sterile, Vol. 2.3ml, REF 220267, Sorenson Medical, Inc., West Jordan, Utah. |
Code Information |
REF 220267, all serial numbers |
Recalling Firm/ Manufacturer |
Sorenson Medical, Inc. 1375 W 8040 S West Jordan UT 84088-8320
|
For Additional Information Contact |
801-352-6267
|
Manufacturer Reason for Recall |
Sterile disposable cassette used with infusion pump may cause an under-infusion when used in a low flow rate therapy of less than 10 ml per hour.
|
FDA Determined Cause 2 |
Other |
Action |
Consignees were notified by letter on 10/10-11/2006 and told to immediately discontinue use/quarantine all affected cassettes. . |
Quantity in Commerce |
4956 units |
Distribution |
Worldwide, including USA, Belgium, France, Germany, Italy, Singapore, Venezuela. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = FRN and Original Applicant = SORENSON MEDICAL, INC.
|
|
|
|