Class 2 Device Recall VerifyNow P2Y12 Assay Device Kit

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall VerifyNow P2Y12 Assay Device Kitsee related information
Date Initiated by FirmSeptember 05, 2006
Date PostedDecember 06, 2006
Recall Status1 Terminated 3 on March 26, 2012
Recall NumberZ-0228-2007
Recall Event ID 36707
510(K)NumberK051231 
Product Classification Platelet assay, automated - Product Code JOZ
ProductVerifyNow P2Y12 Assay Device Kit, Part Number 85054
Code Information Lot Numbers: WK0002A , WK0003A, WK0003B, WK0004A ,   WK0005A , WK0006A, WK0006AR, WK0006B, WK0006BR,   WK0007A, WK0008A, WK0009A, WK0010A ,WK0010B, 10WK000C, WK0011A, WK0012A and WK0013A
FEI Number 3001376288
Recalling Firm/
Manufacturer
Accumetrics Inc
3985 B Sorrento Valley Blvd
San Diego CA 92121
For Additional Information ContactRoberta G. Hoffman
888-919-9333 Ext. 218
Manufacturer Reason
for Recall
The VerifyNow P2Y12 assay can report an erroneous result instead of an error message when a sample with low hematocrit is run.
FDA Determined
Cause 2
Other
ActionThe 333 consignees were notified by regular mail or by E-mail on 9/5/2006 -- 9/13/2006. There were four versions of the notification letter tailored to (1) domestic customers; (2) domestic distributors; (3) international customers; and (4) international distributors. Each notification was accompanied by a 'fax back' form which customers were asked to sign and return to Accumetrics as an acknowledgement they received the notification letter. The notification letter instructs consignees that test results with the VerifyNow P2Y12 Assay and all VerifyNow assays should be interpreted in conjunction with other clinical and laboratory data available to the clinician. Caution is advised when testing patients outside the ranges documented in the package insert.
Quantity in Commerce333
DistributionWorldwide Distribution- USA, and countries of Australia, Canada, France, Germany, Italy, Japan, Korea, Malaysia, Netherlands, New Zealand, Norway, Poland, Singapore, Spain, Sweden, Switzerland, & United Kingdom
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = JOZ
-
-