| Class 2 Device Recall Multi Diagnost Eleva | |
Date Initiated by Firm | July 23, 2007 |
Date Posted | September 25, 2007 |
Recall Status1 |
Terminated 3 on September 17, 2012 |
Recall Number | Z-0912-2007 |
Recall Event ID |
44910 |
510(K)Number | K023441 |
Product Classification |
X-Ray System - Product Code IZI
|
Product | Philips Multi Diagnost Eleva |
Code Information |
Site numbers 103056, 104136, 521113 102666, 104083, 104530, 105193, 105194, 105394, 504340, 504341, 505039, 505345, 505346, 506155, 506236, 506247, 519574, 532087, 532140, 532151, 532213, 532703, 533312, 533473, 533559, 534247, 535714, 535767, 536019, 536091, 536146, 536720, 536935, 537083, 537669, 537787, 538440, 539457, 539598, 539990, 540252, 541263, 541474 |
Recalling Firm/ Manufacturer |
Philips Medical Systems North America Co. Phillips 22100 Bothell Everett Hwy Bothell WA 98021-8431
|
For Additional Information Contact | Sarah Baxter 425-487-7665 |
Manufacturer Reason for Recall | Units do not comply with the requirement to provide automatic beam limitation within +/- 3 degrees of horizontal, vertical, and perpendicular to the plane of the image receptor during radiographic procedures. |
FDA Determined Cause 2 | Other |
Action | On 7/23/07 Philips mailed a letter dated July 9, 2007 titled Electronic Product Radiation Warning. The letter advises customers of the issue and asks that customers apply the enclosed cautionary label to the device. The letter also explains that there will be a system modification to be installed. On 8/20/07 the firm sent the same letter, along with a letter dated 8/20/07 to customers thought to have either scrapped their units and the customer had previously indicated they were no longer using the unit. The 8/20/07 letter requested the customer contact Philps if they are currently using the device. |
Quantity in Commerce | 44 units |
Distribution | Units were distributed to 1543 hospitals and medical centers throughout the US. No international distribution from this firm. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = IZI
|
|
|
|