Date Initiated by Firm |
October 04, 2007 |
Date Posted |
November 27, 2007 |
Recall Status1 |
Terminated 3 on July 22, 2009 |
Recall Number |
Z-0290-2008 |
Recall Event ID |
45462 |
Product Classification |
Femoral Holder/Driver - Product Code HWR
|
Product |
Odyssey Femoral Holder/Driver, Part Number K001-7410, Non-sterile, 1 each, Wright Medical Technology, Inc., 5677 Airline Road, Arlington, TN 38002 |
Code Information |
Lot Numbers: 047367193, 067380510, and 077414021 |
Recalling Firm/ Manufacturer |
Wright Medical Technology Inc 5677 Airline Road Arlington TN 38002-0100
|
For Additional Information Contact |
Debbie Daurer 901-867-4601
|
Manufacturer Reason for Recall |
Breakage: breaking of the connecting foot that is on the distal surface of the instrument
|
FDA Determined Cause 2 |
Device Design |
Action |
The firm notified their distributors of the recall on October 4, 2007, with an Urgent Medical Device Recall letter explaining the problem and requesting return of the product from their inventory and from their consignees. The hospitals were notified of the recall on October 4, 2007 using the same Urgent Device Recall Letter. |
Quantity in Commerce |
66 units |
Distribution |
Worldwide-USA and Australia |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|