Class 2 Device Recall Mevatron MD2 |
|
Date Initiated by Firm |
August 07, 2007 |
Date Posted |
February 14, 2008 |
Recall Status1 |
Terminated 3 on December 01, 2010 |
Recall Number |
Z-0678-2008 |
Recall Event ID |
46157 |
510(K)Number |
K882729
|
Product Classification |
Linear Medical Accelerator - Product Code IYE
|
Product |
Siemens Mevatron MD2 Medical Charged-Particle Radiation Therapy System; (Interlock #51 Jaw Field Position) Model Number: 09401654, Siemens Medical Solutions, Concord, CA 94520 |
Code Information |
Serial Numbers: 1949, 1991, 1996, 2045, 2053, 2067, 2113, 2117, 2192, 2199, 2235, 2238, 2241, 2252, 2316, 2374, 2163, 2205, 2217, 2249, 2283, 2305, 2346, 2397, 2423, 2427, 2362, 2392, 2425, 2426, 2086, 2118, 2139, 2152, 2189, 2231, 2234, 2261, 2281, 2328, and 2438 |
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc 4040 Nelson Ave Concord CA 94520-1200
|
For Additional Information Contact |
925-246-8407
|
Manufacturer Reason for Recall |
Incorrect field size -- Products manufactured prior to 1999 may experience jaw field size errors during interlock resulting in potential mistreatment if a light field check is not performed. This problem is present in machines with serial numbers less than 3094, as they do not contain redundant sensors for jaw position.
|
FDA Determined Cause 2 |
Device Design |
Action |
The firm has sent consignees a Customer Safety Advisory Notice on August 7, 2007 informing them that a misplacement of the X and/or Y jaws can occur if interlock #51 Jaw Field Position is asserted and then cleared with the left-hand key rest. The firm recommends that if an interlock issue is encountered tha treatment of the patient be discontined until this issue is resolved. |
Quantity in Commerce |
41 units |
Distribution |
Worldwide Distribution |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = IYE and Original Applicant = SIEMENS MEDICAL LABORATORIES, INC.
|
|
|
|