| Date Initiated by Firm | February 12, 2008 |
| Date Posted | March 28, 2008 |
| Recall Status1 |
Terminated 3 on June 26, 2009 |
| Recall Number | Z-1361-2008 |
| Recall Event ID |
46706 |
| Product Classification |
Template for Clinical Use - Product Code HWT
|
| Product | Biomet Vanguard Width Checker, Part Number: 32-487062, 67.5/75 Sizing Wing, Biomet Orthopedics, Inc., Warsaw, IN; The device is a preparatory instrument used for femoral knee insertion. |
| Code Information |
Lot Number-212910. |
Recalling Firm/ Manufacturer |
Biomet, Inc. 56 E Bell Dr Warsaw IN 46581-0587
|
| For Additional Information Contact | 800-348-9500 |
Manufacturer Reason for Recall | The instrument's sizing line is in the wrong place. |
FDA Determined Cause 2 | Process control |
| Action | Distributors were notified via Urgent Medical Device Removal Notice letter dated 2/12/08 to locate the instruments in their possession and at hospital sites and to return them to Biomet. Hospitals with the instruments on hand will be provided with a recall letter dated 2/12/08 which describes the nature of the problem and the removal of these instruments. The Puerto Rico consignee was notified by Biomet UK, Ltd. via letter on 2/26/08. |
| Quantity in Commerce | 35 (31 to U.S. and 4 international). |
| Distribution | Worldwide: USA, Canada, China and Finland. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|