• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Internal handles and electrodes for LIFEPAK 9, LIFEPAK 9P, LIFEPAK 12, AND LIFEPAK 20 Defibrillator/

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Internal handles and electrodes for LIFEPAK 9, LIFEPAK 9P, LIFEPAK 12, AND LIFEPAK 20 Defibrillator/see related information
Date Initiated by FirmJanuary 29, 2008
Date PostedSeptember 16, 2008
Recall Status1 Terminated 3 on May 04, 2011
Recall NumberZ-2036-2008
Recall Event ID 46749
Product Classification Low-energy DC-Defibrillator - Product Code LDD
ProductInternal Defibrillation Electrodes: PN: 805355-10 thru 14 and 805355-020 thru -024.
Code Information Part Number 805355 - Lot Numbers: 0001, 0060, 0080, 0090, 0100, 0102, 0104, 0106, 0107, 0123, 0197, 0200, 0203, 0204, 0205, 0206, 0207, 0300, 0301, 0304, 0305, 0306, 0307, 0398, 0399, 0404, 0405, 0407, 0499, 0500, 0501, 0502, 0503, 0504, 0505, 0507, 0600, 0601, 0604, 0605, 0606, 0607, 0698, 0700, 0703, 0704, 0705, 0706, 0707, 0800, 0802, 0803, 0804, 0805, 0806, 0807, 0895, 0899, 0900, 0901, 0903, 0904, 0905, 0906, 0907, 1000, 1001, 1002, 1004, 1005, 1006, 1007, 1023, 1097, 1099, 1101, 1103, 1104, 1105, 1106, 1107, 1201, 1202, 1204, 1205, 1206, 1207, 1298, 1300, 1302, 1304, 1305, 1306, 1307, 1398, 1400, 1402, 1403, 1404, 1405, 1406, 1407, 1498, 1499, 1501, 1502, 1505, 1506, 1601, 1602, 1603, 1604, 1605, 1606, 1607, 1700, 1701, 1702, 1703, 1704, 1705, 1706, 1707, 1797, 1799, 1801, 1802, 1804, 1805, 1806, 1807, 1898, 1899, 1903, 1905, 1906, 1907, 1999, 2000, 2001, 2002, 2003, 2004, 2005, 2006, 2101, 2103, 2104, 2105, 2106, 2107, 2202, 2203, 2205, 2206, 2207, 2298, 2299, 2300, 2301, 2302, 2304, 2305, 2306, 2399, 2402, 2403, 2405, 2406, 2407, 2499, 2500, 2501, 2502, 2503, 2505, 2506, 2507, 2599, 2600, 2601, 2602, 2603, 2605, 2606, 2607, 2698, 2699, 2702, 2800, 2802, 2806, 2807, 2899, 2900, 2901, 2903, 2904, 2905, 2906, 2907, 2999, 3000, 3001, 3002, 3005, 3006, 3007, 3098, 3099, 3100, 3101, 3102, 3104, 3105, 3106, 3107, 3197, 3198, 3203, 3204, 3205, 3206, 03294, 3299, 3302, 3303, 3305, 3306, 3398, 3399, 3400, 3401, 3403, 3404, 3405, 3406, 3407, 3498, 3499, 3500, 3501, 3503, 3505, 3507, 3597, 3599, 3602, 3603, 3605, 3606, 3607, 3697, 3698, 3699, 3700, 3701, 3702, 3705, 3706, 3707, 3798, 3799, 3805, 3806, 3807, 3899, 3900, 3901, 3905, 3906, 3907, 3998, 4000, 4003, 4005, 4006, 4007, 4099, 4100, 4101, 4102, 4103, 4105, 4106, 4107, 4197, 4199, 4200, 4201, 4203, 4204, 4206, 4207, 4298, 4299, 4300, 4302, 4304, 4305, 4306, 4307, 4399, 4401, 4405, 4406, 4498, 4499, 4500, 4502, 4503, 4506, 4507, 4600, 4601, 4603, 4605, 4606, 4607, 4699, 4701, 4702, 4704, 4705, 4706, 4797, 4799, 4803, 4805, 4806, 4900, 4904, 4906, 4907, 4998, 5000, 5001, 5002, 5003, 5005, 5006, 5103, 5105, 5106, 5199, 5200, 5203, 5204, 5205, 5206, 5299, 5301.
FEI Number 3015876
Recalling Firm/
Manufacturer
Physio Control, Inc.
11811 Willows Rd Ne
Redmond WA 98052-2003
For Additional Information Contact
800-442-1142
Manufacturer Reason
for Recall
Based on testing, sterilization methods described in labeling may be ineffective and may cause damage or corrosion to the paddles or handles. If handles and/or paddles are not sterilized adequately and are used on a patient, there is a risk of infection. If handle discharge button is damaged during sterilization, defibrillation therapy may be prevented.
FDA Determined
Cause 2
Labeling design
ActionStarting 2/5/08, Physio-Control sent letters to customers which notifies them not to use Flash Gravity Steam, 12/88 Ethylene Oxide, STERRAD 100 machine, or Gravity Steam to sterilize. The firm recommends the use of 100% EO, STERRAD 100S machine, or Prevacuum steam for sterilization. The letter also recommends that after sterilization, the handles and paddles should be examined carefully for damage. If damaged, the product should be removed from use. The firm will follow up with a revised Sterilization Guideline document and requests posting the letter in the sterilization department in the meantime. Contact technical support at 800-442-1142 if you have questions. On 7/23/08 the firm issued new sterilization guidelines that supersede all others.
DistributionDistributed to hospitals and medical centers. Nationwide, DOD, and international: ARGENTINA, BAHRAIN, BANGLADESH, BERMUDA, BOLIVIA, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, EGYPT, EL SALVADOR, FRANCE, GERMANY, GREECE, GUATEMALA, GUYANA, HONG KONG, INDIA, IRAN, ISRAEL, JAPAN, KOREA, LEBANON, MALAYSIA, MALTA, MEXICO, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, QATAR, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SRI LANKA, SYRIA, THAILAND, TURKEY, UNITED ARAB EMIRATES, URUGUAY, VENEZUELA, VIRGIN ISLANDS - US, YEMEN, ZIMBABWE
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
-
-