Class 2 Device Recall Baxa ExactaMix 600 Pharmacy compounding system |
|
Date Initiated by Firm |
February 11, 2008 |
Date Posted |
August 19, 2008 |
Recall Status1 |
Terminated 3 on March 16, 2012 |
Recall Number |
Z-1371-2008 |
Recall Event ID |
46771 |
Product Classification |
Automated pharmacy compounding system - Product Code NEP
|
Product |
Exacta-Mix 600 Pharmacy Compounding System is used in the pharmacy to compound multiple source ingredients into on final solution based on user-specified sequences and volumes. |
Code Information |
Model 071B-RS232, all serial numbers |
Recalling Firm/ Manufacturer |
Baxa Corporation 14445 Grasslands Dr Englewood CO 80112-7062
|
For Additional Information Contact |
303-617-2242
|
Manufacturer Reason for Recall |
Flush solution from TPN compounder may be added to patient's TPN bag.
|
FDA Determined Cause 2 |
Software design |
Action |
Consignees were notified by letter (Exacta-Mix Compounder Software Safety Alert) on 02/11/2008 advising them of the necessity to install a waste bag prior to a machine flush. |
Quantity in Commerce |
221 units |
Distribution |
Nationwide, including military facilities in CA, FL, HI, MS, NC, TX, VA and VA facilities in NY, OR, SC, UT. Foreign distribution to Australia, Canada, China and New Zealand. |