| Class 2 Device Recall Terumo "Custom" Cardiovascular Procedure Kit | |
Date Initiated by Firm | January 29, 2008 |
Date Posted | March 28, 2008 |
Recall Status1 |
Terminated 3 on July 07, 2009 |
Recall Number | Z-1313-2008 |
Recall Event ID |
46869 |
510(K)Number | K982891 |
Product Classification |
Cardiovascular Procedure Kit - Product Code OEZ
|
Product | Cardiovascular Procedure Kit, P/N 70579-04, St. Louis University Hospital Pump & Table Pack, Terumo Cardiovascular, Ashland, MA. (the Pump & Table pack contains a weighted flexible sucker, Catalog No. 804113) |
Code Information |
Lot Numbers: GC20, GE24, GF22, GG12, GH17, GL25, GN06, HA22, HD19, HF07, HG18 and HM08. |
Recalling Firm/ Manufacturer |
Terumo Cardiovascular Systems Corp 6200 Jackson Rd Ann Arbor MI 48103-9586
|
For Additional Information Contact | 800-521-2818 |
Manufacturer Reason for Recall | Loose Flash at tip of Weighted Flexible Sucker: A loose strand of stainless steel left on the device during manufacturing by an outside supplier may result in an unretrieved metal fragment remaining in the patient. |
FDA Determined Cause 2 | Nonconforming Material/Component |
Action | Consignees were notified via recall letter, Urgent Medical Device Removal, dated 1/29/08 via fax or email to cease using the weighted sucker in the packs and to use an alternative sucker until replacement product arrives from Terumo. |
Quantity in Commerce | 252 |
Distribution | Nationwide-USA including states of Missouri, Tennessee and Wisconsin-Custom Kits for 3 hospitals. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|