| Date Initiated by Firm | March 28, 2008 |
| Date Posted | May 09, 2008 |
| Recall Status1 |
Terminated 3 on June 25, 2012 |
| Recall Number | Z-1601-2008 |
| Recall Event ID |
47436 |
| Product Classification |
Heparin Access Flush - Product Code NZW
|
| Product | Tyco Healthcare Monoject Prefill100U/mL Heparin
Lock Flush Syringe 5mL
REF # 8881590125, Tyco Healthcare, Mansfield, MA 02048 |
| Code Information |
Lot Numbers: 7041194; 7102804; 7041204; 7113034; 7051534; 7113044; 7051544; 7113054; 7051554 ; 7113104; 7071924 ; 7113114; 7072034; 7113154; 7072044; 8010064; 7072054; 8010114; 7072064 ; 8010134; 7072154; 8010174; 7082284; |
| FEI Number |
1282497
|
Recalling Firm/ Manufacturer |
Covidien LP 15 Hampshire St Mansfield MA 02048-1113
|
| For Additional Information Contact | SAME 508-261-8000 |
Manufacturer Reason for Recall | Scientific Protein Laboratories (SPL), disclosed that two lots of Heparin Sodium, USP Active Pharmaceutical Ingredient has a heparin-like contaminant |
FDA Determined Cause 2 | Material/Component Contamination |
| Action | Tyco Healthcar notifed customers by letter sent via US Postal Service beginning 3/28/08, through 4/1/08. The letter requests that they immediately stop using and return the product. Distributors are requested to in-turn notify their customers. An instruction sheet was attached for product return and replacement.
All customers are required to respond to the notice regardless of remaining inventory.
. |
| Quantity in Commerce | 7,959,224 units |
| Distribution | Nationwide Distribution |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|