| Date Initiated by Firm | May 21, 2008 |
| Date Posted | September 16, 2008 |
| Recall Status1 |
Terminated 3 on April 30, 2012 |
| Recall Number | Z-1870-2008 |
| Recall Event ID |
48447 |
| Product Classification |
Mechanical Operating-Room Table - Product Code FWX
|
| Product | Medacta AMIS Universal Table for Leg Positioner, Model Number: 01.15.10.0172. The AMIS universal table for leg positioner and AMIS Leg Positioner are used during Anterior Minimally Invasive Surgery for hip arthroplasty. The system aids the surgical procedure by providing support and alignment of the limb. |
| Code Information |
All product of Model Number 01.15.10.0172 are subjectto this safety notification |
| FEI Number |
3006639916
|
Recalling Firm/ Manufacturer |
Medacta Usa Inc
4725 Calle Quetzal
Camarillo
CA
930129101
|
Manufacturer Reason for Recall | Medacta is issuing this Recall because improper connection of the table clamping mechanism may result in the AMIS table becoming separated from the surgical table resulting in injury to the patient. |
FDA Determined Cause 2 | Labeling design |
| Action | The Notification was sent via FedEx overnight on 05/27/2008 to the Director of Risk Management, Director of Biomedical, and Director of Operating Room at each institution with a table.
The Safety Notification included:
(1) Revised Instructions for Use that include (a) Improved instructions for the attachment of the clamps, and
(b) Instructions for the use of new safety straps to provide a fail-safe attachment means.
(2) A hang tag for labeling the devices.
(3) A reply card indicating that the label has been attached to the table.
Customers have been instructed in the proper adjustment of the clamps for the universal table as well as in the use of the safety straps. Customers have also been provided with a hang tag that reinforces the use of the safety straps when attaching the universal table.
An additional copy of the notification will be provided to the sales representatives responsible for these hospitals. Included in these packets are instructions for the sales representatives to visit each hospital and provide hands-on training to the operating room personnel regarding proper attachment of the universal table, use of the clamps, and use of the straps.
If you have questions, contact Medacta USA at 1-800-901-7836. |
| Quantity in Commerce | 20 Units |
| Distribution | Nationwide to CA, CO, FL, ID, NV, TX & UT; Worldwide to: Belgium, France, Germany, Italy & Switzerland |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|