| Class 2 Device Recall Naviscan PET Systems Inc ("Naviscan") PEM Flex Solo II PET Scanner. | |
Date Initiated by Firm | June 19, 2008 |
Date Posted | September 11, 2008 |
Recall Status1 |
Terminated 3 on March 13, 2009 |
Recall Number | Z-2229-2008 |
Recall Event ID |
48694 |
510(K)Number | K032063 |
Product Classification |
PET Scanner - Product Code KPS
|
Product | Naviscan PET Systems, Inc ("Naviscan") PEM Flex Solo II PET Scanner. The PEM Flex Solo II images small body parts with optimized PET technology. |
Code Information |
Serial Numbers: P002-B-070316-01, P002-B-070423-01, P002-B-070531-01, P002-B-070629-01, P002-B-070712-0l, P002-B-070718-01, P002-B-070910-01, P002-B-070911-01, P002-B-071015-01, P002-B-080107-01, P002-B-080108-01, P002-B-080207-01, and P002-B-080307-01 |
Recalling Firm/ Manufacturer |
Naviscan PET Systems 11180 Roselle St Ste 100 San Diego CA 92121-1211
|
For Additional Information Contact | Heather Jalisi 858-332-0942 |
Manufacturer Reason for Recall | This field correction is being initiated as a result of the firm's investigation into a report from a user indicating that the motorized compression exceeded 25 pounds of compression force during the pre-scan positioning of the patient. The investigation identified that a secondary fault protection did not perform as expected. |
FDA Determined Cause 2 | Device Design |
Action | An Urgent Medical Device Field Correction Notice was sent to all PEM Flex Solo II customers about this issue via Federal Express Priority Service on June 19, 2008.
Customers were informed of the:
(1) Issue: Potential for motorized compression to exceed 25 pounds of compression force if continued beyond the primary safety limit, This could lead, to patient discomfort or injury.
(2) Advisory: Naviscan recommends that patient breast positioning be completed by ceasing motorized-compression when the primary safety limit of 15 pounds pre-compression force is reached. This is indicated by the automatic stopping of the paddle movement as controlled by the primary safety system. "Fine-tuning" positioning, should be completed by using the manual adjustment knobs.
(3) Correction: A Naviscan Field Service Engineer will contact each customer to schedule a field correction visit. If customers have any questions they were instructed to contact Naviscan Customer Support at 1-888-628-4159. |
Quantity in Commerce | 13 |
Distribution | Nationwide to CA, FL, MA, MO, NY, OK, PA, TX, and WA |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = KPS
|
|
|
|