• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall One Touch Ping Glucose Monitoring System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall One Touch Ping Glucose Monitoring Systemsee related information
Date Initiated by FirmAugust 04, 2008
Date PostedNovember 07, 2008
Recall Status1 Terminated 3 on November 12, 2008
Recall NumberZ-0283-2009
Recall Event ID 49475
510(K)NumberK080639 
Product Classification Insulin Infusion Pump - Product Code LZG
ProductOne Touch Ping Glucose Monitoring System Combines the functionality of OneTouch(R) Ping (TM) Insulin Pump and a OneTouch(R) Ping (TM) Meter Remote through radio frequency (RF) communication. Together, they provide an option to help make insulin delivery more discreet and flexible. Also, allows the most recent blood glucose results from the OneTouch(R) Ping (TM) Meter Remote to be automatically entered into bolus insulin calculations to cover carbohydrates in food or to correct for a high blood glucose level.
Code Information Part numbers: 100-430-00, 100-431-00, 100-432-00, 100-434-00, and 100-435-00. Serial numbers: 01-02000-15, 02-02001-15, 03-02002-15, 04-02003-15, 04-02010-15, 05-02004-15, 05-02011-15, 06-02012-15, 06-02100-15, 07-02013-15, 07-02101-15, 08-02007-15, 08-02102-15, 09-02008-15, 09-02110-15, 10-02009-15, 10-02104-15, 10-02111-15, 11-02112-15, 12-02032-15, 12-02106-15, 12-02113-15, 12-02120-15, 13-02033-15, 13-02040-15, 13-02107-15, 13-02121-15, 13-02202-15, 14-02027-15, 14-02034-15, 14-02108-15, 14-02203-15, 14-02210-15, 15-02028-15, 15-02035-15, 15-02042-15, 15-02123-15, 15-02204-15, 16-02029-15, 16-02036-15, 16-02050-15, 16-02117-15, 16-02124-15, 16-02131-15, 16-02205-15, 16-02300-15, 17-02037-15, 17-02051-15, 17-02118-15, 17-02132-15, 17-02213-15, 17-02220-15, 18-02038-15, 18-02126-15, 18-02140-15, 18-02221-15, 19-02039-15, 19-02053-15, 19-02060-15, 19-02127-15, 19-02208-15, 19-02222-15, 20-02047-15, 20-02054-15, 20-02142-15, 20-02209-15, 20-02216-15, 20-02223-15, 20-02304-15, 21-02048-15, 21-02129-15, 21-02136-15, 21-02143-15, 21-02150-15, 21-02217-15, 21-02224-15, 21-02231-15, 22-02049-15, 22-02056-15, 22-02063-15, 22-02144-15, 22-02151-15, 22-02225-15, 22-02232-15, 23-02057-15, 23-02071-15, 23-02152-15, 23-02219-15, 23-02233-15, 23-02240-15, 24-02058-15, 24-02065-15, 24-02072-15, 24-02139-15, 24-02160-15, 24-02241-15, 25-02059-15, 25-02066-15, 25-02080-15, 25-02147-15, 25-02228-15, 25-02235-15, 26-02148-15, 26-02155-15, 26-02162-15, 26-02236-15, 26-02243-15, 26-02250-15, 26-02331-15, 27-02149-15, 27-02156-15, 27-02163-15, 27-02237-15, 27-02244-15,  28-02076-15, 28-02090-15, 28-02157-15, 28-02164-15, 28-02171-15, 28-02245-15, 28-02319-15, 28-02333-15, 29-02091-15, 29-02158-15, 29-02172-15, 29-02246-15, 29-02253-15, 29-02260-15, 30-02092-15, 30-02166-15,  30-02173-15, 30-02180-15, 30-02247-15, 30-02254-15, 30-02261-15, 30-02342-15, 31-02093-15, 31-02167-15, 31-02248-15, 31-02255-15, 31-02262-15, 31-02336-15, 32-02094-15, 32-02168-15, 32-02175-15, 32-02256-15, 32-02263-15, 32-02351-15, 33-02088-15, 33-02169-15, 33-02176-15, 33-02190-15, 33-02345-15, 34-02096-15, 34-02177-15, 34-02184-15, 34-02191-15, 34-02258-15, 35-02178-15, 35-02185-15, 35-02192-15, 35-02259-15, 36-02098-15, 36-02186-15, 36-02193-15, 36-02348-15, 37-02099-15, 37-02195-15, 39-02189-15, 39-02196-15, 40-01901-15, 40-02197-15, 41-02198-15, 42-01910-15, 42-02199-15, 43-01911-15, 44-01912-15, 45-01906-15, 45-01913-15, 45-01920-15, 48-01909-15, 49-01924-15, 49-01931-15, 50-01918-15, 50-01925-15, 51-01926-15, 51-01940-15, 52-01927-15, 52-01934-15, 52-01941-15, 53-01928-15, 53-01942-15, 54-01936-15, 54-01943-15, 54-01950-15, 55-01951-15, 56-01864-15, 56-01938-15, 56-01952-15, 57-01939-15, 57-01946-15, 57-01953-15, 57-01960-15, 58-01866-15, 58-01873-15, 58-01947-15, 58-01961-15, 59-01874-15, 59-01948-15, 59-01948-15, 60-01868-15, 60-01882-15, 60-01949-15, 60-01963-15, 60-01970-15, 61-01869-15, 61-01876-15, 61-01883-15, 61-01890-15, 61-01957-15, 62-01877-15, 62-01884-15, 62-01891-15, 62-01958-15, 63-01885-15, 63-01892-15, 63-01959-15, 63-01973-15, 64-01886-15, 64-01893-15, 64-01974-15, 64-01981-15, 65-01887-15, 67-01889-15, 67-01896-15, 67-01991-15, 68-01897-15, 68-01985-15, 68-01992-15, 69-01898-15, 69-01979-15, 69-01993-15, 70-01899-15, 70-01987-15, 70-01994-15, 71-01995-15, 72-01996-15, 73-01997-15, 74-01998-15, and 75-01999-15.
Recalling Firm/
Manufacturer
Animas Corporation
200 Lawrence Dr
West Chester PA 19380
For Additional Information ContactRita McIntyre
610-644-8990
Manufacturer Reason
for Recall
display old, inaccurate values for bolus amount delivered and amount of planned bolus totals
FDA Determined
Cause 2
Software design
ActionOn 8/6/08, the recalling firm telephoned all users to inform them of problem and the need to return the system.
Quantity in Commerce251 units
DistributionNationwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = LZG
-
-