|
Class 2 Device Recall DatexOhmeda S/5 Anesthesia Delivery Unit |
|
Date Initiated by Firm |
October 30, 2008 |
Date Posted |
July 30, 2009 |
Recall Status1 |
Terminated 3 on May 17, 2012 |
Recall Number |
Z-1635-2009 |
Recall Event ID |
50800 |
510(K)Number |
K050676 K042154 K973985
|
Product Classification |
Gas-Machine for Anesthesia - Product Code BSZ
|
Product |
Datex-Ohmeda S/5 Anesthesia Delivery Unit
Intended to provide general inhalation anaesthesia and ventilatory support to a wide range of patients. Intended for volume or pressure control, pressure support and synchronized intermittent mandatory (SIMV) ventilation modes. |
Code Information |
6505591 6504143 6504144 6505588 6488527 6482851 6478332 6478333 6478334 6482385 6478325 6474359 6472276 6472278 6472279 6474357 6472280 6465283 6465282 6472271 6453235 6453236 6438719 6438718 6438723 6438723 6424397 6438712 6438712 6433525 6433526 6433527 6424397 6423152 6424394 6423156 6423157 6423158 6424393 6423153 6423154 6419336 6419337 6419338 6414823 6403878 6403880 6403883 6408473 6408474 6408113 6403881 6399280 6399281 6395497 6393520 6395487 6395489 6393514 6393515 6382828 6382829 6382830 6382831 6382833 6390074 6390075 6390436 6390437 6390438 6391069 6391070 6391071 6391072 6391073 6391074 6380586 6380587 6380582 6374511 6374512 6374513 6370003 6369999 6368798 6368052 6368053 6365585 6367921 6363659 6363651 6355792 6355793 6355794 6351433 6351434 6351435 6351436 6351438 6351439 6353705 6353706 6353707 6353709 6355786 6355787 6353701 6353702 6353703 6351440 6351441 6351442 6353695 6353704 6353708 6358409 6358408 6358407 6358406 6358405 6353697 6353696 6351437 6241024 6241028 6241032 6246176 6222219* 6241030 6241027 6236404 6241026 6215604** 6215609** 6215607** 6241021 6246177 6241031 6241039 6241029 6236405 6241020 6236407 6215621 6215607 6215609 6215604 6190700 6190701 6190702 6190704 6125604 6125605 6190692 6190693 6190694 6190695 6190698 6190699 6125601 6190676 6125603 6125606 6125607 6182956 6190619 6190621 6182958 6182944 6182945 6182946 6125595 6125596 6125516* 6125519* 6125590 6125591 6125592 6125593 6125594 6125584 6125587 6125576 6125579 6125580 6125577 6125546 6125548 6125527 6125526 6125516 6125519 6125504 6125505 6125506 6125507 6125508 6125509 6125510 6125511 6125502 6125627 6125499 6125500 6125628 6125629 6125630 6125471 6125472 6125473 6125474 6125475 6125476 6125483 6125484 6125485 6125486 6125497 6125498 6125631 6125459 6125443 6097443 6097444 6097445 6097452 6097455 6097457 6097458 6097463 6097464 6097465 6097466 6097467 6097468 6097469 6097470 6097474 6097476 6097481 6097454 6097387 6067703 6085924 6097375 6097376 6097377 6097378 6097379 6097380 6097381 6097382 6097383 6097384 6097426 6097427 6097428 6097429 6097430 6097431 6097432 6097433 6097434 6097435 6097436 6097437 6097438 6097439 6097440 6097441 6097442 6097409 6097410 6097411 6097412 6097413 6097414 6097415 6097416 6097417 6097418 6097419 6097420 6097421 6097422 6097423 6097424 6097425 6097393 6097401 6097403 6097404 6097405 6097406 6097407 6097408 6097504 6097394 6097395 6097391 6097392 6085915 6085916 6085918 6085925 6086605 6097358 6097360 6097362 6097363 6097371 6097365 6097366 6097368 6097369 6097370 6086599 6086607 6086590 6086604 6086592 6067669 6067671 6067679 6067708 6067670 6067693 6067641 6067680 6067681 6067682 6067715 6067648 6067649 6067656 6067660 6067678 6067723 6067662 6067710 6067645 6067646 6067647 6067642 6067655 6067658 6067639 6067640 6067650 6067719 6067722 6067638 6067634 6067636 6067627 6067628 6067726 6048809 6048836 6048837 6067724 6048824 6048825 6048826 6048811 6048813 6048817 6048818 6048820 6048821 6048822 6048841 6048842 6048831 6048834 6048835 6048810 6048823 6048830 6048814 6048812 6048815 6048819 6048832 6048783 6048784 6048785 6048792 6048796 6048799 6048802 6048803 6048804 6048786 6048788 6048789 6048790 6048791 6048793 6048794 6048798 6048797 6048800 6048807 6048808 6048806 6048766 6048805 6048801 6048771 6048846 6048795 6035761 6048855 6048856 6048760 6048761 6035762 6035773 6048853 6048854 6035759 40300736 6035765 6035774 6035784 6048756 6048757 6048758 6035775 6035760 6035772 6020230 40300726 40300727 40300728 40300730 40300738 40300739 40300740 40300741 40300742 40300743 6035751 6035757 6035758 6025124 6025128 6035737 6035740 6035741 6035742 6035743 6035744 6035745 6035747 6025108 6025110 6025111 6025125 6025126 6025127 6025131 6025132 6025137 6035738 6025109 6025114 6025115 6025116 6025117 6025119 6025120 6025122 6025123 6025130 6025092 6025106 6025107 6025133 6025134 6025136 6017506 6020222 6025138 6025097 6025099 6025103 6025104 6025105 6025102 6025096 6025100 6025101 6424400 6424401 6295272 6295271 6295270 6295269 6291479 6321383 6321389 6358404 6358403 6358402 6358401 6355800 6355799 6355797 6355796 6333584 6393511 6438715 6424395 6424396 6449787 6125573 6125572 6125571 6125570 6125569 6125568 6125567 6125566 6017406 6017411 6067661 6067659 6097386 6097385 6125589 6190691 6246184 6048778 6048777 6048776 6182948 6321388 6321386 6474360 6459102 6459100 6086594 6086593 6085937 6085936 6085935 6085934 6085932 6085931 6085930 6085928 6125520 6265709 6125614 6399277 6399278 6399279 6125565 6048774 6067709 6097367 6067677 6438721 6420594 6403882 6393516 6391786 6391785 6391784 6391783 6355798 6341969 6295278 6295275 6295274 6295273 6253232 6125462 6097361 6085938 6419341 6420589 6420590 6420591 6420592 6449782 6280792 6280795 6280796 6280797 6280798 6280799 6085914 6085926 6048772 6048773 6449781 6449783 6449786 6459095 6459096 6086586 6125461 6125460 6125585 6125578 6182954 6182953 6182952 6182951 6182950 6182949 6236403 6233478 6233471 6233470 6233469 6233467 6265706 6265704 6262565 6262564 6262561 6262560 6020221 6017503 6017501 6017497 6017496 6017495 6017494 6017493 6017432 6017431 6017430 6017428 6017416 6017419 6017421 6017427 6017429 6017498 6017500 6020218 6474365 6474366 6449785 6423155 6419342 6395488 6395490 6358413 6363374 6363650 6358412 6329449 6329450 6228755 6215611 6190620 6182947 6097498 6086596 6086595 6086591 6086577 6086576 6067637 6035780 6025089 6017508 6291478 6463643 6465271 6454603 6459093 6433530 6420593 6393512 6391075 6380585 6355788 6125436 6097460 6097462 6086613 6086584 6035755 6017414 6017418 6017412 6017413 6017415 6013323 6017403 6017404 6486760 6472272 6472273 6472274 6454594 6438724 6438725 6438726 6419343 6420597 6420598 6307164 6307161 6271286 6228753 6228752 6225495 6225494 6215620 6215601 6215600 6215599 6190663 6190662 6190658 6190641 6190640 6190639 6125442 6125441 6125437 6125435 6125434 6097461 6097459 6097456 6097453 6048833 6048827 6035752 6025093 6013077 6035756 6048762 6125518 6125588 6190673 6190674 6215616 6215618 6215619 6215622 6259822 6312312 6312313 6312314 6321382 40000892 40013604 40019528 40022745 40027753 40032309 40059668 40070388 40114916 6438731 6449780 6222208 6215633 6215631 6067692 6067691 6067690 6067688 6067687 6502373 6490251 6465272 6465273 6449788 6438722 6438713 6433532 6433524 6424398 6423149 6367923 6367922 6353700 6341962 6341961 6353698 6341968 6341964 6321378 6321377 6280788 6271287 6253229 6236397 6236396 6233476 6233475 6233472 6190644 6190645 6190643 6125597 6125575 6125564 6125563 6125562 6125561 6125626 6125514 6125447 6125446 6125445 6097374 6067663 6086574 6067672 6067665 6067720 6048775 6048768 6048851 6048850 6048852 6035754 6035753 6035763 6020223 6020229 6017407 6125480 6125479 6262559 6265708 6265707 6365583 40000567 40000901 40000920 40000891 6013076 6013075 6013074 6035783 6035782 6035781 6048782 6097483 6097482 6097480 6097479 6097478 6097477 6097475 6097473 6097472 6097471 6125448 6190642 6190638 6190657 6190656 6190654 6190653 6190652 6190651 6190650 6190649 6190648 6190655 6228773 6236402 6236401 6246182 6246181 6246180 6321380 6321379 6393517 6393518 6393519 6414822 6035739 6025121 6025113 6025112 6035750 6035749 6035748 6035746 6067725 6067630 6067629 6086581 6086585 6086583 6085927 6097373 6097372 6125481 6125559 6125557 6125581 6182937 6182939 6182938 6215603 6215602 6222225 6222220 6228778 6307162 4011377 6202490 40059756 40149175 40300725 6504142 6438729 6438727 6395496 6395495 6395494 6380583 6368056 6365573 6365572 6365571 6363653 6363652 6341965 6321387 6321387 6312310 6291477 6291473 6262557 6262558 6257639 6222222 6222218 6222217 6222216 6222215 6222211 6215627 6182943 6182941 6215613 6215612 6215615 6215614 6182942 6190647 6125545 6125522 6125470 6125469 6125468 6125467 6125466 6125465 6125464 6125463 6125433 6097493 6097492 6097490 6097488 6097487 6097486 6097485 6097484 6097447 6097446 6085917 6097357 6086611 6086588 6085933 6086612 6086610 6067706 6067705 6067702 6067699 6067698 6026725 6025129 6025095 6025094 6020233 6020232 6020228 6020227 40004336 6505590 6488526 6408478 6408479 6374510 6304985 6304984 6280787 6262563 6262562 6257638 6222210 6182940 6125513 6125512 6125477 6125444 6097390 6086575 6067631 40113228 40011275 40067322 40010626 40010629 40000302 40000552 40000548 6017510 6017511 6020224 6020225 6020226 6048779 6048780 6048781 6086600 6086601 6086602 6086603 6097355 6097356 6097364 6190668 6190669 6190670 6190671 6190672 6228774 6329451 6329452 6329454 6048844 6048770 6438728 6438730 6265705 6067714 6067711 6067676 6067675 6067668 6067667 6067666 6086609 6086598 6086597 6502375 6454602 6438716 6438717 6438720 6433528 6433529 6423150 6423151 6410767 6410768 6382825 6382826 6380588 6382824 6365581 6365582 6368054 6368055 6265711 6265712 6269507 6269508 6269509 6271283 6271284 6241033 6241034 6253226 6253227 6253228 6241035 6241037 6241038 6246178 6246183 6190679* 6190684* 6190685* 6190705* 6190707 6222207 6233466 6182962 6190681* 6190682* 6190683* 6182961 6182960 6182959 6182936 6125612 6125611 6125610 6125609 6125608 6125544 6125543 6125542 6125541 6125540 6125539 6125538 6125537 6125536 6125535 6125534 6125533 6125532 6125531 6125530 6125529 6125525 6125524 6125523 6125521 6125496 6125495 6125494 6125493 6125492 6125491 6125490 6125489 6125488 6125487 6125458 6125457 6125456 6125455 6125454 6125453 6125452 6125451 6125450 6125449 6097503 6097502 6097501 6097500 6097499 6097497 6097496 6097495 6097494 6097491 6097489 6067717 6067716 6067713 6067704 6067674 6067673 6048845 6048843 6048767 6048765 6048763 6025091 6025090 6020235 6020234 6020231 6465281 6465284 6291470 6291469 6291468 6291467 6291466 6291465 6291463 6291462 6291461 6291460 6291459 6291458 6291457 6291456 6291455 6291454 6291453 6291452 6291451 6291450 6290461 6290459 6290458 6290457 6290456 6290455 6290454 6290453 6285899 6285898 6285897 6285896 6285895 6285894 6285893 6285892 6285891 6285890 6285887 6285886 6285885 6280801 6280800 6280791 6280790 6280789 6228772 6228771 6086608 6086589 6067686 6067685 6067684 6472275 6438714 6424399 6419334 6419335 6414824 6414825 6391076 6391778 6391779 6391780 6391781 6391782 6391787 6382832 6380580 6380581 6380579 6374514 6376698 6376699 6376700 6376701 6376702 6370001 6365577 6365578 6333585 6321381 6312315 6291475 6257640 6215610 6190689 6190627 6190626 6190625 6190624 6190623 6190622 6125503 6125501 6017509 6017507 6097388 6097389 40080040 6465276 6465277 6463644 6465270 6459099 6474361 6478324 6496354 6502374 6496288 6472277 6490249 6490250 6465279 6465280 6465278 6474363 6474364 6496355 6459102 6474360 6459102 6474360 6459100 6474358 6493414 6504140 6505589 6504141 6509346 6496356 6496357 6496358 6478331 6496359 6478323 6465274 6465275 40000558 40000559 6496350 6496351 6496352 6496353 109909 810380 6365584 40022353 40031904 40049771 40070277 40070278 40070279 40070280 40070281 40070282 40070283 40070308 40070341 40070342 40070348 40070349 40070350 40070369 40070371 40070375 40070376 40070377 40070379 40070381 40070384 40070385 40070387 40071103 40074258 40074280 40075049 40075050 40075051 40075052 40075055 40075060 40075061 40075062 40075063 40075065 40075067 40075076 40075081 40075087 40075089 40075267 40078677 40078678 40078679 40078683 40078686 40078687 40078689 40078690 40078700 40078712 40078714 40078719 40078720 40078723 40078725 40078727 40078729 40078731 40078736 40078737 40078738 40078739 40078742 40078743 40078744 40078749 40078751 40078763 40078765 40078766 40078767 40078768 40078769 40078771 40078772 40084506 40084510 40084517 40084518 40084519 40084522 40084525 40084526 40084527 40085652 40085679 40085680 40085682 40085733 40088575 40088581 40088582 40088583 40088584 40089331 40089341 40089371 40089372 40089373 40089377 40089382 40089383 40089384 40089386 40089399 40089400 40089401 40089403 40089404 40089405 40089406 40089407 40091299 40092360 40092361 40092362 40092363 40092364 40092366 40092367 40094150 40094151 40094157 40094763 40094764 40095434 40095765 40095766 40095767 40095768 40095769 40095770 40095771 40095772 40095773 40095774 40095775 40095776 40095777 40095778 40095779 40095781 40095782 40095783 40095784 40095785 40095786 40095787 40095788 40095789 40095790 40095791 40095792 40095793 40095795 40095796 40095797 40095798 40095799 40095800 40095801 40095802 40095803 40095804 40095805 40095806 40095807 40095808 40095809 40095810 40095811 40095812 40095813 40095814 40097273 40097274 40097275 40097276 40097277 40097278 40097279 40097280 40097281 40097282 40097283 40097284 40097285 40097286 40097287 40097288 40097289 40097290 40097291 40097299 40097300 40097301 40097302 40097303 40097304 40097305 40097306 40097307 40097308 40097309 40097310 40097311 40097312 40097313 40098745 40098746 40098748 40098749 40100481 40100482 40100483 40100502 40100503 40102522 40103814 40103815 40103816 40105918 40105919 40105920 40105921 40105922 40105924 40105925 40105926 40105929 40105930 40105931 40107233 40107969 40107970 40107971 40107972 40107973 40107974 40107975 40108820 40108822 40108823 40108824 40108825 40108826 40112688 40112689 40113205 40113206 40113207 40113208 40114418 40114419 40114425 40114426 40114427 40114428 40114429 40114430 40114431 40116024 40116025 40116026 40116027 40116028 40116029 40116030 40116031 40116032 40116033 40116034 40116035 40116036 40116037 40116038 40120942 40120943 40120944 40120945 40120946 40120947 40120948 40120949 40120950 40120951 40120952 40120953 40120954 40120955 40120956 40120957 40120958 40120959 40120960 40120965 40120966 40120967 40120969 40120970 40120971 40120973 40120974 40120975 40120976 40122635 40124796 40127149 40127150 40127626 40127627 40127628 40129959 40129960 40129961 40129962 40129963 40225827 40225820 40225819 40225808 40225804 40225807 40225806 40224018 40224019 40224011 40224014 40224009 40224006 40224008 40224016 40224007 40224010 40224013 40224012 40224017 40224015 40223999 40224000 40224001 40224002 40223046 40223044 40223045 40223043 40223047 40221728 40221721 40221722 40221723 40221717 40202547 40202548 40202549 40202552 40213987 40209850 40209849 40209858 40209856 40209855 40209854 40209857 40209841 40209847 40209840 40209843 40209826 40209844 40209839 40209845 40209820 40209819 40202544 40160811 40160807 40209831 40209762 40202492 40202541 40202546 40202488 40202543 40209764 40202498 40202493 40202538 40202490 40209823 40209846 40209842 40209822 40202475 40202479 40202507 40202483 40202471 40202478 40202491 40202500 40202481 40202485 40202489 40202484 40202487 40202480 40202506 40202474 40202476 40202482 40202486 40202499 40202504 40202477 40202473 40202472 6182955 40202528 40202529 40202517 40202524 40202516 40202522 40202523 40202521 40202519 40202520 40202518 40202539 40202542 40202545 40202540 40202556 40202551 40202550 40202515 40202555 40202554 40202553 40202534 40202533 40202532 40202537 40202535 40202536 40202531 40202530 40202513 40178146 40178153 40178157 40178151 40178162 40178159 40178161 40178154 40178160 40178149 40178165 40178155 40178158 40178156 40178148 40178164 40178152 40178150 40178142 40178143 40178147 40178145 40178134 40178135 40178129 40178141 40178136 40178123 40178122 40178127 40178125 40178128 40178124 40178126 40178132 40178130 40178133 40178131 40178120 40178117 40178118 40178116 40178119 40178121 40178112 40178110 40178113 40178108 40178109 40178114 40178107 40178111 40178115 40178105 40178103 40178104 40178106 40178097 40178095 40178096 40178093 40178094 40178098 40178092 40178101 40178100 40178099 40178091 40178090 40178089 40178042 40178043 40178059 40178030 40178033 40178026 40178024 40178021 40178027 40178032 40178028 40178025 40178019 40178023 40178031 40178022 40178020 40178046 40178051 40178048 40178052 40178049 40178045 40178053 40178057 40178050 40178055 40178054 40178056 40178047 40178058 40178039 40178034 40178035 40178029 40178040 40178037 40178038 40178041 40178036 40178044 40178001 40178002 40177998 40178003 40178000 40178004 40178005 40177999 40177997 40178006 40177989 40177994 40177996 40177986 40177993 40177995 40177988 40177991 40177992 40177990 40177987 40177985 40177968 40177957 40177958 40177959 40177961 40177960 40177970 40177971 40177969 40177966 40177967 40177983 40177984 40177981 40177980 40177982 40177962 40177963 40177964 40177965 40177955 40177956 40177954 40177952 40177953 40177951 40177949 40177950 40177926 40177925 40177923 40177924 40177927 40177915 40177903 40177917 40177920 40177902 40177919 40177905 40177916 40177904 40177901 40177921 40177922 40177899 40177900 40177918 40177906 40177907 40177908 40177909 40177910 40177911 40177897 40177896 40177898 40177891 40177893 40177872 40177873 40177875 40177870 40177874 40177871 40177829 40177853 40177852 40177857 40177859 40177858 40177854 40177837 40177838 40177856 40177855 40174154 40174128 40174123 40174092 40174119 40174124 40174131 40174160 40174115 40174158 40174088 40174122 40174127 40174120 40174091 40174161 40174087 40174125 40174163 40174085 40170259 40174152 40174130 40174090 40174147 40174129 40174116 40174162 40174153 40174089 40174149 40174155 40174148 40174118 40174121 40174157 40177786 40177787 40174126 40174151 40174150 40174109 40174107 40174106 40174105 40174104 40174103 40174102 40170340 40174156 40170341 40170323 40174159 40174086 40174117 40177791 40177792 40177790 40177789 40177788 40177781 40174066 40174065 40174064 40174063 40177793 40177785 40177782 40177778 40174068 40177777 40177784 40177776 40177779 40177780 40174067 40177775 40177783 40170328 40174114 40174111 40170339 40174110 40174112 40174108 40174113 40174101 40170330 40170331 40170332 40170337 40170333 40170336 40170334 40170335 40170338 40167400 40167402 40167405 40167404 40167401 40167406 40167403 40167399 40167398 40174146 40174145 40174144 40174140 40174137 40174133 40174136 40174141 40174142 40174139 40174134 40174135 40174132 40174143 40174138 40167384 40167383 40167373 40167379 40167377 40167378 40167375 40167376 40167382 40167380 40167381 40167374 40167372 40167371 40167365 40167363 40167366 40167364 40167367 40167368 40167370 40167369 40167354 40165686 40165685 40165696 40165687 40164551 40165690 40165684 40165697 40165681 40165688 40165694 40165682 40165683 40165691 40165695 40165689 40167359 40167351 40167352 40167353 40167350 40167348 40167349 40167355 40167356 40167357 40167358 40165727 40165714 40165713 40165718 40165704 40165716 40165709 40165717 40165706 40165708 40165712 40165707 40165711 40165710 40165705 40165719 40165726 40165725 40165715 40164545 40164542 40164549 40164548 40164550 40164544 40164547 40164546 40164541 40164543 40164476 40164506 40164504 40164473 40164474 40164507 40164508 40164475 40164505 40164509 40164518 40164520 40164519 40164515 40164512 40164511 40164510 40164516 40164517 40164514 40164513 40163534 40162383 40162382 40163533 40163532 40164462 40164466 40164467 40163561 40163545 40164465 40163558 40164464 40164463 40163540 40162338 40162362 40162353 40162344 40162356 40162354 40162343 40162355 40162363 40162364 40162361 40162345 40162365 40162379 40162375 40162372 40162333 40162347 40162371 40162374 40162376 40162373 40162377 40162378 40162380 40162335 40162332 40162342 40162336 40162334 40162331 40162320 40162312 40162316 40162315 40162314 40162321 40162322 40162313 40162306 40162308 40162319 40162318 40162317 40162329 40162323 40162327 40162325 40162326 40162328 40162330 40162324 40162310 40162309 40162311 40160837 40160836 40160835 40160840 40160844 40162305 40162304 40162299 40162301 40162302 40162303 40162307 40162291 40160832 40160843 40160838 40160839 40160841 40160842 40160812 40160805 40160804 40159395 40160753 40160754 40160813 40160814 40160815 40160802 40160801 40160800 40160822 40160825 40160828 40160826 40160827 40160824 40160823 40160833 40160821 40160819 40160817 40160818 40160820 40160748 40160745 40160747 40160746 40159415 40159406 40159407 40159403 40159410 40159411 40159414 40159399 40159412 40159416 40159402 40159408 40159404 40159413 40159405 40159409 40159392 40159367 40159377 40159368 40159371 40159375 40159373 40159376 40159398 40159400 40159401 40159397 40159394 40159396 40159380 40159372 40159379 40159381 40159378 40159370 40159366 40159365 40159374 40159363 40159362 40159364 40159369 40159358 40159359 40159360 40159361 40157579 40157580 40157581 40157577 40157575 40157576 40157578 40157547 40157546 40157548 40157545 40157549 40157556 40157544 40157530 40157528 40157529 40157527 40157518 40157517 40157512 40157519 40157520 40157512 40157525 40157521 40157523 40157524 40157522 40157526 40157513 40157515 40157516 40157514 40157511 40157509 40157510 40155811 40155814 40155815 40155810 40155813 40155817 40155812 40155809 40155818 40155821 40155822 40155823 40155819 40155820 40155824 40155816 40155773 40155777 40155774 40155775 40155776 40155778 40155779 40155783 40155782 40155780 40155781 40155750 40155752 40155759 40155766 40155755 40155754 40155757 40155758 40155753 40155751 40155737 40155738 40155739 40154443 40154441 40154438 40154431 40154430 40154432 40154439 40154435 40154434 40154436 40154440 40154437 40154433 40154444 40154442 40154394 40154393 40154399 40154397 40154398 40154396 40154400 40154395 40154364 40154358 40154356 40154357 40154355 40154348 40154360 40154361 40154347 40154354 40154350 40152917 40152928 40152924 40152925 40152922 40152921 40152923 40152926 40152920 40152883 40152910 40152916 40152918 40152885 40152915 40152919 40152913 40152884 40152886 40152881 40152907 40152911 40152914 40152908 40152882 40152906 40152876 40152880 40152873 40152909 40152869 40152905 40152877 40152879 40152868 40152870 40152878 40152872 40152874 40152875 40152871 40150873 40150870 40150869 40150872 40150871 40150875 40150876 40150878 40150874 40150879 40150864 40150865 40150868 40150867 40150866 40150857 40150860 40150861 40150863 40150859 40150858 40150862 40150854 40150850 40150852 40150855 40150853 40150851 40150856 40150804 40150803 40150830 40150832 40150831 40150820 40150813 40150824 40150811 40150819 40150817 40150814 40150825 40150809 40150816 40150807 40150823 40150810 40150812 40150805 40150806 40150818 40150815 40150822 40150821 40150808 40150802 40150826 40150828 40150829 40150827 40150799 40149157 40149151 40149156 40149154 40149155 40149144 40149180 40149159 40149161 40149160 40149158 40149163 40149162 40149128 40149127 40149131 40149130 40148583 40148580 40148581 40148585 40148586 40149113 40149111 40148587 40149109 40149112 40149110 40148582 40148584 40148554 40148556 40148555 40145354 40145356 40145355 40145358 40145357 40145337 40145339 40145338 40145326 40145325 40145328 40145333 40145334 40145332 40145329 40145331 40145327 40145330 40143441 40143443 40143442 40145304 40145307 40145299 40145302 40145310 40143438 40143424 40143423 40143437 40143418 40143416 40143421 40143419 40143422 40143436 40143417 40143415 40143439 40143420 40142769 40142737 40142735 40142736 40142738 40141308 40141313 40141310 40141311 40141316 40141312 40141315 40141307 40141314 40141309 40139542 40139433 40139434 40139437 40139432 40139436 40136748 40139435 40139438 40139431 40136746 40136754 40136749 40136753 40136751 40136752 40139430 40136741 40136740 40136738 40136755 40136759 40136743 40136742 40136745 40136744 40136763 40136764 40136756 40136760 40136750 40136747 40136734 40136731 40136739 40136735 40136737 40136732 40136730 40136703 40136757 40136762 40136761 40136758 40136715 40136736 40136716 40136721 40136726 40136723 40136720 40136722 40136728 40136717 40136719 40136702 40136718 40136709 40136727 40136706 40136724 40136733 40136701 40136729 40136710 40136708 40136707 40136712 40136711 40136714 40136704 40136696 40136697 40136694 40136698 40136688 40136700 40136695 40136686 40136677 40136685 40136699 40136705 40136713 40136679 40136691 40136683 40136675 40136684 40136681 40136674 40136682 40136680 40136678 40136676 40136666 40136673 40136669 40136662 40136668 40136660 40136670 40136663 40136671 40136672 40136661 40136667 40136659 40136644 40136647 40136651 40136650 40136652 40136645 40136646 40136643 40136638 40136640 40136648 40136639 40136642 40136637 40136641 40136649 40136625 40136590 40136613 40136612 40136614 40136611 40132167 40136596 40132166 40132226 40132225 40132266 40132228 40132227 40132200 40132237 40132230 40132220 40132219 40132218 40132267 40132198 40132234 40132216 40132261 40132208 40132207 40132206 40132205 40132212 40132263 40132262 40132204 40132193 40132203 40132192 40132183 40132217 40132265 40132271 40132256 40132195 40132252 40132268 40132158 40132248 40132269 40132251 40132194 40132201 40132250 40132249 40132202 40132270 40132157 40132156 40132159 40132213 40132128 40132130 40132129 40132145 40132125 40132126 40132141 40132127 40132138 40132142 40132143 40132144 40132122 40132121 40132124 40132123 40132093 40132085 40129937 40129942 40129936 40129928 40129927 40129931 40129933 40129932 40129904 40129907 40129908 40129894 40129902 40129893 40129890 40129895 40129891 40129901 40129892 40129896 40129900 40129899 40129903 40129905 40129898 40129897 40129906 40127667 40127655 40127653 40127669 40127657 40127685 40127666 40127668 40127656 40127684 40127686 40127687 40127689 40129877 40129871 40129872 40127688 40129876 40129875 40129873 40129874 40127663 40127664 40127652 40127658 40127659 40127636 40127635 40127634 40127637 40127638 40127632 40127633 40127641 40127644 40127603 40127599 40127592 40127604 40127607 40127606 40127610 40127605 40127609 40127598 40127608 40127600 40127602 40127597 40127596 40127594 40127157 40127159 40127158 40127593 40127160 40127601 40127595 40127590 40127591 40127156 40127152 40127155 40127153 40127154 40127151 40124862 40124866 40124861 40124795 40124837 40124860 40124858 40124855 40124863 40122659 40124835 40124850 40124843 40124856 40124845 40124846 40124844 40124847 40122652 40122654 40122655 40122653 40122650 40122651 40122643 40122647 40122646 40122638 40122642 40122645 40122639 40122641 40122640 40122644 40122648 40122649 40122636 40122629 40122627 40122628 40122622 40122618 40122620 40122588 40122590 40122591 40122592 40122593 40122616 40122181 40122182 40122185 40122191 40122186 40122189 40122190 40122292 40122187 40122193 40122201 40120193 40122200 40122204 40122202 40122203 40120963 40120941 40120964 40120968 40120927 40120933 40120934 40120938 40120939 40120930 40120932 40120940 40120935 40120929 40120936 40120928 40120931 40120937 40120920 40120914 40120924 40120925 40120919 40120918 40120921 40120909 40120908 40120915 40120910 40120917 40120911 40120912 40120913 40120916 40120185 40120903 40120180 40120905 40120179 40120906 40120184 40120176 40120177 40120181 40120178 40120182 40120170 40120174 40120172 40120175 40120183 40120173 40119868 40119864 40120171 40119872 40119867 40119871 40119869 40119870 40119873 40120169 40119844 40118125 40118115 40118114 40118116 40118117 40117242 40117241 40117243 40117244 40117238 40117237 40117239 40117240 40116021 40116022 40116020 40116023 40116008 40116014 40116015 40116016 40116017 40116018 40116019 40115991 40115990 40115996 40115993 40115994 40115995 40115992 40115998 40115997 40115999 40116000 40116001 40116002 40116003 40116004 40116006 40116005 40116009 40116007 40116010 40116012 40116011 40116013 40114922 40114923 40114926 40114925 40114928 40114927 40114931 40114932 40114933 40114924 40113771 40113770 40113768 40113767 40113772 40113773 40113774 40113755 40113747 40113748 40113761 40113737 40113749 40113762 40111804 40111808 40111809 40112661 40112657 40111793 40111791 40111792 40111794 40112660 40112658 40111807 40112659 40111802 40111781 40111783 40111795 40110700 40110703 40110702 40111776 40111775 40111777 40111778 40111779 40111780 40111782 40110685 40110673 40110686 40110687 40110689 40110688 40110656 40110657 40110655 40110654 40110671 40110658 40110659 40110660 40110661 40110663 40110662 40110664 40110668 40110667 40110684 40110669 40110670 40110672 40109542 40109538 40109539 40109540 40110665 40110666 40109531 40109534 40109532 40109546 40109535 40109544 40109536 40109550 40109537 40109541 40109543 40108832 40108833 40108835 40108837 40108841 40108842 40108844 40108845 40108843 40108834 40108836 40108838 40108839 40108840 40108846 40108850 40108848 40108849 40108853 40109523 40109526 40109524 40109525 40108821 40108817 40107993 40107994 40107990 40107992 40107991 40107995 40107996 40107998 40107983 40107984 40107981 40107982 40107985 40107977 40107976 40107978 40107979 40107980 40106582 40106581 40106583 40106584 40106585 40107959 40107963 40107964 40106578 40106577 40106579 40106580 40106573 40106574 40106575 40106576 40106568 40106570 40106569 40106572 40106571 40106552 40106542 40106543 40106545 40106544 40106546 40106547 40106549 40106551 40105944 40105952 40105932 40105953 40105954 40105933 40105945 40105946 40104739 40104740 40104738 40104744 40104741 40104747 40104742 40104743 40104745 40104746 40104734 40104728 40104733 40104735 40104737 40104736 40104726 40103810 40103817 40103820 40103821 40103795 40103791 40103809 40103794 40103796 40103792 40103790 40102542 40102543 40100508 40100510 40100505 40100506 40098754 40100507 40100509 40098750 40098752 40098753 40098751 40098736 40098738 40098733 40098729 40098735 40098732 40098737 40098708 40098730 40098731 40098727 40098726 40098734 40098711 40098716 40098725 40098728 40097319 40098718 40097314 40097316 40097317 40097318 40098717 40097321 40097315 40098714 40098719 40097292 40097293 40097294 40097295 40097298 40097297 40097296 40095430 40095428 40095405 40095424 40095427 40095431 40095429 40095432 40095433 40095413 40095426 40095425 40095421 40095423 40095420 40095417 40095412 40095415 40095411 40095422 40095418 40095416 40095419 40095409 40095401 40095406 40095408 40095402 40095410 40095400 40095407 40095403 40095404 40095414 40094143 40094146 40094147 40094144 40094148 40094139 40094140 40092372 40094149 40092370 40092369 40094141 40094142 40092368 40092371 40094145 40092290 40092291 40092286 40092287 40089420 40089421 40089398 40089402 40089363 40089345 40089359 40089348 40089361 40089362 40089344 40089343 40089347 40089340 40089360 40089350 40089349 40089346 40089334 40089342 40089336 40089339 40089326 40088586 40089327 40089337 40089328 40089335 40089338 40088585 40089324 40089333 40088574 40088550 40088547 40089325 40088552 40088546 40088551 40088544 40088545 40088553 40088548 40088538 40085736 40085735 40082989 40085734 40085693 40085691 40085705 40085723 40085694 40085697 40085707 40085692 40085703 40085700 40085710 40085711 40085712 40085704 40085702 40085706 40085689 40085696 40085685 40085684 40085686 40085687 40085698 40085690 40085681 40085683 40085688 40085695 40085656 40085666 40085638 40084524 40084523 40085637 40084537 40085644 40085643 40085642 40084536 40084535 40084538 40084509 40084508 40084507 40084505 40084512 40082981 40082979 40082986 40082980 40082975 40082988 40082987 40083017 40083005 40083008 40083018 40083015 40083020 40083019 40083023 40083026 40083022 40083013 40083024 40083025 40083009 40083028 40083014 40083010 40083016 40083011 40083021 40083027 40083012 40082938 40082942 40082936 40082941 40082950 40082938 40082942 40082939 40082940 40082937 40082935 40080035 40080063 40080042 40080060 40080036 40080056 40080064 40080057 40080039 40080037 40080041 40078733 40080032 40080023 40080020 40080078 40080079 40080070 40078740 40078750 40080030 40080029 40080031 40080021 40080026 40080073 40080051 40080025 40080068 40078755 40078761 40078703 40075068 40074291 40075090 40074254 40070304 40070306 40070298 40070301 40070302 40070299 40070297 40070307 40067361 40067360 40067365 40067329 40067330 40067345 40067347 40067346 40067359 40067366 40067362 40067364 40067367 40067368 40067315 40067339 40067344 40067316 40067349 40067343 40067319 40067317 40067310 40067311 40067312 40067294 40065816 40067299 40067293 40065819 40065818 40067301 40065057 40065043 40065817 40065072 40065811 40065076 40065813 40065073 40065080 40065815 40065810 40065075 40065074 40065066 40065059 40065058 40065078 40065077 40065079 40065032 40065045 40065046 40065031 40065040 40065037 40065647 40065042 40065051 40065052 40065041 40065050 40065033 40065035 40065034 40065039 40065044 40065036 40064533 40064531 40064530 40064532 40114420 40149148 40149150 40150846 40150839 40167386 40177892 40177894 40177890 40177887 40177886 40177883 40224023 40224022 40084520 40085639 40085640 40085641 40085654 40085655 40089376 40089379 40089381 40092329 40092330 40092331 40092333 40095180 40102524 40102525 40102526 40102528 40111784 40111786 40111788 40112677 40117225 40118123 40118124 40122656 40122657 40122658 40139473 40139475 40139477 40139478 40139513 40139516 40139517 40139518 40148557 40149152 40149165 40149168 40150840 40174100 40177884 40177885 40177888 40177895 40214019 40214020 40065060 40065064 40065065 40202470 40074261 40074287 40074260 40065063 40065062 40065061 40071121 40071137 40178077 40178073 40178075 40178076 40178074 40103822 40103823 40127642 40127640 40127639 40132223 40132224 40132231 40132236 40132273 40149134 40149133 40149132 40150787 40150788 40150789 40214012 40223035 40223034 40223998 40067374 40223030 40224770 40213972 40178014 40178013 40178016 40178015 40178012 40177932 40177933 40177930 40177929 40177931 40177928 40177881 40167345 40167344 40167343 40167342 40167341 40167340 40167339 40167338 40157570 40157571 40150784 40150785 40149208 40149204 40149207 40149206 40149203 40149205 40139519 40139485 40139499 40139484 40139491 40139489 40132173 40132177 40132176 40132245 40132181 40132242 40132243 40132235 40132239 40132240 40132241 40132277 40132180 40132274 40132247 40132276 40132246 40132244 40132232 40132196 40132108 40132107 40114424 40114423 40114422 40114421 40112702 40112701 40112703 40104713 40104721 40104720 40104712 40104718 40104714 40104709 40104716 40104722 40104719 40104715 40104711 40104710 40104717 40104723 40119845 40091310 40091301 40132190 40143409 40091304 40094163 40094164 40122582 40145340 40092284 40132189 40132184 40070320 40070316 40070315 40070317 40091306 40091305 40080074 40092274 40132186 40225818 40091300 40091302 40132222 40177827 40177828 40070319 40070318 40122583 40132187 40178072 40085677 40092348 40129956 40145309 40177841 40109533 40145312 40114919 40065055 40106540 40106539 40132117 40132188 40132185 40177947 40122587 40122589 40122586 40122585 40122584 40127650 40127648 40127654 40127649 40127651 40149138 40075077 40071123 40075085 40075075 40075079 40075083 40075084 40075082 40075070 40075073 40075072 40075071 40075056 40075057 40092335 40067331 40098740 40098739 40106550 40113209 40132199 40139460 40139459 40143430 40143429 40165723 40177973 40177975 40177976 40177977 40154429 40108808 40108809 40092300 40092292 40092293 40065054 40065053 40067389 40071109 40071120 40071127 40071134 40074276 40091283 40091284 40091287 40106548 40112693 40112692 40113750 40118126 40163524 40163525 40170324 40170325 40223984 40132258 40141297 40141296 40177889 40167392 40160806 40160808 40160809 40160810 40139543 40139547 40139552 40139556 40124829 40112682 40112700 40104729 40104730 40104731 40104732 40070300 40070303 40105939 40105940 40108004 40108006 40108005 40106556 40108814 40108813 40108812 40108811 40108807 40108806 40108805 40119844 40120898 40136627 40145342 40150800 40177839 40177840 40178138 40178137 40214057 40214065 40214060 40214059 40214054 40213966 40214062 40214061 40214063 40214064 40300001 40108831 40108829 40108830 40108828 40129926 40129922 40129923 40129924 40129930 40129929 40139481 40139483 40139482 40139467 40177863 40177862 40177861 40177860 40178069 40178068 40178067 40178066 40113753 40113752 40127645 40170326 40170327 40300028 40223997 40223996 40221716 40209859 40209860 40202468 40202469 40202467 40177818 40167385 40167317 40162390 40163523 40162387 40162389 40162388 40157563 40157564 40157560 40157566 40157565 40157561 40157559 40157567 40157568 40157562 40157558 40157557 40155795 40155784 40155788 40155787 40155789 40155793 40155785 40155790 40155791 40155786 40155794 40155792 40154427 40154426 40154428 40154425 40150801 40149169 40148551 40141294 40141292 40141291 40141293 40141290 40139527 40139495 40139494 40132162 40132169 40132170 40132164 40132163 40132165 40129964 40127671 40127670 40129912 40129913 40127673 40127676 40127672 40127675 40127674 40124817 40124864 40124851 40124854 40124852 40124853 40124848 40122621 40122624 40122623 40122625 40114912 40114911 40114910 40113731 40113730 40113734 40113735 40110692 40110677 40110675 40110691 40110676 40110690 40110674 40110679 40110682 40110683 40110681 40110680 40110678 40109549 40109547 40109548 40109553 40109552 40109551 40109555 40109554 40109557 40109556 40104750 40104751 40102534 40102533 40102535 40102536 40102532 40100501 40095162 40095161 40095159 40095160 40095168 40098741 40095151 40095150 40095149 40095148 40092341 40092323 40092318 40092317 40092316 40092314 40092315 40092313 40092312 40092311 40092308 40092307 40092306 40091315 40091314 40091281 40089397 40085664 40085659 40085653 40085645 40085647 40085648 40085657 40084513 40085649 40085646 40085658 40080101 40080100 40080099 40080088 40080108 40080086 40080106 40080105 40080107 40080087 40080089 40080090 40080098 40078692 40078732 40078695 40078691 40078698 40078694 40078717 40074263 40070323 40067333 40225810 40225811 40223993 40223994 40223995 40223985 40223986 40223038 40223039 40209853 40209834 40209800 40209801 40177938 40174094 40174095 40174096 40167360 40167361 40167320 40167321 40164538 40164539 40163548 40163549 40159390 40159391 40152929 40154345 40154346 40154351 40154353 40149194 40149198 40149199 40149200 40149201 40149202 40149172 40149173 40149175 40149176 40145301 40145303 40145306 40145308 40143404 40143405 40143407 40141300 40141302 40141303 40141304 40141305 40136664 40136665 40139449 40139450 40139451 40139439 40139440 40139441 40139442 40139443 40139444 40139445 40139446 40139447 40139448 40132197 40132210 40132214 40132233 40132135 40132136 40132137 40132140 40132089 40132090 40132091 40129878 40129879 40129880 40129881 40129882 40129909 40129911 40127131 40127132 40127133 40127134 40127135 40127136 40127137 40124857 40124859 40124820 40124818 40124819 40124821 40124822 40122573 40122574 40120972 40120192 40120899 40120900 40120901 40120902 40117252 40117257 40113739 40113740 40112690 40112691 40112694 40112695 40112696 40107962 40107966 40107967 40107968 40095163 40095164 40095165 40095166 40095167 40094175 40094179 40094181 40094182 40094183 40094184 40094185 40094186 40085715 40085719 40085722 40085725 40085713 40085714 40085718 40085720 40085721 40070337 40085651 40085660 40085662 40085663 40085669 40085671 40085672 40085675 40070345 40071139 40075053 40071132 40075044 40075054 40075078 40075080 40075088 40070352 40070354 40070355 40070356 40070346 40070338 40070339 40070340 40070343 40070344 40085650 40085667 40070347 40070351 40070353 40067390 40067392 40070238 40070243 40070244 40067375 40067376 40067377 40067378 40067379 40067385 40067386 40067387 40070239 40070242 40107961 40109519 40109520 40122575 40143406 40145305 40223992 6013322 40300027 40225812 40223040 40223041 40223042 40221727 40221726 40221725 40221724 40202463 40202465 40202464 40202462 40202461 40202460 40178166 40178163 40177948 40177835 40177836 40177834 40170312 40170311 40170309 40170310 40170308 40165729 40164535 40164534 40164536 40164533 40164532 40164537 40164531 40164530 40164529 40164528 40164527 40160752 40160751 40159357 40159356 40159351 40159355 40159352 40159354 40159353 40159348 40159350 40159349 40157582 40154401 40154402 40154362 40154363 40150845 40150844 40149174 40149146 40149147 40148573 40145341 40142729 40142728 40141327 40141325 40141326 40141324 40139506 40139510 40139508 40139509 40139507 40139505 40139472 40139471 40139470 40132272 40132221 40132114 40132115 40132116 40129958 40129957 40129916 40129917 40129918 40129914 40129921 40129925 40129883 40129884 40129886 40129885 40127647 40127630 40127631 40127629 40124842 40124840 40124841 40124838 40124839 40122595 40122615 40122594 40114908 40114909 40114907 40114905 40114906 40114904 40108815 40108816 40106536 40106537 40106538 40103797 40102556 40102547 40089385 40089380 40088579 40088578 40088576 40088577 40085673 40085674 40085678 40085676 40085670 40083029 40083030 40080082 40080083 40080109 40080067 40078730 40078726 40078764 40078713 40078724 40078728 40070365 40070383 40070391 40071102 40071101 40071106 40071107 40071095 40070389 40071105 40071104 40070272 40071275 40070276 40070274 40070273 40067307 40067306 40067308 40067309 40065071 40065069 40065068 40065070 40065067 40067300 40067302 40067325 40067336 40067391 40070245 40070246 40070247 40070248 40070249 40070250 40070251 40070252 40070253 40070309 40070310 40070311 40070321 40070366 40070367 40070372 40070380 40071117 40071118 40071126 40071130 40071141 40074266 40074268 40074269 40074274 40074282 40074285 40074299 40074301 40075042 40075043 40082943 40082945 40082947 40082948 40082949 40083032 40092299 40092301 40092302 40092339 40092714 40092715 40092716 40092717 40092718 40092719 40092720 40092721 40092722 40092723 40095156 40095172 40095173 40095174 40095175 40095176 40103798 40103804 40107960 40107986 40107987 40107988 40108827 40108847 40111805 40114913 40114914 40114915 40114917 40120186 40120188 40120961 40122576 40122577 40122578 40122579 40127611 40127612 40127625 40129910 40129915 40132134 40132209 40132275 40136615 40136616 40136617 40136626 40136628 40136631 40136632 40139492 40139511 40139512 40139526 40142771 40142772 40143402 40143433 40143434 40143435 40143440 40143444 40143445 40143448 40145314 40145315 40145384 40145385 40145386 40145387 40148539 40148540 40148541 40148542 40148543 40148544 40148545 40148546 40148547 40148548 40148549 40148550 40148552 40148553 40149181 40149182 40149183 40149184 40149185 40149186 40154373 40154374 40155760 40155761 40157569 40159347 40159382 40159383 40159384 40162341 40162346 40162348 40162357 40162358 40162359 40162360 40164469 40164470 40164471 40164472 40164521 40164522 40164523 40164524 40164525 40170295 40170296 40170297 40170298 40170299 40177804 40177805 40177807 40177809 40177815 40177864 40177865 40177940 40177941 40177942 40177943 40213968 40213969 40221719 40223037 40225817 40300000 40070263 40082946 40082944 40083031 40083033 40112681 40150790 40213970 40300719 40108810 40067393 40070328 40074259 40074289 40074297 40074298 40075069 40078702 40078745 40078756 40080027 40082959 40088541 40088542 40088549 40089329 40089330 40092334 40092342 40094172 40094173 40094174 40094187 40095154 40095155 40102519 40102520 40104756 40104758 40104759 40104760 40104762 40104763 40105936 40105937 40108818 40111798 40122626 40124823 40127643 40129934 40132077 40132211 40139493 40139521 40141301 40149170 40149171 40150791 40152912 40152932 40155726 40155727 40155728 40155729 40155730 40155731 40155732 40155733 40155796 40155797 40155798 40159346 40160767 40160768 40170317 40177823 40177824 40177825 40177826 40177845 40177934 40177974 40177978 40177979 40178083 40178084 40178085 40202512 40202514 40209797 40209798 40209799 40098755 40100474 40100476 40100475 40098756 40105934 40105935 40111790 40111789 40111785 40150836 40150835 40150834 40177820 40177822 40177821 40067303 40067305 40067304 40067388 40070260 40070261 40070314 40070312 40070313 40070335 40070331 40070334 40070333 40070336 40070332 40071098 40074300 40074284 40080055 40080052 40080053 40080075 40080077 40083007 40083006 40082951 40082961 40082962 40082952 40082954 40082955 40082956 40082960 40084504 40084502 40084503 40089332 40103802 40103803 40113732 40113733 40119865 40119866 40119863 40120191 40124803 40124801 40124799 40124802 40124800 40124798 40124867 40124865 40124868 40127130 40127624 40127623 40132131 40139457 40139456 40139453 40139455 40139454 40139452 40139497 40139503 40139504 40139498 40139502 40139496 40139501 40139500 40139520 40142740 40145335 40145383 40145382 40149125 40149123 40149124 40149122 40149121 40149120 40149119 40149118 40149117 40149116 40154369 40154372 40154368 40154370 40154371 40162298 40162300 40162293 40162294 40162297 40162295 40162296 40163547 40163542 40163541 40177802 40177800 40177801 40177803 40177843 40177844 40178011 40178010 40178008 40178009 40178007 40178017 40202527 40202525 40202526 40209836 40209851 40209852 40214000 40214003 40202496 40202495 40209767 40202494 40209765 40209766 40202497 40213974 40213973 40300003 40300002 40300725 40067353 40067348 40067352 40067351 40067350 40071097 40071096 40071094 40074271 40074256 40071142 40071135 40071143 40071112 40078681 40078735 40078696 40078722 40078715 40078709 40078710 40084528 40082982 40082984 40082985 40082983 40084533 40084532 40085661 40085708 40085709 40085717 40088564 40088554 40088560 40088557 40088555 40088559 40088558 40088539 40091286 40091290 40091289 40091285 40092304 40092303 40092305 40092338 40092337 40092336 40094167 40094188 40100516 40100515 40100517 40102549 40105938 40108819 40111796 40111799 40111797 40111801 40112683 40112685 40112672 40112673 40113776 40113775 40112676 40114399 40113779 40113778 40114400 40118118 40118120 40120907 40124836 40127147 40127146 40127148 40127683 40127682 40127680 40127678 40127681 40127677 40127679 40129955 40132106 40139458 40143428 40143426 40143427 40143425 40145317 40145318 40145321 40145322 40145320 40145319 40149139 40149179 40149178 40150796 40150798 40150797 40150793 40150795 40150794 40150833 40154375 40159385 40162384 40165722 40165720 40165721 40174093 40174097 40174099 40174098 40170253 40177882 40178018 40209791 40209786 40209787 40209782 40209793 40209783 40209784 40209792 40209788 40209789 40209785 40209790 40209809 40209832 40213990 40225816 40225815 40225814 40225813 40300723 40300724 40300004 40224021 40224020 40214040 40214039 40214018 40209763 40209780 40209778 40202466 40202459 40202458 40202457 40178168 40178167 40178144 40177972 40177851 40177850 40177849 40177847 40177819 40177808 40177817 40170314 40170313 40167397 40167396 40167395 40167337 40167336 40167335 40167334 40167333 40167332 40167331 40167330 40167329 40167328 40167327 40167326 40167325 40167324 40167323 40165703 40165702 40165701 40165700 40165699 40165698 40165693 40165692 40160834 40159330 40159329 40159328 40159327 40159326 40159325 40159324 40159323 40159322 40159321 40159320 40159319 40159318 40159317 40157584 40157583 40159344 40159343 40159342 40159341 40159340 40159339 40159338 40159337 40159336 40159335 40159334 40159333 40159332 40159331 40159345 40155808 40155807 40155806 40155805 40155804 40155803 40155802 40155801 40155800 40155799 40152867 40152863 40152862 40150883 40150882 40150881 40150880 40150843 40150786 40149143 40149142 40149141 40149140 40148572 40148571 40148570 40148569 40148568 40148567 40148566 40148565 40148564 40148563 40148562 40148561 40148560 40145381 40145380 40145379 40145378 40145377 40145376 40145375 40145374 40145373 40145372 40145371 40145370 40145369 40145368 40145367 40145366 40145365 40145364 40145363 40145362 40145361 40145360 40145359 40141320 40141321 40141299 40141298 40139533 40139532 40139530 40139555 40139554 40139553 40139550 40136657 40136656 40136655 40136654 40136653 40132215 40132087 40129919 40127621 40127620 40127619 40127618 40127617 40127616 40127615 40127614 40127613 40102529 40122562 40122215 40122213 40122212 40122211 40122210 40122209 40122208 40122207 40122194 40122206 40122205 40122199 40122198 40122197 40122196 40122195 40122214 40122177 40122176 40122184 40122183 40122182 40122180 40122179 40122178 40119858 40119857 40119856 40119854 40119850 40119849 40119859 40119855 40119853 40119852 40119851 40119848 40119847 40114411 40114410 40114409 40114408 40114407 40114406 40114405 40114404 40114403 40114402 40114401 40113760 40113759 40113758 40113757 40113756 40113754 40113751 40114417 40114416 40114415 40114414 40114413 40114412 40112674 40112699 40112698 40112697 40112680 40112679 40112675 40112678 40106567 40106566 40106565 40106564 40106563 40106562 40106561 40106560 40106559 40106558 40106557 40106555 40106554 40106553 40105950 40105949 40105948 40105947 40105959 40105958 40105957 40105956 40105955 40105951 40105943 40105942 40105917 40105916 40105915 40104764 40085665 40102530 40100513 40100512 40098713 40098724 40098723 40098722 40098712 40098710 40098709 40098707 40098721 40098720 40098715 40097322 40097320 40092354 40092353 40092352 40092351 40092350 40092349 40091291 40091288 40088587 40088573 40088572 40088571 40088570 40088567 40088566 40088565 40088563 40088562 40088561 40088556 40084501 40084500 40084515 40084514 40084236 40084531 40084530 40084529 40082934 40080104 40080093 40080102 40080094 40080103 40080097 40080096 40080095 40080092 40080033 40080028 40078773 40078754 40078753 40074296 40074293 40074288 40071138 40071124 40071089 40071088 40071087 40071085 40071084 40071083 40070360 40070359 40070358 40070357 40070269 40070268 40070265 40070264 40214001 40209835 40209827 40178102 40177946 40167316 40165730 40165728 40164540 40163546 40160750 40160749 40152931 40152930 40150792 40132092 40129940 40129939 40129938 40129935 40129870 40124808 40124804 40124809 40124807 40124806 40124805 40119862 40119861 40119860 40067295 40067296 40067298 40067313 40067332 40067334 40067337 40067338 40067340 40067341 40067342 40070259 40070262 40070266 40070267 40070270 40070271 40070322 40071100 40071136 40078758 40078759 40078760 40085699 40085701 40088540 40088543 40091294 40092275 40092288 40092289 40092321 40092322 40092357 40092358 40092359 40094152 40094165 40094166 40094170 40094171 40094176 40094180 40095157 40095158 40095169 40095170 40095171 40098743 40098744 40098747 40100479 40102521 40102545 40102546 40103807 40103808 40105941 40111771 40111803 40111806 40112663 40112667 40112668 40112669 40113763 40113764 40117253 40117258 40139522 40139523 40139551 40143401 40143447 40145313 40145336 40149114 40149126 40149129 40149136 40149137 40149145 40170329 40174071 40065814 40067297 40067335 40070285 40070326 40070327 40070382 40071108 40071129 40074255 40074257 40074264 40074270 40074272 40074273 40074275 40074277 40074283 40074290 40074292 40074302 40085668 40088580 40094177 40094178 40095177 40095178 40095179 40100496 40100500 40102548 40106541 40107989 40114432 40114433 40114434 40117251 40122580 40122581 40122660 40124794 40136603 40139462 40139463 40139464 40139465 40139466 40139469 40142770 40145300 40163544 40163550 40163551 40163552 40163553 40164468 40177842 40177939 40178139 40178140 40209796 40214014 40214015 40214016 40214017 40221718 40221720 400070325 40224769 40224768 40224766 40224767 40129887 40129888 40082963 40082965 40082966 40092285 40092283 40124815 40124816 40164526 40224771 40104748 40104749 40104752 40104753 40104754 40104755 40104761 40127138 40127139 40136634 40136635 40154352 40213992 40213993 40213994 40213995 40213996 40213997 40213998 40213999 40214002 40214004 40214005 40214006 40214007 40214008 40214009 40214010 40214011 40214013 40225824 40225825 40225826 40213967 40159393 40167319 40167318 40092309 40092282 40092280 40092279 40092276 40092278 40092273 40092281 40092328 40092325 40092324 40092326 40092327 40102537 40103801 40103800 40103799 40103819 40103818 40103811 40120923 40120926 40120922 40124810 40124812 40124814 40124811 40124813 40129943 40129944 40129941 40132105 40132103 40132097 40132099 40132094 40132098 40132102 40132101 40132100 40132104 40132096 40132095 40132255 40132257 40132254 40132253 40136630 40136621 40136622 40136624 40136619 40136629 40136620 40136623 40136618 40141295 40148578 40148576 40148577 40148574 40148579 40148575 40154378 40154367 40154380 40154379 40154382 40154376 40154366 40155764 40155763 40155762 40160816 40177914 40300015 40300011 40300014 40300022 40300010 40300020 40300009 40300026 40300023 40300013 40300012 40300018 40300025 40300019 40300016 40300017 40300007 40300008 40300024 40224763 40224764 40224761 40224762 40224760 40224759 40224758 40224757 40224756 40224755 40224754 40224753 40224752 40224751 40224750 40224749 40224748 40202503 40202505 40202502 40202501 40213975 40213978 40213977 40213976 40213979 40213980 40213981 40213982 40213983 40213984 40213985 40209818 40209817 40209814 40209812 40209811 40209810 40209808 40209807 40178082 40178080 40178078 40177880 40177879 40177878 40177794 40177810 40177812 40177816 40177813 40177798 40177797 40177796 40177795 40170321 40170319 40170318 40167391 40167390 40167389 40167388 40167387 40162369 40162370 40162368 40162367 40162366 40160760 40160766 40160765 40160764 40160763 40160756 40160762 40160758 40160761 40160759 40155749 40155748 40155747 40155746 40155745 40155744 40155742 40155741 40155743 40155740 40154388 40154389 40154391 40154387 40154386 40154390 40154384 40154392 40154383 40149197 40149196 40149193 40149195 40149192 40149190 40149189 40149191 40149188 40149187 40145345 40145353 40145348 40145347 40145350 40145352 40145344 40145349 40145346 40145351 40145295 40145294 40145292 40145293 40141323 40141322 40141319 40141318 40132151 40132155 40132152 40132153 40132146 40132149 40132150 40132148 40132147 40132154 40129945 40129951 40129953 40129954 40129952 40129948 40129950 40129946 40129949 40129947 40124824 40124827 40124830 40124834 40124828 40122572 40122570 40122569 40122568 40122567 40122565 40122563 40117267 40117268 40117269 40117270 40118113 40118112 40117260 40117262 40117263 40117264 40117266 40117265 40117261 40117256 40117259 40117255 40117234 40117236 40117235 40117233 40117232 40117231 40117230 40117229 40117228 40117226 40117227 40117224 40117223 40114942 40114944 40114940 40114945 40114941 40114939 40114938 40114937 40114936 40114934 40113741 40113745 40113742 40113743 40113738 40113746 40103834 40103836 40103832 40103831 40104708 40103837 40103828 40103839 40103829 40104707 40104705 40104706 40103827 40103826 40103824 40103825 40103812 40103813 40102512 40102518 40102511 40102514 40102515 40100518 40102513 40102516 40102517 40100522 40100519 40102509 40100520 40102508 40102507 40102510 40100523 40100511 40100514 40100494 40100497 40100492 40100499 40100490 40100493 40100491 40100495 40100489 40100488 40100486 40100484 40100498 40100480 40091312 40091321 40091319 40091313 40092310 40091320 40091325 40091317 40091318 40091322 40091311 40091309 40091308 40091307 40091297 40091293 40091292 40091279 40091282 40091278 40091277 40091276 40089422 40089423 40089417 40089415 40089416 40089413 40089414 40089419 40089410 40089411 40089412 40089409 40089395 40089408 40089396 40089393 40089391 40089390 40089394 40089392 40089389 40089387 40089388 40089367 40089368 40089366 40089365 40089364 40089353 40089352 40089355 40089351 40085726 40085731 40085732 40085730 40085728 40085729 40082991 40082990 40085724 40082976 40082974 40082969 40082973 40082977 40082978 40082970 40082971 40082972 40082967 40082968 40082964 40078674 40078707 40078708 40078747 40078685 40078746 40078770 40071116 40071114 40070373 40071125 40071122 40071113 40071111 40070368 40070362 40070364 40214044 40214050 40214047 40214048 40214046 40214049 40214045 40209848 40178070 40067326 40162352 40162350 40162349 40162351 40159388 40159387 40159386 40159389 40157533 40157532 40157531 40154417 40154416 40154415 40154412 40154411 40154413 40154408 40154409 40154410 40154407 40154404 40154405 40154406 40154424 40154420 40154423 40154422 40154419 40154418 40154414 40154403 40067327 40067328 40067369 40067370 40132229 40132259 40132083 40132082 40132084 40127145 40127144 40127143 40127140 40127142 40127141 40067373 40071140 40074253 40074262 40074265 40074267 40074278 40074279 40074281 40074294 40074295 40082992 40083000 40089378 40108851 40108852 40109527 40109528 40109529 40109530 40139486 40139487 40139488 40139490 40170315 40065081 40070361 40070374 40070390 40075047 40075086 40080119 40082953 40082957 40082958 40167394 40167393 40088568 40088569 40112655 40112664 40112665 40112666 40139528 40139539 40139540 40214021 40214022 40224765 40224003 40223990 40223987 40223988 40223989 40223991 40223032 40214058 40209838 40177945 40177944 40177867 40177866 40177869 40177868 40177832 40177833 40177831 40177830 40065038 40067314 40067318 40067320 40067321 40067322 40067323 40067324 40067354 40067355 40067356 40067357 40067358 40067371 40067372 40070363 40167346 40167347 40167322 40165724 40163543 40162337 40162340 40162339 40150849 40150847 40150848 40071092 40071099 40071119 40071131 40071133 40075045 40075046 40078673 40078675 40078676 40078680 40078682 40078684 40078688 40078693 40078697 40078701 40078705 40132260 40132264 40132139 40132132 40132133 40132078 40129965 40132079 40132076 40132081 41132080 40132086 40129969 40129967 40129968 40129966 40078706 40078711 40078734 40078748 40080022 40080024 40080034 40080050 40080065 40080066 40080069 40080071 40080072 40080076 40082993 40082994 40082995 40082996 40082997 40083001 40083002 40083003 40089356 40089357 40089358 40089369 40089374 40089375 40091323 40091324 40092294 40092295 40092296 40092297 40092298 40094159 40094161 40104724 40114946 40117245 40117246 40117247 40117248 40118119 40139479 40139480 40141273 40142730 40142731 40142732 40142733 40142734 40142773 40142774 40142775 40142776 40142777 40143432 40145296 40145297 40145298 40149135 40170260 40170261 40170262 40170264 40170265 40170267 40170300 40170301 40170302 40170303 40170304 40170305 40170306 40170307 40177799 40177806 40065047 40065048 40070386 40078741 40080058 40080062 40080085 40080111 40080112 40080113 40080114 40080115 40080116 40080117 40080118 40082998 40082999 40083004 40091316 40092343 40092344 40092345 40092346 40092347 40102523 40102527 40102531 40104725 40104727 40105928 40107997 40107999 40108000 40108001 40108002 40108003 40108007 40108008 40108804 40112654 40112656 40112662 40114929 40114930 40118121 40132238 40142741 40142742 40142743 40142744 40142745 40142746 40142747 40142748 40142749 40142750 40142751 40142752 40142753 40142754 40142755 40142756 40142757 40142758 40142759 40142760 40142761 40142762 40142763 40142764 40142765 40142766 40142767 40142768 40145323 40145324 40157485 40157486 40157488 40157489 40157490 40157491 40157492 40157493 40157494 40157495 40157496 40157497 40157498 40157499 40157500 40157501 40157502 40157503 40157504 40157505 40157506 40157507 40157508 40157534 40167409 40167410 40167411 40167412 40177814 40177912 40177913 40209821 40209828 40209829 40209830 40213988 40213989 40070284 40213991 40223028 40223029 40225823, and 40170316. |
Recalling Firm/ Manufacturer |
GE Healthcare 9900 Innovation Drive Wauwatosa WI 53226
|
Manufacturer Reason for Recall |
GE Healthcare has received reports of breathing circuit or tubing misconnections of the Anesthesia Delivery Unit (ADU), that may impact patient safety.
|
FDA Determined Cause 2 |
Other |
Action |
A GE Healthcare "Urgent Medical Device Correction" letter was sent to consignees on 1/16/09. The letter was addressed to Health Care Administrator, Chief of Anesthesia and Director of Clinical Engineering. The letter described the Safety Issue; Affected Product Details; Safety Instructions, which include clinical personnel being properly trained on the correct connections of breathing circuit tubing and proper execution of System Checkout according to the ADU URMs; Product Correction and Contact Information. |
Quantity in Commerce |
6,621 |
Distribution |
USA including states of AL, AK, AZ, AR, CA, CO, DC, FL, GA, HI, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, NE, NV, NH, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, VT, VA, WA, and WI and countries of Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bolivia, Bosnia & Herzegovina, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Egypt, England, Estonia, Finland, France, Germany, Greece, Hong Kong, Iceland, India, Iraq, Ireland, Isle of Man (UK), Israel, Italy, Japan, Jordan, Kazahkstan, Kuwait, Latin America, Latvia, Lebanon, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Paraguay, Peru, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Uruguay, and Yemen. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = BSZ and Original Applicant = DATEX-ENGSTROM, INC. 510(K)s with Product Code = BSZ and Original Applicant = DATEX-OHMEDA, INC.
|
|
|
|