Date Initiated by FirmJanuary 06, 2009
Date PostedJuly 22, 2009
Recall Status1 Terminated 3 on June 25, 2012
Recall NumberZ-1375-2009
Recall Event ID 50901
510(K)NumberK993768 
Product Classification Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate - Product Code MEH
ProductCitation TMZF HA Hip Stem - LT. Hip stems are single use devices intended for cementless fixation. The products are intended for use in primary reconstruction of the proximal femur or revision of a previous total hip arthroplasty.
Code Information 6265-5105 CITATION TMZF SIZE 5 LEFT 24706601 6265-5105 CITATION TMZF SIZE 5 LEFT 25044901 6265-5105 CITATION TMZF SIZE 5 LEFT 25044902 and 6265-5105 CITATION TMZF SIZE 5 LEFT 25068301. 
Recalling Firm/
Manufacturer
Stryker Howmedica Osteonics Corp.
325 Corporate Dr
Mahwah NJ 07430
For Additional Information ContactColleen O'Meara
201-831-5000
Manufacturer Reason
for Recall
Product deviation : Specific lots of sprayed HA hip stems did not meet Stryker's Internal Material Specification for tensile bond strength and crystallinity.
FDA Determined
Cause 2
Process control
ActionAn "Urgent Product Recall" letter was sent via Federal Express on January 6, 2009 with return receipt to Stryker Branches/Agencies, Hospital Risk Management, Hospital Chief of Orthopaedics and Surgeons. The letter described the affected product, issue, potential hazards and risks. Stryker Branches and Hospital Risk Management were asked to examine their inventory for affected product. Stryker Representatives were instructed to contact customers to reconcile and return recalled product to Stryker Orthopaedics. All letters include a response form to be completed and returned indicating that they have been informed of the recall and the amount of product on hand, if any. Direct questions about the recall to Stryker Orthopaedics at 1-201-831-5718.
Quantity in Commerce1,628 total stems, all varieties.
DistributionNationwide Distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.