| | Class 2 Device Recall ThermoFisher Scientific Microm STP 420D |  |
| Date Initiated by Firm | June 03, 2009 |
| Date Posted | April 12, 2010 |
| Recall Status1 |
Terminated 3 on June 29, 2011 |
| Recall Number | Z-1320-2010 |
| Recall Event ID |
52175 |
| Product Classification |
Processor, tissue, automated - Product Code IEO
|
| Product | ThermoFisher Scientific Microm STP 420D automated tissue processor. |
| Code Information |
All units. |
Recalling Firm/ Manufacturer |
Thermo Fisher Scientific 4481 Campus Dr Kalamazoo MI 49008-2590
|
Manufacturer Reason for Recall | The firm has received complaints regarding specimen integrity (higher rate of specimen fragmentation) as well as complaints that the chamber may stop rotating prematurely, resulting in sample deterioration. |
FDA Determined Cause 2 | Other |
| Action | Consignees were notified of the problem by letter, on or about 6/3/09, updating the instructions for use. The units were subsequently upgraded on or before 9/8/09. For additional questions and information, please contact ThermoFisher's Technical Support department , Subject: PT Module Advisory Notice. |
| Quantity in Commerce | 16 |
| Distribution | Hawaii, Kentucky, Louisiana, Michigan, Missouri, Nebraska, New York, North Carolina, Ohio, Pennsylvania, Tennessee, Texas and West Virginia. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|