| Class 2 Device Recall Bausch & Lomb Stellaris Vacuum Phaco Pack | |
Date Initiated by Firm | January 08, 2010 |
Date Posted | May 28, 2010 |
Recall Status1 |
Terminated 3 on April 02, 2012 |
Recall Number | Z-1713-2010 |
Recall Event ID |
54200 |
510(K)Number | K063331 |
Product Classification |
unit, phacofragmentation - Product Code HQE
|
Product | Bausch & Lomb Stellaris Vacuum Phaco Pack, Model BL5111, 6 packs/box, RX, Sterile. The firm name on the label is Bausch & Lomb Incorporated, Rochester, NY.
The plastic needle wrench contained in the package is used to tighten or untighten phacoemulsification needles on phaco handpieces of the Millenium and Stellaris machines. The Millenium and Stellaris are used by ophthalmic surgeons to emulsify and aspirate cataract material during cataract surgery, to cut and aspirate vitreous material during anterior and posterior segment surgery, and to drive a variety of handpieces during posterior segment surgery. |
Code Information |
U1577 and U1605 |
Recalling Firm/ Manufacturer |
Bausch & Lomb Inc 1400 North Goodman Street Rochester NY 14609-3547
|
For Additional Information Contact | Chris Willis 949-297-3511 |
Manufacturer Reason for Recall | Users may incorrectly assemble the plastic needle wrench to the phacoemulsification needle resulting in the generation of plastic particulate |
FDA Determined Cause 2 | Component design/selection |
Action | The recalling firm, Bausch & Lomb, issued an "URGENT: PHACO NEEDLE WRENCH CORRECTION" letter dated January 7, 2010, via first class mail to all customers. The letter describes the product, problem and actions to be taken by the customers. The letter also provides instructions and pictorial representations demonstrating proper seating of the needle to the wrench as well as a description of the appropriate force to tighten the needle to the handpiece. The customers were instructed to please read and follow the attached instructions entitled (Potential for Particulate Matter in the eye and proper assembly and use of needle wrench) and fill-out and return/fax at 949-716-8362, the Bausch & Lomb Phacoemulsification Needle Wrench Particulate correction form.
If you have any questions or concerns please contact, Chris Willis or Stephen Haight at 1-800-387-3284. |
Quantity in Commerce | 9,967/6-pack boxes |
Distribution | Worldwide distribution: USA and Germany, France, Italy, Norway, Sweden, Finland, The Netherlands, Switzerland, U.K., Austria, Luxembourg, Ireland, Hong Kong, New Zealand, India, Australia, and Japan. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|