| | Class 2 Device Recall JOURNEY " UNI TIBIAL INSERT RIGHT MEDIAL/ LEFT LATERAL, |  |
| Date Initiated by Firm | January 23, 2009 |
| Date Posted | March 03, 2010 |
| Recall Status1 |
Terminated 3 on March 25, 2011 |
| Recall Number | Z-0946-2010 |
| Recall Event ID |
54269 |
| 510(K)Number | K052265 |
| Product Classification |
Knee implant - Product Code HSX
|
| Product | JOURNEY " UNI TIBIAL INSERT RIGHT MEDIAL/ LEFT LATERAL, A/P 46 MM, M/L 27 MM, REF 71422255, QTY: (1), STERILE EO, SMITH & NEPHEW, INC. MEMPHIS, TN 38116 |
| Code Information |
All Lot Numbers |
Recalling Firm/ Manufacturer |
Smith & Nephew Inc 1450 E. Brooks Rd Memphis TN 38116
|
| For Additional Information Contact | David Archer 901-399-5427 |
Manufacturer Reason for Recall | The firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct. |
FDA Determined Cause 2 | Device Design |
| Action | All affected Smith & Nephew Sales Representatives were notified of problem and the recall overnight mail on 12/23/2009. All affected Smith & Nephew International Distributors were notified via e-mail and telephone on 01/23/2009. They were instructed to immediately quarantine the product for return. For questions, please contact the firm, Leah Easley at 901-399-6137. |
| Quantity in Commerce | 1907 units |
| Distribution | US and Argentina, Switzerland, Australia, Canada, Belgium, Italy, Spain, Thailand, Austria; PROC, Finland, Germany, Malaysia, Portugal, Netherlands, Singapore, and England |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|