|
Class 2 Device Recall FHC microTargeting Guideline 4000 |
|
Date Initiated by Firm |
December 28, 2009 |
Date Posted |
March 16, 2010 |
Recall Status1 |
Terminated 3 on March 18, 2010 |
Recall Number |
Z-1143-2010 |
Recall Event ID |
54278 |
510(K)Number |
K071364
|
Product Classification |
Depth Electrode - Product Code GZL
|
Product |
FHC microTargeting Guideline 4000 ver 1.4 (GL4K) Software, Catalog # MT-GL4K.
Intended to assist in functional neurological procedures. |
Code Information |
Software ver 1.4, Serial Numbers: 19454, 14423, 17349, 19215, 15573, 19424, 17583, 17589, 17588, and 17581. This version of software was released in August 2009 |
Recalling Firm/ Manufacturer |
FHC, Inc. 1201 Main Street Bowdoinham ME 04008
|
For Additional Information Contact |
same as recall contact 207-666-8190
|
Manufacturer Reason for Recall |
The raster tab for the GL4K version 1.4 software may assign incorrect depths for waveform recordings and if used to identify the implantation target for the DBS lead, the implant may be placed in the wrong location.
|
FDA Determined Cause 2 |
Software change control |
Action |
FHC notified customers by telephone and by an Urgent: Medical Device Correction email, dated December 2009, beginning December 28th. Customers have been instructed to keep the instruction with the GL4K until FHC can provide a revision to the software; and, to use only "distance from zero" or "distance traveled" modes (not "distance from target" mode) during procedures using the raster tab. Customers are to notify their FHC representative to confirm receipt of the notification. A software patch has been verified and validated and will be installed on all affected units. The new version of the corrected software is Guideline 4000 version 1.41. Questions should be directed to Jason Moyer, product manager or Keri Seitz, recall coordinator at 207-666-8190. |
Quantity in Commerce |
10 |
Distribution |
Worldwide Distribution -- USA, including states of GA, MA, NC, NJ, OH, RI, TN, TX, and WV and Bucharest, Romania. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = GZL and Original Applicant = FHC, INC.
|
|
|
|