Class 2 Device Recall ACCUTNI AND ACCUTNI CALIBRATORS ON THE ACCESS IMMUNOASSAY SYSTEM 33340, 33345

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall ACCUTNI AND ACCUTNI CALIBRATORS ON THE ACCESS IMMUNOASSAY SYSTEM 33340, 33345 see related information
Date Initiated by Firm February 06, 2010
Date Posted April 12, 2010
Recall Status1 Terminated 3 on May 09, 2012
Recall Number Z-1318-2010
Recall Event ID 54729
510(K)Number K010429  K021814  
Product Classification immunoassay method, troponin subunit - Product Code MMI
Product Access Immunoassay Systems Accu Tnl Reagent Kit when used in conjunction with UniCel Dxl Access Immunoassay Systems; Reagent Part Numbers: A78803 & 33340; Calibrator Part Number: 33345

The intended use: The Access AccuTnI assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cardiac troponin I (cTnI) levels in human serum and plasma using the Access Immunoassay Systems to aid in the diagnosis and treatment of myocardial infarction and cardiac muscle damage. Cardiac Troponin I determination also aids in the risk stratification of patients with unstable angina or non-ST segment elevation acute coronary syndromes with respect to relative risk of mortality, myocardial infarction, or increased probability of ischemic events requiring urgent revascularization procedures.
Code Information Lot Numbers:  716707, 720541,  720542,  721451,  721452,  722677,  722678,  821024,  821025,  821515, 821663, 823314, 823523, 823524, 823850, 825783, 825784, 826258, 826259, 827876, 828479, 829974, 831218, 850000,  850003, 850004, 850005, 850010, 850011, 850013, 908322, 908323, 908917, 908918, 909803, 909804, 911338, 911339, 911748, 912628, 912629, 913131, 913132, 913265, 913274, 913500, 913821, 914632, 915132, 915417, 915670, 915850, 916720, 916721,  916721A,  917611, 918460 Note: These Lot Numbers represent all reagent lots released to field that could be used in conjunction with Unicel DxI Access Immunoassay Systems. 
Recalling Firm/
Manufacturer
Beckman Coulter Inc.
200 S Kraemer Blvd
Brea CA 92821-6208
For Additional Information Contact Nora Zerounian
714-961-3634
Manufacturer Reason
for Recall
The recall was initiated because Beckman Coulter has confirmed customer reports that different results have been obtained using the same patient samples on Access/Access 2 and UniCel Dxl platforms. Values obtained with UniCel Dxl systems have been demonstrated to have a positive bias compared to values obtained with Access or Access 2 systems.
FDA Determined
Cause 2
Other
Action An "URGENT: PRODUCT CORRECTIVE ACTION (PCA)" letter dated February 5, 2010, was sent to the customers. The letter describes the product, problem and action to be taken by the customers. The customers are recommended to immediately discontinue using AccuTnI Reagents and Calibrators on the UniCel DxI system until further notice. If this is not possible, the customers are directed to notify their stakeholders of the bias until further notice. The customers should complete, fax (786-639-4000) or mail the enclosed RESPONSE FORM. If consignees need assistance or have any questions regarding the notification, they were instructed to contact Technical Support at 1-800-854-3633 in the United States and Canada.
Quantity in Commerce 196,071 kits
Distribution Worldwide distribution: USA and Canada
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = MMI and Original Applicant = BECKMAN COULTER, INC.
-
-