|
Class 2 Device Recall Dimension Vista System |
|
Date Initiated by Firm |
May 05, 2010 |
Date Posted |
September 15, 2010 |
Recall Status1 |
Terminated 3 on October 25, 2010 |
Recall Number |
Z-2423-2010 |
Recall Event ID |
55763 |
510(K)Number |
K051087
|
Product Classification |
in vitro diagnostics - Product Code JJE
|
Product |
Siemens Healthcare Diagnostics, Inc., Dimension Vista Intelligent Laboratory System with Alpha Pinnacle 10kVA UPS External Battery Pack. Part numbers: 070200001, 070200011, 070200021; Dimension Vista 1500 and 3000T instruments. |
Code Information |
Part numbers: 070200001, 070200011, 070200021; Dimension Vista 1500 and 3000T instruments within the serial number range of B01060900055 through B01090100414R0, excluding B01081100388R0, B01081100389R0, B01081200396R0, B01081200402R0, B01081200403R0, B01081200404R0, B01081200406R0, B01081200408R0, B01081200410R0, B01081200411R0, B01090100412R0, B01090100413R0. |
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc. 500 GBC Drive, Mailstop 514 PO BOX 6101 Newark DE 19714-6101
|
For Additional Information Contact |
302-631-6311
|
Manufacturer Reason for Recall |
Defective batteries used in the Alpha Pinnacle P10 UPS External Battery Pack may overheat and cause a strong sulfuric odor to be released and a corrosive sulfuric acid solution leak to occur.
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
Siemens issued an Urgent Field Safety Notice in May 2010 to customers with information on what to do if the battery leaks, overheats, or vents an odor. Firm also issued a Support Bulletin to field service personnel to inspect and replace affected units. |
Quantity in Commerce |
228 units |
Distribution |
Worldwide distribution: USA, Australia, Austria, Belgium, Canada, France, Germany, Italy, Japan, Malaysia, Netherlands, New Zealand, Portugal, Slovenia, Spain, and Switzerland. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = JJE and Original Applicant = DADE BEHRING, INC.
|
|
|
|