Date Initiated by Firm | June 07, 2010 |
Date Posted | August 25, 2010 |
Recall Status1 |
Terminated 3 on September 29, 2012 |
Recall Number | Z-2274-2010 |
Recall Event ID |
55933 |
510(K)Number | K781846 |
Product Classification |
introducer - Product Code KGZ
|
Product | Arrow Percutaneous Sheath Introducer Kit with Integral Hemostasis Valve/Side Port for use with 7.5 Fr. Catheters; Catalog number AJ-09801 |
Code Information |
Lot numbers RF0018145, RF0021641, RF9044164, RF9069739, RF9070837, RF9100262 |
Recalling Firm/ Manufacturer |
Arrow International Inc 2400 Bernville Road Reading PA 19605
|
For Additional Information Contact | 610-378-0131 |
Manufacturer Reason for Recall | Sterility not assured. The stopcock handle position was inadvertently changed from an open to a closed position, which prevents sterilization of the closed portion of the device. Sterility of the affected product cannot be guaranteed. |
FDA Determined Cause 2 | Process change control |
Action | June 7, 2010 Notification letter to customer instructed them to return the product to Arrow. Arrow will destroy the returned product. |
Quantity in Commerce | 3000 kits |
Distribution | All affected product was distributed to one foreign consignee: Arrow Japan Ltd. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = KGZ
|