| Class 2 Device Recall BIPLANE XRay Imaging Systems with Velara Generator | |
Date Initiated by Firm | February 04, 2010 |
Date Posted | September 22, 2010 |
Recall Status1 |
Terminated 3 on September 23, 2011 |
Recall Number | Z-2449-2010 |
Recall Event ID |
56093 |
510(K)Number | K031333 K033737 K984545 |
Product Classification |
System, X-Ray, Angiographic - Product Code IZI
|
Product | BIPLANE X-ray Imaging Systems included Allura XPER FD10/10, FD20/10, FD20/20, Integris BH5000, and Integris Allura 12-15 Biplane.
The BIPLANE system has two C-arms and it allows the operator to image in two planes: Lateral and Frontal. It has a C-arm on the stand and C-arm suspended from the ceiling.
The product's brochure is labeled in part: "Philips Healthcare Global Information Center...P.O. Box 1286...5602 BG Eindhoven...The Netherlands". |
Code Information |
Each unit is identified with a SITE number: 76992, 103526, 103581, 103921, 104020, 104021, 104068, 104386, 104497, 104628, 104926, 104950, 104953, 104959, 105184, 105399, 105581, 105585, 105654, 504655, 504674, 504733, 504742, 504779, 504868, 504921, 505067, 505068, 505120, 505356, 505380, 505530, 505611, 505703, 505903, 506008, 506026, 506069, 506119, 506509, 506655, 519118, 519136, 519196, 519197, 519678, 519730, 519981, 520275, 520582, 520666, 520688, 520756, 520793, 520949, 520964, 521149, 521325, 521525, 522136, 522456, 530396, 530653, 530654, 530660, 531649, 531758, 531799, 531926, 531939, 532152, 532218, 532240, 532241, 532383, 532614, 533288, 533290, 533397, 534131, 534156, 534786, 535345, 535761, 535775, 536028, 536031, 536034, 536035, 536037, 536042, 536171, 536715, 536789, 536899, 537175, 537593, 537631, 537804, 538028, 538095, 538099, 538157, 538172, 538387, 538409, 538514, 538516, 539213, 539326, 539628, 540184, 540237, 540461, 540721, 541237, 541238, 541239, 541276, 541577, 541589, 541590, 542382, 542413, 542557, 542714, 542963, 543289, 543411, 543690, 543879, 543953, 543989, 544306, 544395, 544543, 544635, 544656, 544668, 544844, 545025, 545262, 545396, 545518, 545583, 545604, 545753, 545882, 545884, 545997, 546013, 546119, 546149, 546569, 546906, 547173, 547215, 547431, 547642, 547715, 547769, 547823, 547922, 548005, 548020, 548036, 548082, 548140, 548288, 548383, 548618, 549202, 549242, 549368, 549370, 549903, 550198, 550691, 550758, 550812, 550935, 551855, 552355, 552599, 552639, 552969, 553685, 553699, 553766, 554763, 554802, 554970, 555293, 555299, 555356, 555834, 558604, 558945, 41443586, 41443609, 41443627, 41443647, 41443727, 41443765, 41443804, 41443909, 41444645, 41444647, 41444650, 41444689, 41444752, 41444816, 41445118, 41445432, 41445448, 41445501, 41445538, 41445615, 41445628, 41446230, 41446262, 41446773, 41455885, 41455897, 41455920, 41455925, 41455933, 41455944, 41455953, 41660520, 41780160, 41793290, 41795344, 41796154, 42046093, 42046440, 42057324, 42064058, 42067824, 42394072, 42479274, 42508759, 42553772, 42564341, 42625859, 42735090, 42891020, 42891095, 42891704, 43043666, 43058934, 43164789, 43173591, 43315510, 43436747, 43443622, 43453735, 43635004, 43670742, 43848923, 44028554, 44066424, 44584972, 44616211, 44654768, 44735751, and 44892370. |
Recalling Firm/ Manufacturer |
Philips Medical Systems North America Co. Phillips 22100 Bothell Everett Hwy Bothell WA 98021-8431
|
For Additional Information Contact | Bob Pettis 425-487-7602 |
Manufacturer Reason for Recall | A lock-up might occur inside the Velara Generator due to intermittent shorts inside the system. The problem can only resolved by a warm or cold system restart. In exceptional cases the power converters of the KVmA unit in the generator are damaged and the system can not be restarted. |
FDA Determined Cause 2 | Software design |
Action | On 02/04/2010, Philips Healthcare began sending out the URGENT-Device Correction Notices to their consignees. The firm issued two notifications, one for the Allura Integris and Xper FD10C, FD10F, FD10F/C and FD20 Monoplane X-Ray imaging systems with the Velara Generator. The second notification is for the Allura Integris Xper FD10 and FD20 Biplane X-ray imaging systems with the Velara Generator.
This notice informs the consignees that a system locks up during a critical interventional case and it might impose a serious injury to patients. This lock up issue could only be resolved by a warm or cold system restart. A warm restart would take about 90 seconds or a cold restart would take about four minutes, and the device could not be used during this time.
If the system does not work after the restart, the consignees should contact the Philips Healthcare Customer Service Representative at 1-800-772-9377 #5 (then follow prompts). A service representative will visit each consignee and conduct a software and hardware upgrade. |
Quantity in Commerce | 267 |
Distribution | Nationwide Distribution |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = IZI
|
|
|
|