  Class 2 Device Recall Inion CPS see related information
Date Initiated by Firm April 14, 2010
Date Posted December 16, 2010
Recall Status1 Terminated 3 on June 06, 2012
Recall Number Z-0683-2011
Recall Event ID 55861
Product Classification Plate, bone - Product Code JEY
Product 2.0 mm Inion CPS, 2.0 x 5 mm screw, REF SCR-1224, sterile, Manufacturer, INION Oy, Finland.
Code Information Lot numbers of 0700000 and above.
Recalling Firm/
Manufacturer
Stryker Craniomaxillofacial Division
750 Trade Centre Way
Ste 200
Portage MI 49002-0482
For Additional Information Contact
877-534-2464
Manufacturer Reason
for Recall
Inadequate Instructions for Use. The stated strength retention is ambiguous and the maximum immersion time of the plates/meshed is not defined.
FDA Determined
Cause 2
Other
Action Stryker issued an "URGENT MEDICAL DEVICE CORRECTION" letter dated April 14, 2010. The notice informed users of the change in the instructions for use (IFU), and included a copy of the revised IFU. The consignees were informed that the affected product does not need to be returned. Stryker can be contacted about this action at 877-534-2464 (option "Regulatory Affairs", reference Product Field Action #2009-445)
Quantity in Commerce 8631
Distribution Nationwide Distribution: Throughout the US
Total Product Life Cycle TPLC Device Report