 Class 2 Device Recall Inion CPSsee related information
Date Initiated by FirmApril 14, 2010
Date PostedDecember 16, 2010
Recall Status1 Terminated 3 on June 06, 2012
Recall NumberZ-0687-2011
Recall Event ID 55861
Product Classification Plate, bone - Product Code JEY
Product3.1 mm Inion CPS, 3.1 x 14 mm screw, REF SCR-1228, sterile, Manufacturer, INION Oy, Finland.
Code Information Lot numbers of 0700000 and above.
FEI Number 3005101424
Recalling Firm/
Manufacturer
Stryker Craniomaxillofacial Division
750 Trade Centre Way
Ste 200
Portage MI 49002-0482
For Additional Information Contact
877-534-2464
Manufacturer Reason
for Recall
Inadequate Instructions for Use. The stated strength retention is ambiguous and the maximum immersion time of the plates/meshed is not defined.
FDA Determined
Cause 2
Other
ActionStryker issued an "URGENT MEDICAL DEVICE CORRECTION" letter dated April 14, 2010. The notice informed users of the change in the instructions for use (IFU), and included a copy of the revised IFU. The consignees were informed that the affected product does not need to be returned. Stryker can be contacted about this action at 877-534-2464 (option "Regulatory Affairs", reference Product Field Action #2009-445)
Quantity in Commerce8631 all products
DistributionNationwide Distribution: Throughout the US
Total Product Life CycleTPLC Device Report