|
Class 2 Device Recall Statspin Express 4 Centrifuge |
|
Date Initiated by Firm |
August 18, 2010 |
Date Posted |
May 10, 2011 |
Recall Status1 |
Terminated 3 on September 01, 2020 |
Recall Number |
Z-2169-2011 |
Recall Event ID |
57446 |
Product Classification |
Centrifuges, (micro, ultra, refrigerated) for clinical use - Product Code JQC
|
Product |
Statspin Express 4 Centrifuge Model Number: SSH4 Intended Use: Rapid Separation of plasma and serum from primary gel collection tubes |
Code Information |
Serial Numbers: 737M51000100 737M51000101 737M51000102 737M51000103 737M51000104 737M51000105 737M51000106 737M51000107 737M51000108 737M51000109 737M51000111 737M51000112 737M51000113 737M51000114 737M51000115 737M51000116 737M51000117 737M51000118 737M51000119 737M51000120 737M51000121 737M51000122 737M51000123 737M51000124 747M51000125 747M51000127 747M51000128 747M51000129 747M51000130 747M51000131 747M51000132 747M51000133 747M51000134 747M51000135 747M51000136 747M51000137 747M51000138 747M51000139 747M51000140 747M51000141 747M51000142 747M51000143 747M51000144 747M51000145 747M51000146 747M51000147 747M51000148 747M51000149 751M51000150 751M51000151 751M51000152 751M51000153 751M51000154 751M51000155 751M51000156 751M51000157 751M51000158 751M51000159 751M51000160 751M51000161 751M51000162 751M51000163 751M51000164 751M51000165 751M51000166 751M51000167 751M51000168 751M51000169 751M51000170 751M51000171 751M51000172 751M51000174 752M51000175 752M51000176 752M51000177 752M51000178 752M51000179 752M51000180 752M51000181 752M51000182 752M51000183 752M51000184 752M51000185 752M51000186 752M51000187 752M51000188 752M51000189 752M51000190 752M51000191 752M51000192 752M51000193 752M51000194 752M51000195 752M51000196 752M51000197 752M51000198 752M51000199 802M51000201 802M51000202 802M51000203 802M51000204 802M51000205 802M51000206 802M51000207 802M51000208 802M51000209 802M51000210 802M51000211 802M51000212 802M51000213 802M51000214 802M51000215 802M51000216 802M51000217 802M51000218 802M51000219 802M51000220 802M51000221 802M51000222 802M51000223 802M51000224 805M51000225 805M51000226 805M51000227 805M51000228 805M51000229 805M51000230 805M51000231 805M51000232 805M51000233 805M51000234 805M51000235 805M51000236 805M51000237 805M51000238 805M51000239 805M51000240 805M51000241 805M51000242 805M51000243 805M51000244 805M51000245 805M51000246 805M51000247 805M51000248 805M51000249 806M51000250 806M51000251 806M51000252 806M51000253 806M51000254 806M51000255 806M51000256 806M51000257 806M51000258 806M51000259 806M51000260 806M51000261 806M51000262 806M51000263 806M51000264 806M51000265 806M51000266 806M51000267 806M51000268 806M51000269 806M51000270 806M51000271 806M51000272 806M51000273 806M51000274 807M51000275 807M51000276 807M51000277 807M51000278 807M51000279 807M51000280 807M51000281 807M51000282 807M51000283 807M51000284 807M51000285 807M51000286 807M51000287 807M51000288 807M51000289 807M51000290 807M51000291 807M51000292 807M51000293 807M51000294 807M51000295 807M51000296 807M51000297 807M51000298 807M51000299 807M51000300 807M51000301 807M51000303 807M51000304 807M51000305 807M51000306 807M51000307 807M51000308 807M51000309 807M51000310 807M51000311 807M51000313 807M51000314 807M51000315 807M51000316 807M51000317 807M51000318 807M51000319 807M51000320 807M51000321 807M51000322 807M51000323 807M51000324 808M51000325 808M51000326 808M51000327 808M51000328 808M51000329 808M51000330 808M51000331 808M51000332 808M51000333 808M51000334 808M51000335 808M51000336 808M51000337 808M51000338 808M51000339 808M51000340 808M51000341 808M51000342 808M51000343 808M51000344 808M51000345 808M51000346 808M51000347 808M51000348 808M51000349 810M51000350 810M51000351 810M51000352 810M51000353 810M51000354 810M51000355 810M51000356 810M51000357 810M51000358 810M51000359 810M51000360 810M51000361 810M51000362 810M51000363 810M51000364 810M51000365 810M51000366 810M51000367 810M51000368 810M51000369 810M51000370 810M51000371 810M51000372 810M51000373 810M51000374 811M51000375 811M51000376 811M51000377 811M51000379 811M51000380 811M51000381 811M51000382 811M51000383 811M51000384 811M51000385 811M51000386 811M51000387 811M51000388 811M51000389 811M51000390 811M51000391 811M51000392 811M51000393 811M51000394 811M51000395 811M51000396 811M51000397 811M51000398 811M51000399 813M51000400 813M51000401 813M51000402 813M51000403 813M51000404 813M51000405 813M51000406 813M51000407 813M51000408 813M51000409 813M51000410 813M51000411 813M51000412 813M51000413 813M51000414 813M51000415 813M51000416 813M51000417 813M51000418 813M51000419 814M51000420 814M51000421 814M51000422 814M51000423 814M51000424 814M51000425 814M51000426 814M51000427 814M51000428 814M51000429 814M51000430 814M51000431 814M51000432 814M51000433 814M51000434 814M51000435 814M51000436 814M51000437 814M51000438 814M51000439 817M51000440 817M51000441 817M51000442 817M51000443 817M51000444 817M51000445 817M51000446 817M51000447 817M51000448 817M51000449 817M51000450 817M51000451 817M51000452 817M51000453 817M51000454 817M51000455 817M51000456 817M51000457 817M51000458 817M51000459 818M51000460 818M51000461 818M51000462 818M51000463 818M51000464 818M51000465 818M51000466 818M51000467 818M51000468 818M51000469 818M51000471 818M51000472 818M51000473 818M51000474 818M51000475 818M51000476 818M51000477 818M51000478 818M51000479 820M51000480 820M51000481 820M51000482 820M51000483 820M51000484 820M51000485 820M51000486 820M51000487 820M51000488 820M51000489 820M51000490 820M51000491 820M51000492 820M51000494 820M51000495 820M51000496 820M51000497 820M51000499 821M51000500 821M51000501 821M51000502 821M51000503 821M51000504 821M51000505 821M51000506 821M51000507 821M51000508 821M51000509 821M51000510 821M51000511 821M51000513 821M51000514 821M51000515 821M51000516 821M51000517 821M51000518 821M51000519 823M51000520 823M51000521 823M51000522 823M51000523 823M51000524 823M51000525 823M51000526 823M51000527 823M51000528 823M51000529 823M51000530 823M51000531 823M51000532 823M51000533 823M51000534 823M51000535 823M51000536 823M51000537 823M51000538 823M51000539 823M51000540 823M51000541 823M51000542 823M51000543 823M51000544 823M51000545 823M51000546 823M51000547 823M51000548 823M51000549 823M51000550 823M51000551 823M51000552 823M51000553 823M51000554 825M51000555 825M51000556 825M51000557 825M51000558 825M51000559 825M51000560 825M51000561 825M51000563 825M51000564 825M51000565 825M51000566 825M51000567 825M51000568 825M51000569 825M51000570 825M51000571 825M51000572 825M51000573 825M51000574 827M51000575 827M51000576 827M51000577 827M51000578 827M51000579 827M51000580 827M51000581 827M51000582 827M51000583 827M51000584 827M51000585 827M51000586 827M51000587 827M51000588 827M51000589 827M51000590 827M51000591 827M51000592 827M51000593 827M51000594 830M51000595 830M51000596 830M51000597 830M51000598 830M51000599 830M51000600 830M51000601 830M51000602 830M51000603 830M51000604 830M51000605 830M51000606 830M51000607 830M51000608 830M51000609 830M51000610 830M51000611 830M51000612 830M51000613 830M51000614 830M51000615 830M51000616 830M51000617 830M51000618 830M51000619 830M51000620 830M51000621 830M51000622 830M51000623 830M51000624 830M51000625 830M51000628 830M51000630 830M51000631 830M51000632 830M51000634 830M51000636 830M51000637 830M51000638 832M51000626 832M51000627 832M51000629 832M51000633 832M51000635 832M51000639 832M51000640 832M51000641 832M51000642 832M51000643 832M51000644 832M51000650 832M51000654 834M51000645 834M51000646 834M51000647 834M51000648 834M51000649 834M51000652 834M51000653 834M51000655 834M51000656 834M51000657 834M51000658 834M51000659 834M561000651 835M51000660 835M51000661 835M51000662 835M51000663 835M51000664 835M51000665 835M51000666 835M51000667 835M51000668 835M51000669 835M51000670 835M51000671 835M51000672 835M51000673 835M51000674 835M51000676 835M51000677 835M51000678 835M51000679 835M51000680 835M51000681 835M51000682 835M51000683 835M51000685 835M51000686 835M51000687 835M51000688 835M51000689 835M51000690 835M51000691 835M51000692 835M51000693 835M51000694 836M51000695 836M51000696 836M51000697 836M51000698 836M51000699 836M51000701 836M51000702 836M51000703 836M51000704 836M51000705 836M51000706 836M51000707 836M51000708 836M51000709 838M51000710 838M51000711 838M51000712 838M51000713 838M51000714 838M51000715 838M51000716 838M51000717 838M51000718 838M51000719 838M51000720 838M51000721 838M51000722 838M51000723 838M51000724 838M51000725 838M51000726 838M51000727 838M51000728 838M51000729 838M51000730 838M51000731 838M51000732 838M51000733 838M51000734 838M51000735 838M51000736 838M51000737 838M51000738 838M51000739 841M51000740 841M51000741 841M51000742 841M51000743 841M51000744 841M51000745 841M51000746 841M51000747 841M51000748 841M51000749 841M51000750 841M51000751 841M51000752 841M51000753 841M51000754 842M51000755 842M51000756 842M51000757 842M51000758 842M51000759 842M51000760 842M51000761 842M51000762 842M51000763 842M51000764 842M51000765 842M51000766 842M51000767 842M51000768 842M51000769 842M51000770 842M51000771 842M51000772 842M51000773 842M51000774 842M51000775 842M51000776 842M51000777 842M51000778 842M51000779 842M51000780 842M51000781 842M51000782 842M51000783 842M51000784 843M51000785 843M51000786 843M51000787 843M51000788 843M51000789 843M51000790 843M51000791 843M51000792 843M51000793 843M51000794 843M51000795 843M51000796 843M51000797 843M51000798 843M51000799 845M51000800 845M51000801 845M51000802 845M51000803 845M51000804 845M51000806 845M51000807 845M51000808 845M51000809 845M51000810 845M51000811 845M51000812 845M51000813 845M51000814 846M05100817 846M51000815 846M51000816 846M51000818 846M51000819 846M51000820 846M51000821 846M51000823 846M51000824 846M51000825 846M51000827 846M51000828 846M51000829 846M51000830 846M51000831 846M51000832 847M51000833 847M51000834 847M51000836 847M51000837 847M51000838 847M51000839 847M51000840 847M51000841 847M51000842 847M51000843 847M51000844 847M51000845 847M51000846 847M51000847 849M51000848 849M51000849 849M51000851 849M51000852 849M51000853 849M51000854 849M51000855 849M51000856 849M51000857 849M51000858 849M51000859 849M51000860 849M51000861 849M51000862 850M51000863 850M51000864 850M51000865 850M51000866 850M51000867 850M51000868 850M51000869 850M51000870 850M51000871 850M51000872 850M51000873 850M51000874 850M51000875 850M51000876 850M51000877 851M51000878 851M51000879 851M51000880 851M51000881 851M51000882 851M51000883 851M51000884 851M51000885 851M51000886 851M51000887 851M51000888 851M51000889 851M51000890 851M51000891 851M51000892 853M51000893 853M51000894 853M51000895 853M51000896 853M51000897 853M51000898 853M51000899 853M51000900 853M51000901 853M51000902 902M51000903 902M51000904 902M51000905 902M51000906 902M51000907 902M51000908 902M51000909 902M51000910 902M51000911 902M51000912 902M51000913 902M51000914 902M51000915 902M51000916 902M51000918 902M51000919 902M51000920 902M51000921 902M51000922 904M51000923 904M51000924 904M51000925 904M51000926 904M51000927 904M51000928 904M51000929 904M51000930 904M51000931 904M51000932 904M51000933 904M51000934 904M51000936 904M51000937 904M51000938 904M51000939 904M51000940 904M51000941 904M51000942 905M51000943 905M51000944 905M51000945 905M51000946 905M51000947 905M51000948 905M51000949 905M51000950 905M51000951 905M51000952 905M51000953 905M51000954 905M51000955 905M51000956 905M51000957 905M51000958 905M51000959 905M51000960 905M51000961 905M51000962 906M51000963 906M51000964 906M51000965 906M51000966 906M51000967 906M51000968 906M51000969 906M51000970 906M51000971 906M51000972 906M51000973 906M51000975 906M51000976 906M51000977 906M51000978 906M51000979 906M51000980 906M51000981 906M51000982 907M51000983 907M51000984 907M51000985 907M51000986 907M51000987 907M51000988 907M51000989 907M51000990 907M51000991 907M51000992 907M51000993 907M51000994 907M51000995 907M51000996 907M51000997 907M51000999 907M51001000 907M51001001 907M51001002 907M51001017 909M51001003 909M51001004 909M51001005 909M51001006 909M51001007 909M51001008 909M51001009 909M51001010 909M51001011 909M51001012 909M51001013 909M51001014 909M51001015 909M51001018 909M51001019 909M51001020 909M51001021 909M51001022 910M51001003 910M51001004 910M51001005 910M51001006 910M51001007 910M51001008 910M51001009 910M51001010 910M51001011 911M51001023 911M51001024 911M51001025 911M51001026 911M51001027 911M51001028 911M51001029 911M51001030 911M51001031 911M51001032 911M51001034 911M51001035 911M51001036 911M51001037 911M51001038 911M51001039 911M51001040 911M51001041 911M51001042 911M51001043 911M51001044 911M51001045 912M51001046 912M51001047 912M51001048 912M51001049 912M51001050 912M51001051 912M51001052 912M51001053 912M51001054 912M51001055 912M51001056 912M51001057 912M51001058 912M51001059 912M51001060 912M51001061 912M51001062 912M51001063 912M51001064 912M51001065 912M51001066 912M51001067 912M51001068 912M51001069 912M51001070 913M51001071 913M51001072 913M51001073 913M51001074 913M51001075 913M51001076 913M51001077 913M51001078 913M51001079 913M51001080 913M51001081 913M51001082 913M51001083 913M51001084 913M51001085 913M51001086 913M51001087 913M51001088 913M51001089 913M51001090 913M51001091 913M51001092 913M51001093 913M51001094 913M51001095 913M51001096 913M51001097 913M51001098 913M51001099 913M51001100 913M51001101 914M51001102 914M51001103 914M51001104 914M51001105 914M51001106 914M51001107 914M51001108 914M51001109 914M51001110 914M51001111 914M51001112 914M51001113 914M51001114 914M51001115 914M51001116 914M51001118 914M51001119 914M51001120 914M51001121 915M51001122 915M51001123 915M51001124 915M51001125 915M51001126 915M51001127 915M51001129 915M51001130 915M51001131 915M51001132 915M51001133 915M51001134 915M51001135 915M51001136 915M51001137 916M51001138 916M51001139 916M51001140 916M51001141 916M51001142 916M51001143 916M51001144 916M51001145 916M51001146 916M51001147 916M51001148 916M51001149 916M51001150 916M51001151 916M51001152 916M51001153 916M51001154 916M51001155 916M51001156 917M51001158 917M51001160 917M51001161 917M51001162 917M51001163 917M51001164 917M51001165 917M51001166 917M51001167 917M51001168 917M51001169 917M51001170 917M51001172 917M51001173 918M51001174 918M51001175 918M51001176 918M51001177 918M51001178 918M51001179 918M51001180 918M51001182 918M51001183 918M51001184 918M51001185 919M51001186 919M51001187 919M51001188 919M51001189 919M51001190 919M51001191 919M51001192 919M51001193 919M51001194 919M51001195 919M51001196 919M51001197 919M51001198 919M51001199 919M51001200 919M51001201 919M51001202 919M51001203 919M51001204 919M51001205 920M51001206 920M51001207 920M51001208 920M51001209 920M51001210 920M51001211 920M51001212 920M51001213 920M51001214 920M51001215 920M51001216 920M51001217 920M51001218 920M51001219 920M51001220 921M51001221 921M51001222 921M51001223 921M51001224 921M51001225 921M51001226 921M51001227 921M51001228 921M51001229 921M51001230 921M51001231 921M51001232 921M51001233 921M51001234 921M51001235 923M51001236 923M51001237 923M51001238 923M51001239 923M51001240 923M51001241 923M51001242 923M51001243 923M51001244 923M51001245 923M51001246 923M51001247 923M51001248 923M51001249 923M51001250 923M51001251 923M51001252 924M51001253 924M51001254 924M51001255 924M51001256 924M51001257 924M51001258 924M51001259 924M51001260 924M51001261 924M51001262 924M51001263 924M51001264 924M51001265 924M51001266 924M51001267 924M51001268 925M51001269 925M51001270 925M51001271 925M51001272 925M51001274 925M51001275 925M51001276 925M51001277 925M51001278 925M51001279 925M51001280 925M51001281 925M51001282 925M51001283 925M51001284 925M51001285 925M51001286 925M51001287 925M51001288 925M51001289 925M51001290 925M51001291 925M51001292 925M51001293 926M51001294 926M51001295 926M51001296 926M51001297 926M51001298 926M51001299 926M51001300 926M51001301 926M51001302 926M51001303 926M51001304 926M51001305 926M51001306 926M51001307 926M51001308 926M51001309 926M51001310 926M51001311 926M51001312 926M51001313 926M51001314 926M51001315 926M51001316 926M51001317 926M51001318 927M51001319 927M51001320 927M51001321 927M51001322 927M51001323 927M51001324 927M51001325 927M51001326 927M51001327 927M51001328 927M51001329 927M51001330 928M51001331 928M51001332 928M51001333 928M51001334 928M51001335 928M51001336 928M51001337 928M51001338 928M51001339 928M51001340 928M51001341 928M51001342 928M51001343 928M51001344 928M51001345 929M51001346 929M51001347 929M51001348 929M51001349 929M51001350 929M51001351 929M51001352 929M51001353 929M51001354 929M51001355 929M51001356 929M51001357 929M51001358 929M51001359 929M51001360 930M51001361 930M51001362 930M51001363 930M51001364 930M51001365 930M51001366 930M51001367 930M51001368 930M51001369 930M51001370 930M51001371 930M51001372 930M51001373 930M51001374 930M51001375 931M51001376 931M51001377 931M51001378 931M51001379 931M51001380 931M51001381 931M51001382 931M51001383 931M51001384 931M51001385 931M51001386 931M51001387 932M51001388 932M51001389 932M51001390 932M51001391 932M51001392 932M51001393 932M51001394 932M51001395 932M51001396 932M51001397 932M51001398 932M51001399 932M51001400 932M51001401 932M51001402 933M51001403 933M51001404 933M51001405 933M51001406 933M51001407 933M51001408 933M51001409 933M51001410 933M51001411 934M51001412 934M51001413 934M51001414 934M51001415 934M51001416 934M51001417 934M51001418 934M51001419 934M51001420 934M51001421 934M51001422 934M51001423 934M51001424 934M51001425 934M51001426 935M51001427 935M51001428 935M51001429 935M51001430 935M51001431 935M51001432 935M51001433 935M51001434 935M51001435 935M51001436 935M51001437 935M51001438 935M51001439 935M51001440 936M51001442 936M51001443 936M51001444 936M51001446 936M51001447 936M51001448 936M51001449 936M51001450 936M51001451 936M51001452 936M51001453 936M51001454 936M51001455 936M51001456 938M51001457 938M51001458 938M51001459 938M51001460 938M51001461 938M51001462 938M51001463 938M51001464 938M51001465 939M51001466 939M51001467 939M51001468 940M51001469 940M51001470 940M51001471 940M51001472 940M51001473 940M51001474 940M51001475 940M51001477 940M51001478 940M51001479 940M51001480 940M51001481 940M51001482 940M51001483 941M51001484 941M51001485 941M51001486 941M51001487 941M51001488 941M51001489 941M51001490 941M51001491 941M51001492 941M51001493 941M51001494 941M51001495 941M51001496 941M51001497 941M51001498 942M51001499 942M51001500 942M51001501 942M51001502 942M51001503 942M51001504 942M51001505 942M51001506 942M51001507 942M51001508 942M51001509 942M51001510 943M51001511 943M51001512 943M51001513 943M51001514 943M51001515 943M51001516 943M51001517 943M51001518 943M51001519 943M51001520 943M51001521 943M51001522 943M51001523 943M51001524 943M51001525 944M51001526 944M51001527 944M51001528 944M51001529 944M51001530 944M51001531 944M51001532 944M51001533 944M51001534 944M51001535 944M51001536 944M51001537 945M51001538 945M51001539 945M51001540 945M51001541 945M51001542 945M51001543 945M51001544 945M51001545 945M51001546 945M51001547 945M51001548 945M51001549 945M51001550 945M51001552 946M51001553 946M51001554 946M51001555 946M51001556 946M51001557 946M51001558 946M51001559 946M51001560 946M51001561 946M51001562 946M51001563 947M51001564 947M51001565 947M51001566 947M51001567 947M51001568 947M51001569 947M51001570 947M51001571 947M51001572 947M51001573 947M51001574 947M51001575 947M51001576 947M51001577 947M51001578 947M51001579 947M51001580 947M51001581 947M51001582 949M51001583 949M51001584 949M51001585 949M51001586 949M51001587 949M51001588 949M51001589 949M51001590 949M51001591 950M51001592 950M51001593 950M51001594 950M51001595 950M51001596 950M51001597 950M51001598 950M51001599 950M51001600 950M51001601 950M51001602 950M51001603 950M51001604 950M51001605 950M51001606 950M51001607 950M51001608 950M51001609 950M51001610 950M51001611 950M51001612 950M51001613 950M51001614 950M51001615 950M51001616 950M51001617 950M51001618 950M51001619 950M51001620 950M51001621 950M51001622 951M51001623 951M51001624 951M51001625 951M51001626 951M51001627 951M51001628 951M51001629 951M51001630 951M51001631 951M51001632 951M51001633 951M51001634 951M51001635 951M51001636 951M51001637 951M51001638 951M51001639 951M51001640 952M51001641 952M51001642 952M51001643 952M51001644 952M51001645 952M51001646 952M51001647 952M51001648 952M51001649 952M51001650 952M51001651 952M51001652 952M51001653 952M51001654 952M51001655 953M51001656 953M51001657 953M51001658 953M51001659 953M51001660 953M51001661 953M51001662 953M51001663 953M51001664 953M51001665 953M51001666 953M51001667 953M51001668 953M51001669 953M51001670 1001M51001676 1001M51001677 1001M51001678 1001M51001679 1001M51001679 1001M51001680 1001M51001681 1001M51001682 1001M51001683 1001M51001684 1001M51001685 1001M51001686 1001M51001687 1001M51001688 1001M51001689 1001M51001690 1001M51001691 1001M51001692 1001M51001693 1001M51001694 1001M51001695 1001M51001696 1001M51001697 1001M51001698 1001M51001699 1001M51001700 1001M51001701 1001M51001702 1001M51001703 1001M51001704 1001M51001705 1001M51001706 1001M51001707 1001M51001708 1001M51001709 1001M51001710 1001M51001711 1001M51001712 1001M51001713 1001M51001714 1001M51001715 1001M51001716 1001M51001717 1001M51001718 1001M51001719 1001M51001720 1001M51001721 1001M51001722 1001M51001723 1001M51001724 1001M51001725 1001M51001726 1009M51001727 1009M51001728 1009M51001729 1009M51001730 1009M51001731 1009M51001732 1009M51001733 1009M51001734 1009M51001735 1009M51001736 1009M51001738 1009M51001739 1009M51001740 1009M51001741 1009M51001742 1009M51001743 1009M51001744 1009M51001745 1009M51001746 1009M51001747 1010M51001748 1010M51001749 1010M51001750 1010M51001751 1010M51001752 1010M51001753 1010M51001754 1010M51001755 1010M51001756 1010M51001757 1010M51001758 1010M51001759 1010M51001760 1010M51001761 1010M51001762 1010M51001763 1010M51001764 1010M51001765 1010M51001766 1010M51001767 1010M51001768 1010M51001769 1010M51001770 1010M51001771 1010M51001772 1011M51001773 1011M51001774 1011M51001775 1011M51001776 1011M51001777 1011M51001778 1011M51001779 1011M51001780 1011M51001781 1011M51001782 1011M51001783 1011M51001784 1011M51001785 1011M51001786 1011M51001787 1011M51001788 1011M51001789 1011M51001791 1011M51001792 1011M51001793 1011M51001794 1011M51001795 1011M51001796 1012M51001797 1012M51001798 1012M51001799 1012M51001800 1012M51001801 1012M51001802 1012M51001803 1012M51001804 1012M51001805 1012M51001806 1012M51001807 1012M51001808 1012M51001809 1012M51001810 1012M51001812 1012M51001813 1012M51001814 1012M51001815 1012M51001816 1012M51001817 1012M51001818 1012M51001819 1012M51001825 1012M51001827 1012M51001828 1012M51001829 1012M51001830 1012M51001833 1012M51001834 1012M51001835 1012M51001836 1012M51001837 1012M51001838 1012M51001839 1012M51001840 1012M51001841 1012M51001842 1012M51001843 1013M51001709 1013M51001811 1013M51001820 1013M51001821 1013M51001822 1013M51001823 1013M51001824 1013M51001826 1013M51001831 1013M51001832 1014M51001844 1014M51001845 1014M51001846 1014M51001847 1014M51001848 1014M51001849 1014M51001850 1014M51001851 1014M51001852 1014M51001853 1014M51001854 1014M51001855 1014M51001856 1014M51001857 1014M51001858 1014M51001859 1014M51001860 1014M51001862 1014M51001863 1014M51001864 1014M51001865 1014M51001866 1014M51001867 1014M51001868 1014M51001869 1014M51001870 1015M51001861 1015M51001871 1015M51001872 1015M51001873 1015M51001874 1015M51001875 1015M51001877 1015M51001878 1015M51001879 1015M51001880 1015M51001881 1015M51001882 1015M51001883 1015M51001884 1015M51001885 1015M51001886 1015M51001887 1015M51001888 1015M51001889 1015M51001890 1015M51001891 1015M51001892 1016M51001893 1016M51001894 1016M51001895 1016M51001896 1016M51001897 1016M51001898 1016M51001899 1016M51001900 1016M51001901 1016M51001902 1016M51001903 1016M51001904 1016M51001905 1016M51001906 1016M51001907 1016M51001908 1016M51001909 1016M51001910 1016M51001915 1016M51001916 1016M51001917 1016M51001919 1016M51001921 1016M51001922 1016M51001923 1016M51001924 1016M51001926 1016M51001927 1016M51001928 1016M51001929 1016M51001930 1016M51001931 1016M51001932 1016M51001933 1016M51001934 1016M51001935 1016M51001936 1016M51001937 1016M51001938 1016M51001939 1016M51001940 1016M51001941 1016M51001942 1016M51001943 1016M51001944 1016M51001945 1016M51001946 1017M51001911 1017M51001912 1017M51001913 1017M51001914 1017M51001918 1019M51001947 1019M51001948 1019M51001949 1019M51001950 1019M51001951 1019M51001952 1019M51001953 1019M51001954 1019M51001955 1019M51001956 1019M51001957 1019M51001958 1019M51001959 1019M51001960 1019M51001961 1019M51001962 1019M51001963 1019M51001964 1019M51001965 1019M51001966 1019M51001967 1019M51001968 1019M51001969 1020M51001970 1020M51001971 1020M51001973 1020M51001975 1020M51001976 1020M51001977 1020M51001978 1020M51001979 1020M51001980 1020M51001981 1020M51001982 1020M51001983 1020M51001984 1020M51001985 1020M51001986 1022M51001972 1022M51001974 1023M51001987 1023M51001988 1023M51001989 1023M51001990 1023M51001991 1023M51001992 1023M51001993 1023M51001994 1023M51001995 1023M51001996 1023M51001997 1023M51001998 1023M51001999 1023M51002000 1023M51002001 1023M51002002 1023M51002003 1023M51002004 1023M51002005 1023M51002006 1023M51002007 1023M51002008 1023M51002009 1023M51002010 1025M51002011 1025M51002012 1025M51002013 1025M51002014 1025M51002015 1025M51002016 1025M51002017 1025M51002018 1025M51002019 1025M51002020 1025M51002021 1025M51002022 1025M51002023 1025M51002024 1025M51002025 1025M51002026 1025M51002027 1025M51002028 1025M51002029 1025M51002030 1025M51002031 1025M51002032 1025M51002033 1025M51002034 1025M51002035 1025M51002036 1025M51002037 1025M51002038 1025M51002039 1025M51002040 1025M51002041 1025M51002042 1025M51002043 1025M51002044 1025M51002045 1025M51002046 1025M51002047 1025M51002048 1025M51002049 1025M51002050 1025M51002051 1025M51002052 1025M51002054 1025M51002055 1025M51002056 1025M51002057 1025M51002058 1025M51002059 1029M51002053 1029M51002060 1029M51002061 1029M51002062 1029M51002063 1029M51002064 1029M51002065 1029M51002066 1029M51002067 1029M51002068 1029M51002069 1029M51002070 1029M51002071 1029M51002072 1029M51002073 1029M51002074 1029M51002076 1029M51002077 1029M51002078 1029M51002079 1029M51002081 1029M51002082 1029M51002083 1030M51002075 1030M51002084 1030M51002085 1030M51002086 1030M51002087 1030M51002088 1030M51002089 1030M51002090 1030M51002091 1030M51002092 1030M51002093 1030M51002094 1030M51002095 1030M51002096 1030M51002097 1030M51002098 1030M51002100 1030M51002101 1030M51002102 1030M51002103 1030M51002104 1030M51002105 1030M51002106 1030M51002107 1030M51002108 1030M51002109 1030M51002110 1030M51002111 1032M51002112 1032M51002113 1032M51002114 1032M51002115 1032M51002116 1032M51002117 1032M51002118 1032M51002119 1032M51002120 1032M51002121 1032M51002122 1032M51002123 1032M51002124 1032M51002125 1032M51002126 1032M51002127 1032M51002128 1032M51002129 1032M51002130 1032M51002131 1032M51002132 1032M51002133 1032M51002134 1033M51002135 1033M51002136 1033M51002137 1033M51002138 1033M51002139 1033M51002140 1033M51002141 1033M51002142 1034M51002143 1034M51002144 1034M51002145 1034M51002146 1034M51002147 1034M51002148 1034M51002149 1034M51002150 1034M51002151 1034M51002152 1034M51002153 1035M51002154 1035M51002155 1035M51002156 1035M51002157 1035M51002158 1035M51002159 1035M51002160 1035M51002161 1035M51002162 1035M51002163 1035M51002164 1035M51002165 1035M51002166 1035M51002167 1035M51002168 1035M51002169 1035M51002170 1036M51002171 1036M51002172 1036M51002173 1036M51002174 1036M51002175 1036M51002176 1036M51002177 1036M51002178 1036M51002179 1036M51002180 1036M51002181 1036M51002182 1036M51002183 1036M51002184 1036M51002185 1037M51002186 1037M51002187 1037M51002188 1037M51002189 1037M51002190 1037M51002191 1037M51002192 1037M51002193 1037M51002194 1038M51002198 1038M51002199 1038M51002200 1038M51002201 1038M51002202 1038M51002203 1038M51002204 1038M51002205 1038M51002206 1038M51002207 1038M51002208 1038M51002209 1038M51002210 1038M51002211 1038M51002212 1038M51002213 1038M51002214 1038M51002215 1038M51002216 1038M51002217 1038M51002218 1038M51002219 1040M51002220 1040M51002221 1040M51002222 1040M51002223 1040M51002224 1040M51002225 1040M51002226 1040M51002227 1040M51002228 |
Recalling Firm/ Manufacturer |
StatSpin, Inc., dba Iris Sample Processing 60 Glacier Dr Westwood MA 02090-1800
|
For Additional Information Contact |
Gail MaGrath 781-551-0036
|
Manufacturer Reason for Recall |
Rotors may fail due to wire spokes breaking
|
FDA Determined Cause 2 |
Device Design |
Action |
Iris notified customers by letter in August 2010. The letter identified the product, the problem, and the action to be taken by the customer. Customers were requested to install the new rotors and discard the older rotors with the wires. There was no need for customers to return the rotors, but the firm requested customers fax or e-mail the enclosed form to verify that they have installed the new rotors.
For any questions customers should call 1-800-782-8774. |
Quantity in Commerce |
2090 units |
Distribution |
Worldwide Distribution - USA (nationwide) and the countries of Canada, Australia Belgium, Czech Republic, Finland, France, Greece, Kuwait, Norway, Turkey,UK, and Norway. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|