| Class 2 Device Recall Cylindrical Reamer |  |
Date Initiated by Firm | December 23, 2010 |
Date Posted | March 04, 2011 |
Recall Status1 |
Terminated 3 on October 01, 2012 |
Recall Number | Z-1545-2011 |
Recall Event ID |
57605 |
Product Classification |
Reamer - Product Code HTO
|
Product | Cylindrical Bullet Nose Reamer 19.0 mm x 250 mm, REF 475826, Biomet Orthopedics 66 East Bell Drive Warsaw IN.
The product is used to prepare the femoral canal for stem implantation. |
Code Information |
096540 |
Recalling Firm/ Manufacturer |
Biomet, Inc. 56 E Bell Dr Warsaw IN 46582-6989
|
For Additional Information Contact | 574-267-6639 Ext. 1676 |
Manufacturer Reason for Recall | The firm was notified by their supplier that during manufacture of this instrument, the 12 degree cutting edge was not created. Testing found that if the cutting edge is missing, the reamer will not perform as intended. |
FDA Determined Cause 2 | Other |
Action | The firm, BIOMET, sent an" URGENT MEDICAL DEVICE RECALL NOTICE" dated December 23, 2010 to its customers. The notice described the product, problem and actions to be taken by the customer. The customers were instructed to immediately locate, discontinue use of product and return the product to Biomet. Biomet distributors were told to notify hospital personnel of the recall if they had distributed the product to hospital customers. The Customers/Distributors were instructed to confirm receipt of the notice by calling 800-348-9500 extension 3755 and 3983 and to complete and return the Response Form included with the recall notice via fax to 574-372-1683.
Questions related to this notice should be directed to 574-371-3755 or 574-372-3983, Monday through Friday, 8am to 5pm. |
Quantity in Commerce | 51 all products |
Distribution | Worldwide distribution: USA including states of: FL, OH, PA, IN, CA, and NJ; and countries of Argentina, Canada, Mexico, Netherlands and Spain. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|