Class 2 Device Recall Leksell GammaPlan

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Leksell GammaPlansee related information
Date Initiated by FirmOctober 01, 2005
Date PostedFebruary 14, 2011
Recall Status1 Terminated 3 on March 02, 2011
Recall NumberZ-1223-2011
Recall Event ID 57607
510(K)NumberK973441 
Product Classification Radionuclide Radiation Therapy System - Product Code IWB
ProductLeksell GammaPlan, Model 5.34. Intended to be used for planning the dosimetry of treatments in stereotactic radiation therapy.
Code Information SG33620363, US68537171, SG10821625, 3713G01281, SG03321319, SG33520213, US19502892, C160, SG0320704, SG44120295, 00000000000002, SG31630678, SG24820212, SG24820194, SG44221038, SG34620630, SG40920412, SG31020878, A4986A, SG11520160, SG44720158, SG41720310, SG43820213, 3416G09539, SG24820139, US19257221, SG33820630, 19476882, SG13320399H, SG40520126, SG50425261, SG41720317, US66320311, US19490144, SG21220130, US68154038, US19501776, SW ONLY, SG32520836, 3416G08464, SG21220113, SG44925638, SG04120289, SG51220179, SG2030679, US6824174, SG22920793, SG4820232, 3628G06513, 000000000000019, 00000000000004, US19310946, 3416G05185, US69131975, SG42520530, SG23320693, US68104248, SG23720396, SG24820231, SG14720637, SG43520759, US68513855, SG44820136, SG20220701, SG44120292, US19476874, US6863578, SG04020719, 3416G07810, SG43720660, SG44925635, US68538435, US19322820, SG33320960, US19319754, SG31620624, US10102906, SG44925637, SG43520477, US69172955, SG42720416, 3711C04231, US68154724, SG0620611, SG20220192, SG33620337, US19311023, SG35020182, US19239028, SG50425266, SG51920134, 000000000000011, SG44925633, SG24021335, 0304011939, A4986A, SG34220901, SG1220108, SG32120124, US69132547, SG43720550, US19476879, 6146A06124, SG41920535, 6412A00007, US68500234, US19322145, SG20220694, US68115508, SG41620718, SG10720353, 000000000000014, SG43520485, SG42820297, 000000000000015, SG41020267, US68104599, IE24488098, SG11421969, SG41720337, US6836799, US19344670, SG11720734, US68064889, SG20220679, 000000000000016, SG20220699, SG11421978, SG34220707, US19378673, SG34420200, SG20220143, SG50425271, US10025765, US90345020, US19345857, and SG13420315.  
Recalling Firm/
Manufacturer
Elekta, Inc.
4775 Peachtree Industrial Blvd
Bldg 300, #300
Norcross GA 30092-3011
For Additional Information ContactThomas Valentine
770-670-2548
Manufacturer Reason
for Recall
Investigation found that if a user accidentally selects the wrong image/tube position images can be displayed flipped.
FDA Determined
Cause 2
Other
ActionProduct bulletin Leksell GammaPlan - "Leksell GammaPlan version 5.34" (dated October 2005) was sent to affected Customers as notification of the problem. Field Change Order FCO-DCO-05-0006 (dated November 22, 2005) and Technical Notes Leksell GammaPlan Leksell SurgiPlan Important Notice - "Potential issue using flipped angiographic images" (dated October 2005) was sent to Elekta Service Engineers as notification of the problem. The letters described the issue and the FCO asked the Service Engineers to confirm their receipt of the documentation. The Technical Notes also discussed image visualization requirements and listed recommendations.
Quantity in Commerce272 units
DistributionWorldwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = IWB
-
-