| Class 2 Device Recall Leksell GammaPlan | |
Date Initiated by Firm | October 01, 2005 |
Date Posted | February 14, 2011 |
Recall Status1 |
Terminated 3 on March 02, 2011 |
Recall Number | Z-1223-2011 |
Recall Event ID |
57607 |
510(K)Number | K973441 |
Product Classification |
Radionuclide Radiation Therapy System - Product Code IWB
|
Product | Leksell GammaPlan, Model 5.34.
Intended to be used for planning the dosimetry of treatments in stereotactic radiation therapy. |
Code Information |
SG33620363, US68537171, SG10821625, 3713G01281, SG03321319, SG33520213, US19502892, C160, SG0320704, SG44120295, 00000000000002, SG31630678, SG24820212, SG24820194, SG44221038, SG34620630, SG40920412, SG31020878, A4986A, SG11520160, SG44720158, SG41720310, SG43820213, 3416G09539, SG24820139, US19257221, SG33820630, 19476882, SG13320399H, SG40520126, SG50425261, SG41720317, US66320311, US19490144, SG21220130, US68154038, US19501776, SW ONLY, SG32520836, 3416G08464, SG21220113, SG44925638, SG04120289, SG51220179, SG2030679, US6824174, SG22920793, SG4820232, 3628G06513, 000000000000019, 00000000000004, US19310946, 3416G05185, US69131975, SG42520530, SG23320693, US68104248, SG23720396, SG24820231, SG14720637, SG43520759, US68513855, SG44820136, SG20220701, SG44120292, US19476874, US6863578, SG04020719, 3416G07810, SG43720660, SG44925635, US68538435, US19322820, SG33320960, US19319754, SG31620624, US10102906, SG44925637, SG43520477, US69172955, SG42720416, 3711C04231, US68154724, SG0620611, SG20220192, SG33620337, US19311023, SG35020182, US19239028, SG50425266, SG51920134, 000000000000011, SG44925633, SG24021335, 0304011939, A4986A, SG34220901, SG1220108, SG32120124, US69132547, SG43720550, US19476879, 6146A06124, SG41920535, 6412A00007, US68500234, US19322145, SG20220694, US68115508, SG41620718, SG10720353, 000000000000014, SG43520485, SG42820297, 000000000000015, SG41020267, US68104599, IE24488098, SG11421969, SG41720337, US6836799, US19344670, SG11720734, US68064889, SG20220679, 000000000000016, SG20220699, SG11421978, SG34220707, US19378673, SG34420200, SG20220143, SG50425271, US10025765, US90345020, US19345857, and SG13420315. |
Recalling Firm/ Manufacturer |
Elekta, Inc. 4775 Peachtree Industrial Blvd Bldg 300, #300 Norcross GA 30092-3011
|
For Additional Information Contact | Thomas Valentine 770-670-2548 |
Manufacturer Reason for Recall | Investigation found that if a user accidentally selects the wrong image/tube position images can be displayed flipped. |
FDA Determined Cause 2 | Other |
Action | Product bulletin Leksell GammaPlan - "Leksell GammaPlan version 5.34" (dated October 2005) was sent to affected Customers as notification of the problem. Field Change Order FCO-DCO-05-0006 (dated November 22, 2005) and Technical Notes Leksell GammaPlan Leksell SurgiPlan Important Notice - "Potential issue using flipped angiographic images" (dated October 2005) was sent to Elekta Service Engineers as notification of the problem. The letters described the issue and the FCO asked the Service Engineers to confirm their receipt of the documentation. The Technical Notes also discussed image visualization requirements and listed recommendations. |
Quantity in Commerce | 272 units |
Distribution | Worldwide Distribution |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = IWB
|
|
|
|