Class 2 Device Recall Medstorm Adult Radiolucent Electrode

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Medstorm Adult Radiolucent Electrodesee related information
Date Initiated by FirmDecember 17, 2010
Date PostedFebruary 18, 2011
Recall Status1 Terminated 3 on July 13, 2011
Recall NumberZ-1372-2011
Recall Event ID 57736
510(K)NumberK080421 
Product Classification Electrode, electrocardiograph, multi-function - Product Code MLN
ProductMedstorm Adult Radiolucent Zoll Medical (Part # 16381, HeartSync, Ann Arbor, MI. The Adult Radiolucent Multifunction Tin Electrodes are indicated for use in external pacing, defibrillation and monitoring applications as a non-sterile, disposable device for single patient use only. The electrodes provide the conductive interface between the defibrillator and/or the external transcutaneous cardiac pacemaker and the patients skin. When a patient requires defibrillation, cardioversion or external pacing, these electrodes will be applied to the patient and connected to the respective low power mono-phasic or bi-phasic defibrillator (360 joules maximum) or external pacemaker.
Code Information Lot # Y032709-1
FEI Number 1317188
Recalling Firm/
Manufacturer
Heart Sync, Inc
5643 Plymouth Rd
Ann Arbor MI 48105-9586
For Additional Information ContactStephen Shulman
800-828-4681
Manufacturer Reason
for Recall
The seal on the electrode pouch may be open, exposing the electrode to the outside environment. The package does have a caution statement that states, " Do not open package until immediately prior to using electrodes. and Electrodes may dry out when removed from packaging and exposed to air.
FDA Determined
Cause 2
Packaging
ActionHeart Sync, Inc sent an URGENT MEDICAL DEVICE RECALL letter dated December 23, 2010, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to: Examine their inventory and quarantine product that was subject to the recall that was not sealed properly. Contact any customers if they further distributed the product and notify them at once of this product recall. Contact the firm if they had any unused product. The firm would issue a UPS Call Tag to have the product returned to them and the firm would send replacement product as soon as they were notified. Complete and return the Recall Return Response Form to the firm in the attached prepaid envelope. For any questions regarding this recall call 734-213-5530.
Quantity in Commerce300
DistributionWorldwide Distribution - USA including AK, AZ, CA, FL, IL, KS, LA, MA, ND, NM, OH, TX, WA, WI and WV; and Canada
Total Product Life CycleTPLC Device Report