|
Class 1 Device Recall Ambulatory Infusion Pump |
|
Date Initiated by Firm |
February 28, 2011 |
Date Posted |
April 05, 2011 |
Recall Status1 |
Terminated 3 on August 06, 2012 |
Recall Number |
Z-1870-2011 |
Recall Event ID |
57939 |
510(K)Number |
K981816
|
Product Classification |
Pump, infusion - Product Code FRN
|
Product |
Curlin PainSmart IOD Ambulatory Infusion Pump
Provides a means for the volumetric delivery of fluids used in Parenteral, Enteral, Epidural, Subcutaneous and Intravenous applications. |
Code Information |
200016 200017 200030 200035 200038 200041 200043 200044 200049 200050 200052 200054 200055 200057 200059 200072 200082 200084 200087 200099 200102 200104 200106 200112 200131 200140 200141 200145 200149 200150 200151 200152 200163 200165 200169 200170 200175 200175 200177 200178 200188 200191 200192 200193 200194 200195 200196 200197 200202 200203 200204 200205 200206 200207 200208 200209 200214 200215 200217 200218 200219 200220 200221 200222 200225 200226 200227 200228 200229 200230 200231 200232 200236 200237 200239 200240 200241 200242 200243 200244 200248 200250 200253 200254 200255 200255 200257 200258 200263 200266 200267 200269 200270 200273 200275 200276 200280 200280 200281 200282 200287 200288 200289 200290 200292 200292 200294 200295 200296 200297 200298 200299 200307 200309 200311 200312 200313 200314 200315 200316 200319 200320 200321 200322 200323 200324 200325 200326 200329 200330 200331 200332 200333 200334 200335 200336 200340 200341 200348 200349 200350 200351 200352 200353 200358 200359 200360 200361 200362 200363 200364 200366 200369 200370 200371 200372 200373 200374 200375 200376 200380 200381 200382 200383 200384 200385 200386 200387 200390 200391 200392 200393 200394 200395 200396 200397 200400 200401 200402 200403 200404 200405 200406 200407 200411 200412 200413 200414 200415 200416 200417 200419 200422 200423 200424 200426 200427 200428 200429 200430 200433 200434 200435 200436 200437 200438 200439 200440 200444 200445 200446 200447 200448 200449 200450 200452 200455 200457 200458 200459 200460 200461 200462 200463 200466 200467 200468 200469 200470 200471 200472 200473 200476 200477 200478 200479 200480 200481 200482 200483 200486 200487 200489 200490 200491 200536 200537 200538 200563 200564 200565 200566 200567 200568 200572 200573 200576 200577 200578 200579 200580 200582 200583 200584 200587 200588 200589 200590 200591 200592 200594 200595 200598 200599 200600 200602 200603 200604 200606 200607 200610 200611 200612 200613 200614 200615 200616 200617 200620 200621 200622 200623 200624 200625 200626 200627 200631 200632 200637 200638 200639 200640 200642 200643 200725 200769 200844 200852 200878 200898 200936 200943 200947 200948 200950 200951 200952 200953 200956 200958 200962 200964 200965 200966 200967 200968 200968 200969 200974 200977 200978 200980 200981 200982 200983 200984 200989 200991 200995 200997 201001 201003 201004 201008 201025 201029 201030 201032 201033 201033 201037 201043 201059 201061 201070 201073 201074 201081 201082 201084 201090 201092 201096 201099 201100 201102 201104 201105 201110 201112 201116 201117 201118 201119 201121 201122 201126 201127 201129 201131 201132 201133 201134 201137 201142 201144 201145 201147 201148 201154 201157 201158 201162 201166 201172 201173 201177 201179 201180 201181 201185 201189 201190 201191 201192 201193 201194 201195 201198 201200 201201 201202 201203 201204 201205 201206 201212 201214 201215 201216 201217 201219 201221 201222 201225 201226 201228 201230 201231 201234 201235 201236 201240 201242 201244 201246 201247 201248 201250 201251 201254 201255 201256 201257 201258 201261 201263 201264 201268 201269 201272 201272 201274 201275 201276 201277 201280 201281 201282 201283 201287 201289 201301 201302 201305 201306 201307 201308 201309 201310 201311 201312 201315 201316 201317 201318 201319 201320 201321 201322 201326 201327 201328 201329 201331 201332 201333 201335 201338 201339 201340 201341 201342 201343 201344 201346 201351 201352 201353 201354 201355 201375 201376 201377 201381 201382 201383 201384 201385 201386 201387 201388 201391 201392 201393 201394 201395 201396 201397 201398 201401 201402 201403 201404 201405 201415 201416 201418 201421 201422 201423 201424 201425 201426 201443 201444 201449 201462 201464 201469 201471 201474 201475 201477 201481 201482 201483 201485 201486 201487 201488 201489 201494 201495 201496 201497 201498 201499 201500 201501 201506 201507 201508 201509 201510 201511 201513 201514 201519 201520 201521 201522 201523 201524 201525 201526 201529 201530 201531 201532 201533 201534 201535 201537 201540 201541 201542 201543 201544 201545 201546 201547 201551 201552 201553 201554 201555 201556 201557 201558 201562 201563 201564 201566 201568 201569 201570 201571 201574 201575 201576 201577 201578 201579 201581 201582 201585 201586 201587 201588 201589 201590 201592 201593 201597 201598 201599 201600 201601 201602 201603 201604 201607 201608 201609 201610 201611 201612 201613 201614 201617 201618 201619 201620 201621 201622 201623 201625 201628 201629 201630 201631 201634 201635 201636 201637 201641 201642 201643 201644 201645 201646 201647 201648 201651 201652 201653 201654 201655 201656 201657 201658 201661 201662 201663 201664 201665 201666 201668 201670 201673 201675 201676 201677 201680 201682 201686 201688 201697 201701 201702 201703 201704 201706 201707 201709 201768 201809 201811 201818 201820 201821 201821 201821 201826 201826 201826 201827 201828 201830 201840 201841 201844 201845 201846 201847 201848 201849 201850 201851 201854 201855 201856 201857 201858 201859 201865 201867 201871 201872 201873 201874 201875 201876 201877 201878 201895 201901 201909 201910 201911 201913 201915 201916 201922 201923 201924 201925 201928 201932 201933 201937 201940 201943 201944 201945 201946 201947 201949 201950 201957 201958 201959 201960 201961 201962 201963 201964 201967 201968 201969 201970 201971 201972 201973 201974 201977 201978 201979 201980 201981 201982 201983 201984 201987 201989 201990 201991 201992 201997 201998 201999 202002 202003 202004 202006 202007 202011 202012 202013 202016 202017 202018 202019 202020 202021 202022 202023 202026 202027 202028 202029 202030 202031 202032 202034 202039 202041 202047 202048 202049 202050 202051 202052 202057 202058 202059 202071 202072 202073 202074 202075 202093 202094 202095 202096 202121 202133 202136 202138 202145 202149 202153 202166 202187 202189 202190 202191 202194 202225 202227 202229 202230 202231 202235 202242 202247 202249 202250 202251 202252 202253 202254 202255 202258 202259 202260 202261 202262 202263 202265 202266 202270 202271 202272 202287 202288 202289 202290 202291 202295 202296 202297 202298 202299 202300 202302 202303 202307 202308 202309 202310 202311 202312 202313 202314 202317 202318 202319 202320 202321 202324 202325 202366 202378 202379 202379 202383 202384 202385 202389 202391 202408 202409 202411 202415 202416 202417 202418 202421 202422 202426 202429 202430 202438 202438 202439 202456 202462 202483 202494 202503 202508 202510 202512 202513 202523 202526 202527 202529 202533 202535 202536 202539 202541 202543 202544 202545 202554 202556 202557 202558 202563 202564 202571 202572 202576 202577 202580 202587 202598 202599 202600 202604 202605 202609 202619 202621 202654 202656 202660 202664 202665 202666 202673 202682 202688 202690 202696 202698 202701 202702 202704 202705 202714 202715 202716 202723 202729 202730 202731 202732 202737 202738 202739 202740 202741 202742 202743 202744 202747 202748 202749 202750 202751 202753 202755 202756 202761 202763 202764 202766 202769 202771 202775 202778 202785 202788 202792 202793 202797 202798 202802 202804 202824 202825 202826 202827 202829 202835 202836 202838 202844 202852 202853 202854 202856 202857 202858 202859 202862 202863 202864 202865 202866 202867 202868 202869 202873 202874 202875 202877 202878 202879 202880 202881 202884 202885 202886 202887 202888 202889 202890 202892 202902 202903 202904 202905 202907 202908 202909 202910 202913 202914 202915 202917 202921 202922 202923 202924 202927 202943 202947 202948 202949 202950 202952 202953 202956 202957 202958 202959 202960 202962 202963 202964 202967 202968 202969 202973 202974 202976 202977 202979 202984 202985 202986 202993 202994 202998 203000 203001 203006 203012 203013 203014 203015 203016 203017 203018 203022 203024 203028 203029 203030 203031 203032 203033 203037 203038 203039 203040 203042 203065 203068 203078 203082 203084 203087 203088 203090 203095 203096 203097 203109 203110 203111 203112 203113 203114 203115 203116 203121 203122 203123 203124 203125 203126 203127 203130 203134 203139 203143 203146 203147 203153 203158 203163 203171 203174 203175 203176 203182 203184 203187 203189 203194 203195 203196 203197 203199 203202 203204 203205 203215 203218 203221 203222 203223 203224 203225 203226 203234 203235 203236 203237 203238 203240 203241 203242 203245 203246 203247 203248 203249 203250 203252 203255 203276 203281 203282 203283 203284 203286 203287 203288 203296 203297 203300 203301 203302 203304 203306 203307 203311 203312 203313 203314 203318 203319 203322 203323 203327 203331 203332 203336 203337 203339 203341 203342 203345 203348 203350 203352 203353 203354 203355 203356 203359 203362 203363 203365 203367 203368 203369 203370 203381 203383 203384 203385 203386 203387 203388 203390 203396 203397 203399 203400 203401 203402 203403 203404 203408 203409 203410 203411 203412 203420 203422 203425 203435 203436 203438 203439 203440 203441 203442 203443 203446 203447 203448 203450 203451 203452 203453 203454 203457 203458 203459 203460 203461 203462 203463 203464 203467 203468 203469 203470 203471 203472 203474 203475 203479 203480 203481 203482 203483 203484 203485 203486 203490 203491 203492 203493 203494 203495 203496 203497 203500 203501 203502 203503 203504 203506 203507 203508 203512 203513 203517 203518 203519 203521 203524 203525 203528 203529 203530 203531 203532 203533 203534 203535 203538 203539 203541 203542 203543 203544 203545 203546 203549 203550 203551 203552 203553 203554 203555 203556 203560 203561 203562 203563 203564 203565 203568 203569 203572 203573 203574 203577 203578 203580 203582 203583 203587 203590 203591 203592 203593 203594 203595 203596 203600 203601 203602 203604 203605 203606 203607 203608 203612 203617 203620 203623 203625 203626 203627 203628 203631 203632 203633 203634 203635 203637 203638 203639 203644 203645 203651 203663 203664 203665 203667 203668 203674 203675 203680 203682 203683 203684 203685 203687 203690 203691 203692 203694 203695 203697 203698 203699 203702 203703 203704 203705 203706 203707 203708 203709 203713 203714 203715 203716 203717 203718 203719 203721 203726 203727 203729 203730 203732 203733 203734 203735 203738 203739 203740 203741 203742 203743 203744 203745 203748 203749 203750 203751 203752 203753 203756 203757 203760 203761 203762 203763 203765 203766 203767 203768 203771 203772 203773 203774 203775 203777 203778 203779 203782 203783 203784 203785 203786 203787 203788 203789 203792 203793 203794 203795 203798 203799 203800 203801 203809 203810 203811 203812 203813 203814 203815 203816 203819 203823 203828 203832 203833 203837 203846 203848 203851 203852 203853 203854 203855 203856 203857 203858 203861 203862 203863 203864 203865 203867 203869 203871 203874 203875 203876 203877 203878 203879 203881 203883 203891 203892 203893 203894 203895 203896 203897 203898 203902 203903 203904 203905 203907 203909 203927 203929 203943 203944 203945 203946 203948 203949 203950 203951 203956 203957 203958 203959 203961 203964 203966 203967 203970 203971 203973 203974 203975 203977 203979 203980 203983 203986 203989 203990 203991 203994 203995 203996 203999 204001 204002 204004 204005 204006 204007 204008 204011 204013 204016 204018 204020 204029 204031 204033 204042 204043 204044 204045 204046 204047 204048 204049 204052 204053 204054 204055 204056 204057 204058 204059 204062 204063 204064 204065 204066 204067 204068 204069 204072 204073 204074 204076 204078 204079 204080 204081 204084 204085 204088 204089 204090 204091 204093 204094 204101 204102 204104 204105 204106 204107 204108 204109 204113 204114 204115 204117 204118 204119 204120 204121 204124 204126 204127 204128 204129 204130 204131 204132 204135 204136 204137 204138 204139 204140 204141 204142 204145 204146 204147 204148 204149 204150 204153 204154 204158 204159 204160 204161 204162 204163 204164 204165 204169 204170 204171 204172 204173 204175 204177 204178 204182 204183 204184 204185 204186 204187 204188 204189 204192 204193 204194 204196 204198 204199 204200 204201 204204 204205 204206 204209 204210 204211 204212 204213 204222 204223 204224 204225 204226 204227 204228 204229 204233 204234 204235 204236 204237 204238 204239 204241 204245 204246 204248 204249 204250 204251 204252 204254 204259 204261 204262 204264 204267 204269 204271 204272 204275 204276 204277 204278 204279 204281 204282 204286 204292 204295 204297 204298 204300 204301 204302 204304 204309 204311 204312 204313 204315 204316 204318 204319 204322 204323 204325 204326 204328 204329 204330 204331 204334 204335 204336 204337 204338 204339 204340 204341 204344 204345 204346 204347 204348 204351 204354 204355 204358 204359 204360 204361 204362 204368 204369 204371 204375 204376 204377 204378 204379 204380 204381 204382 204386 204387 204389 204390 204391 204392 204393 204395 204399 204400 204401 204402 204403 204404 204405 204406 204410 204412 204413 204417 204418 204419 204420 204421 204424 204425 204426 204427 204429 204430 204431 204432 204435 204436 204437 204438 204439 204441 204442 204443 204447 204448 204449 204450 204451 204452 204453 204454 204458 204459 204460 204461 204462 204464 204465 204466 204470 204471 204472 204474 204476 204477 204478 204479 204485 204486 204490 204491 204492 204493 204494 204497 204500 204501 204502 204503 204504 204505 204506 204507 204510 204512 204513 204514 204515 204516 204518 204519 204523 204524 204525 204526 204527 204528 204531 204532 204536 204537 204539 204540 204541 204542 204543 204544 204547 204548 204549 204550 204551 204552 204553 204554 204558 204560 204561 204562 204563 204565 204566 204567 204572 204574 204575 204576 204578 204580 204585 204586 204592 204594 204595 204596 204597 204598 204599 204600 204604 204607 204614 204619 204621 204622 204623 204626 204629 204631 204632 204633 204634 204635 204637 204639 204644 204646 204650 204651 204652 204653 204654 204655 204659 204660 204661 204663 204665 204671 204672 204675 204683 204684 204685 204689 204691 204693 204697 204698 204715 204719 204720 204721 204729 204735 204740 204741 204750 204753 204755 204756 204757 204761 204761 204762 204777 204784 204789 204790 204794 204798 204799 204806 204819 204821 204822 204823 204824 204826 204827 204828 204831 204832 204833 204834 204835 204836 204839 204842 204847 204848 204849 204852 204855 204857 204858 204860 204863 204864 204865 204866 204867 204869 204871 204874 204880 204881 204882 204884 204885 204886 204887 204888 204891 204892 204893 204894 204895 204898 204899 204901 204905 204907 204908 204909 204911 204912 204915 204916 204920 204921 204922 204923 204924 204925 204927 204928 204938 204939 204940 204942 204943 204945 204947 204948 204953 204954 204957 204958 204960 204962 204965 204967 204973 204975 204983 204986 204987 204988 204990 204995 205003 205004 205005 205006 205009 205013 205015 205017 205027 205028 205029 205030 205032 205034 205035 205036 205041 205042 205043 205044 205046 205047 205048 205049 205056 205059 205061 205062 205063 205064 205065 205067 205071 205073 205074 205075 205076 205081 205082 205085 205089 205090 205091 205092 205093 205094 205095 205096 205099 205100 205104 205105 205106 205107 205108 205110 205115 205116 205121 205126 205127 205128 205129 205130 205133 205134 205135 205136 205137 205139 205140 205141 205144 205145 205146 205147 205148 205149 205150 205151 205154 205155 205156 205157 205158 205159 205160 205161 205164 205165 205166 205167 205168 205169 205171 205172 205175 205176 205177 205178 205179 205180 205181 205182 205186 205187 205188 205189 205190 205191 205192 205193 205196 205197 205198 205199 205200 205201 205202 205203 205206 205207 205208 205209 205210 205211 205212 205213 205216 205217 205218 205219 205220 205221 205222 205223 205226 205228 205229 205230 205231 205232 205236 205241 205244 205247 205249 205252 205253 205254 205255 205256 205261 205263 205264 205268 205272 205275 205276 205279 205292 205293 205294 205298 205299 205300 205302 205304 205313 205315 205320 205321 205322 205323 205324 205325 205328 205329 205330 205331 205332 205333 205334 205335 205338 205339 205340 205341 205342 205344 205345 205347 205350 205352 205354 205359 205363 205365 205366 205367 205372 205377 205379 205382 205385 205388 205390 205395 205402 205403 205409 205411 205413 205414 205417 205425 205430 205438 205439 205443 205449 205450 205451 205452 205455 205456 205457 205458 205459 205460 205461 205463 205466 205467 205470 205471 205472 205473 205474 205475 205478 205479 205480 205481 205482 205483 205484 205485 205488 205489 205490 205491 205492 205493 205495 205496 205499 205500 205501 205502 205503 205504 205505 205506 205509 205510 205511 205512 205513 205515 205516 205517 205520 205521 205522 205523 205524 205525 205526 205527 205530 205532 205533 205534 205535 205536 205537 205538 205541 205542 205543 205544 205545 205546 205547 205548 205551 205552 205553 205555 205556 205557 205558 205559 205562 205563 205564 205565 205566 205567 205568 205569 205572 205573 205574 205575 205576 205577 205578 205579 205582 205584 205585 205586 205587 205588 205589 205590 205593 205594 205595 205596 205597 205598 205599 205600 205603 205604 205605 205606 205607 205608 205609 205610 205613 205614 205615 205616 205617 205618 205619 205620 205623 205624 205625 205626 205627 205628 205630 205632 205645 205652 205663 205664 205665 205666 205671 205672 205675 205676 205679 205680 205681 205682 205683 205684 205688 205689 205690 205692 205693 205694 205695 205696 205699 205701 205704 205705 205707 205708 205709 205711 205716 205717 205719 205720 205721 205723 205725 205726 205730 205731 205732 205733 205734 205735 205742 205743 205748 205749 205750 205751 205752 205754 205755 205756 205759 205760 205761 205761 205763 205764 205765 205766 205769 205770 205771 205772 205773 205774 205775 205776 205779 205780 205781 205782 205783 205784 205785 205786 205789 205790 205791 205792 205793 205795 205796 205797 205801 205802 205803 205804 205805 205806 205807 205808 205811 205812 205813 205814 205815 205816 205817 205818 205821 205822 205823 205824 205825 205826 205827 205834 205837 205838 205839 205841 205842 205843 205844 205845 205856 205857 205859 205860 205862 205869 205870 205871 205874 205875 205876 205878 205879 205880 205884 205885 205900 205901 205902 205903 205904 205906 205907 205908 205912 205913 205914 205920 205921 205922 205923 205927 205934 205936 205938 205939 205940 205941 205942 205943 205946 205947 205948 205949 205950 205953 205954 205956 205960 205961 205962 205963 205964 205965 205966 205967 205970 205971 205972 205975 205976 205978 205979 205980 205983 205984 205985 205986 205987 205988 205990 205991 205994 205996 205997 205998 205999 206008 206009 206010 206013 206015 206016 206017 206018 206021 206022 206023 206027 206028 206029 206030 206031 206032 206033 206034 206037 206038 206039 206040 206041 206043 206044 206045 206048 206049 206050 206051 206054 206059 206062 206064 206071 206072 206074 206075 206077 206079 206080 206081 206085 206086 206087 206088 206089 206090 206091 206092 206096 206097 206099 206100 206101 206102 206103 206104 206107 206108 206109 206110 206111 206112 206113 206114 206118 206119 206121 206123 206127 206130 206133 206135 206143 206146 206147 206148 206150 206151 206155 206166 206192 206196 206196 206198 206203 206228 206247 206252 206269 206270 206271 206274 206276 206277 206278 206279 206292 206300 206301 206303 206315 206317 206324 206329 206339 206343 206359 206362 206363 206365 206373 206376 206392 206393 206397 206406 206412 206425 206426 206427 206451 206455 206468 206469 206471 206474 206482 206483 206491 206494 206499 206501 206502 206503 206508 206520 206526 206527 206528 206534 206535 206536 206537 206540 206544 206548 206549 206554 206555 206559 206561 206562 206567 206569 206572 206573 206576 206576 206587 206588 206591 206592 206593 206594 206595 206615 206616 206618 206623 206624 206625 206627 206643 206649 206650 206652 206657 206660 206663 206666 206669 206684 206687 206693 206701 206702 206704 206705 206706 206708 206709 206713 206725 206727 206731 206734 206739 206740 206744 206755 206760 206764 206765 206766 206767 206768 206770 206771 206778 206779 206779 206780 206781 206782 206785 206786 206789 206792 206793 206798 206803 206806 206808 206809 206813 206814 206816 206817 206818 206819 206820 206821 206825 206827 206828 206829 206830 206831 206832 206833 206840 206843 206844 206847 206852 206853 206854 206855 206858 206861 206872 206878 206884 206888 206892 206893 206901 206904 206905 206907 206910 206912 206916 206917 206929 206932 206938 206939 206942 206944 206949 206964 206968 206969 206973 206974 206975 206976 206977 206978 206981 206984 206986 206987 206988 206989 206994 206995 207010 207011 207012 207013 207014 207016 207017 207021 207028 207030 207032 207033 207035 207043 207050 207054 207063 207064 207065 207068 207083 207084 207085 207086 207090 207091 207092 207095 207104 207105 207106 207107 207117 207117 207118 207121 207122 207123 207124 207125 207128 207129 207130 207131 207134 207138 207139 207141 207156 207158 207159 207160 207163 207166 207167 207174 207178 207179 207180 207187 207188 207189 207192 207195 207203 207204 207206 207208 207209 207210 207211 207213 207216 207218 207220 207221 207222 207224 207225 207226 207231 207232 207233 207234 207235 207235 207236 207237 207244 207245 207251 207252 207253 207254 207255 207256 207259 207260 207261 207262 207263 207264 207266 207267 207270 207271 207272 207273 207274 207275 207276 207277 207280 207281 207282 207283 207284 207285 207286 207287 207290 207291 207292 207293 207294 207295 207297 207298 207301 207302 207303 207304 207305 207306 207307 207308 207311 207312 207313 207314 207317 207318 207319 207320 207323 207324 207325 207327 207329 207330 207332 207333 207336 207337 207338 207339 207340 207341 207342 207343 207346 207347 207348 207349 207350 207352 207353 207354 207357 207358 207359 207360 207361 207362 207363 207364 207368 207369 207371 207372 207379 207380 207381 207384 207390 207391 207392 207393 207394 207395 207396 207399 207402 207403 207404 207405 207406 207407 207408 207409 207412 207414 207415 207416 207417 207418 207419 207420 207423 207424 207425 207426 207429 207431 207432 207433 207436 207437 207438 207439 207440 207441 207442 207443 207446 207447 207448 207449 207450 207451 207453 207454 207457 207458 207459 207460 207463 207464 207465 207467 207472 207473 207474 207475 207476 207477 207478 207479 207482 207483 207484 207485 207486 207487 207488 207489 207492 207493 207493 207494 207495 207496 207497 207498 207505 207507 207509 207510 207511 207514 207517 207519 207522 207523 207524 207525 207526 207527 207528 207529 207533 207534 207535 207536 207537 207538 207539 207540 207543 207545 207546 207547 207548 207549 207550 207551 207554 207555 207557 207558 207559 207560 207561 207562 207573 207574 207577 207578 207579 207581 207582 207583 207586 207588 207589 207590 207591 207592 207593 207595 207602 207603 207604 207605 207606 207607 207607 207608 207609 207610 207611 207611 207612 207612 207613 207614 207617 207618 207619 207620 207627 207629 207630 207631 207633 207634 207637 207638 207640 207641 207642 207646 207650 207651 207652 207653 207654 207655 207656 207657 207660 207661 207662 207663 207664 207665 207666 207668 207676 207678 207679 207680 207681 207682 207683 207684 207688 207689 207690 207691 207692 207697 207698 207701 207704 207705 207706 207707 207708 207709 207719 207720 207723 207724 207725 207726 207727 207728 207729 207730 207733 207734 207735 207736 207737 207738 207739 207740 207757 207758 207759 207760 207761 207763 207764 207766 207769 207770 207771 207773 207774 207775 207776 207777 207782 207784 207784 207785 207785 207786 207786 207787 207790 207791 207792 207793 207794 207795 207796 207797 207808 207812 207813 207815 207816 207817 207818 207819 207822 207823 207824 207825 207826 207827 207828 207829 207832 207833 207834 207835 207836 207837 207838 207839 207842 207843 207844 207845 207846 207847 207848 207849 207852 207853 207854 207855 207856 207857 207858 207859 207862 207863 207864 207865 207866 207867 207868 207869 207874 207876 207877 207878 207879 207880 207881 207882 207885 207886 207887 207888 207889 207890 207891 207892 207895 207896 207897 207898 207899 207900 207901 207902 207905 207906 207915 207916 207917 207918 207919 207923 207926 207927 207928 207929 207930 207933 207934 207935 207938 207939 207940 207941 207942 207943 207944 207945 207948 207949 207950 207951 207952 207953 207954 207955 207958 207959 207960 207961 207962 207964 207965 207966 207969 207970 207971 207973 207974 207975 207976 207977 207980 207981 207982 207983 207984 207985 207986 207987 207991 207993 207994 207995 207999 208000 208001 208004 208008 208009 208010 208011 208012 208013 208014 208015 208018 208019 208020 208021 208022 208023 208024 208025 208028 208030 208031 208032 208033 208034 208035 208036 208040 208041 208042 208043 208044 208046 208047 208050 208056 208057 208058 208061 208062 208063 208064 208065 208068 208069 208070 208071 208072 208073 208074 208075 208078 208079 208080 208081 208082 208083 208084 208086 208091 208092 208093 208097 208098 208099 208100 208101 208104 208106 208107 208109 208110 208111 208112 208113 208116 208117 208118 208119 208120 208121 208123 208124 208127 208128 208129 208130 208131 208132 208134 208135 208140 208141 208142 208147 208151 208152 208153 208154 208157 208158 208159 208160 208161 208162 208163 208164 208167 208168 208169 208170 208171 208172 208173 208174 208177 208178 208179 208180 208181 208182 208184 208185 208188 208189 208190 208191 208192 208193 208194 208195 208204 208205 208206 208207 208208 208209 208210 208211 208214 208215 208218 208219 208220 208221 208222 208223 208227 208228 208229 208230 208231 208232 208233 208234 208242 208243 208246 208247 208248 208249 208250 208251 208254 208255 208256 208257 208258 208260 208261 208262 208265 208266 208267 208268 208269 208270 208271 208272 208283 208284 208285 208286 208287 208288 208289 208294 208297 208299 208300 208301 208302 208303 208304 208305 208308 208309 208310 208311 208312 208313 208314 208315 208318 208322 208323 208324 208325 208326 208327 208328 208331 208332 208333 208334 208335 208336 208337 208341 208348 208349 208350 208351 208352 208353 208354 208355 208358 208359 208360 208361 208362 208363 208364 208365 208368 208369 208370 208371 208372 208373 208374 208375 208378 208379 208380 208381 208382 208383 208384 208385 208388 208389 208390 208391 208392 208393 208394 208395 208398 208399 208401 208402 208403 208404 208405 208406 208409 208410 208411 208412 208413 208414 208415 208416 208419 208420 208421 208422 208423 208424 208425 208426 208429 208430 208431 208432 208433 208434 208435 208436 208439 208440 208441 208442 208443 208444 208445 208446 208449 208450 208451 208452 208453 208454 208455 208456 208459 208460 208461 208462 208463 208464 208465 208466 208469 208470 208471 208472 208473 208474 208475 208476 208479 208480 208481 208482 208483 208484 208486 208487 208490 208491 208492 208493 208494 208495 208496 208497 208500 208501 208502 208503 208504 208505 208506 208507 208510 208511 208512 208513 208514 208515 208516 208517 208520 208522 208523 208524 208525 208526 208527 208528 208531 208532 208533 208534 208535 208536 208537 208538 208541 208542 208543 208544 208545 208546 208547 208548 208551 208553 208554 208555 208556 208557 208559 208560 208563 208564 208565 208566 208567 208568 208569 208570 208573 208574 208575 208576 208577 208578 208579 208580 208583 208584 208585 208586 208587 208588 208591 208592 208595 208596 208597 208598 208599 208600 208601 208602 208605 208606 208607 208608 208609 208610 208611 208612 208615 208616 208617 208618 208619 208620 208621 208622 208625 208626 208627 208628 208629 208630 208631 208633 208636 208637 208638 208639 208640 208641 208642 208643 208646 208647 208648 208649 208650 208651 208652 208653 208656 208657 208658 208659 208660 208661 208663 208664 208668 208670 208671 208672 208673 208674 208675 208676 208679 208680 208681 208682 208683 208684 208685 208686 208689 208690 208691 208692 208693 208694 208695 208696 208699 208700 208702 208703 208704 208705 208706 208707 208710 208711 208712 208713 208714 208715 208716 208717 208721 208722 208723 208724 208725 208726 208727 208728 208731 208732 208733 208734 208735 208736 208737 208738 208741 208742 208743 208744 208745 208746 208747 208748 208751 208752 208753 208754 208755 208756 208757 208758 208761 208762 208763 208764 208765 208766 208767 208768 208771 208772 208773 208774 208775 208776 208777 208778 208781 208782 208783 208784 208785 208786 208787 208788 208791 208792 208793 208794 208795 208796 208797 208798 208801 208802 208803 208804 208805 208806 208807 208808 208812 208813 208814 208815 208816 208817 208818 208819 208822 208823 208824 208825 208826 208827 208828 208829 208838 208841 208842 208843 208844 208845 208846 208847 208850 208851 208852 208853 208854 208855 208856 208857 208860 208861 208862 208863 208864 208866 208867 208869 208872 208873 208874 208876 208877 208878 208879 208880 208883 208884 208886 208887 208888 208889 208890 208891 208894 208895 208896 208897 208898 208899 208901 208902 208905 208906 208907 208908 208909 208910 208911 208912 208915 208916 208917 208918 208934 208935 208936 208937 208940 208941 208974 208976 208978 208979 208983 208984 209000 209001 209002 209003 209004 209005 209006 209007 209017 209018 209019 209020 209021 209022 209023 209024 209028 209029 209030 209031 209032 209033 209034 209035 209038 209039 209040 209041 209042 209043 209044 209045 209049 209050 209051 209052 209053 209054 209055 209056 209059 209060 209061 209062 209063 209064 209065 209066 209070 209071 209072 209073 209074 209075 209076 209077 209080 209081 209083 209084 209085 209086 209087 209088 209099 209100 209101 209102 209103 209105 209106 209107 209111 209112 209115 209119 209121 209122 209123 209124 209127 209128 209129 209130 209131 209132 209133 209134 209137 209138 209139 209141 209142 209143 209144 209148 209151 209152 209153 209154 209155 209157 209158 209159 209162 209163 209164 209165 209166 209167 209168 209169 209172 209173 209174 209175 209176 209177 209178 209179 209182 209183 209184 209185 209186 209187 209188 209189 209192 209193 209194 209195 209196 209197 209198 209199 209202 209203 209204 209205 209206 209207 209208 209209 209212 209213 209214 209215 209216 209217 209218 209219 209222 209223 209224 209225 209226 209227 209228 209229 209232 209233 209234 209235 209236 209237 209238 209239 209242 209243 209244 209245 209246 209247 209248 209249 209253 209254 209255 209256 209257 209258 209259 209260 209264 209265 209266 209268 209270 209271 209272 209273 209276 209277 209283 209285 209287 209288 209289 209290 209296 209299 209301 209307 209308 209311 209312 209313 209317 209319 209320 209324 209325 209326 209328 209329 209332 209333 209335 209336 209337 209340 209341 209342 209345 209348 209349 209350 209351 209352 209352 209353 209356 209357 209358 209359 209360 209362 209366 209367 209370 209371 209372 209374 209375 209376 209377 209378 209382 209384 209385 209386 209387 209388 209389 209391 209395 209396 209397 209398 209400 209401 209402 209403 209407 209408 209409 209410 209411 209412 209413 209414 209418 209419 209420 209421 209424 209425 209426 209427 209433 209434 209437 209438 209439 209440 209441 209443 209451 209454 209455 209456 209457 209458 209459 209460 209464 209465 209466 209468 209469 209471 209472 209473 209478 209481 209485 209486 209487 209488 209490 209490 209493 209494 209495 209496 209497 209498 209499 209501 209505 209506 209507 209508 209522 209526 209532 209533 209537 209539 209540 209547 209548 209578 209579 209580 209583 209584 209585 209586 209587 209589 209590 209591 209611 209612 209616 209617 209624 209625 209626 209627 209653 209656 209658 209659 209662 209663 209664 209665 209677 209678 209679 209680 209684 209685 209686 209687 209690 209691 209694 209695 209696 209697 209698 209699 209703 209704 209706 209707 209710 209715 209716 209733 209736 209737 209738 209739 209740 209747 209749 209757 209777 209780 209781 209783 209786 209787 209789 209793 209798 209799 209800 209802 209803 209806 209807 209810 209819 209820 209821 209822 209823 209824 209825 209826 209832 209833 209834 209835 209836 209837 209838 209839 209843 209844 209845 209846 209847 209849 209850 209851 209856 209857 209858 209860 209861 209862 209863 209864 209867 209868 209869 209872 209874 209875 209876 209877 209880 209881 209894 209898 209899 209906 209911 209913 209922 209924 209925 209926 209927 209941 209980 210008 210013 210017 210019 210020 210021 210023 210024 210033 210036 210037 210039 210040 210041 210042 210043 210044 210048 210049 210050 210051 210052 210053 210054 210055 210056 210065 210066 210067 210068 210069 210070 210071 210079 210081 210082 210083 210084 210085 210086 210087 210090 210091 210092 210093 210094 210095 210112 210113 210116 210117 210118 210119 210120 210121 210122 210123 210126 210127 210128 210136 210139 210141 210149 210152 210168 210169 210170 210171 210172 210173 210174 210175 210178 210179 210180 210181 210183 210185 210186 210188 210191 210193 210194 210195 210196 210197 210198 210199 210202 210203 210206 210207 210208 210209 210210 210212 210215 210216 210217 210220 210221 210222 210223 210224 210227 210228 210229 210230 210231 210234 210235 210236 210239 210240 210241 210242 210243 210244 210245 210246 210249 210250 210251 210252 210253 210254 210255 210256 210260 210261 210264 210265 210266 210271 210272 210273 210276 210277 210278 210279 210280 210281 210282 210283 210286 210287 210288 210289 210290 210293 210294 210303 210306 210307 210308 210309 210310 210311 210312 210313 210316 210317 210319 210320 210321 210322 210323 210324 210327 210328 210329 210330 210331 210332 210333 210336 210340 210347 210352 210353 210354 210354 210359 210363 210366 210367 210368 210369 210370 210371 210372 210373 210376 210377 210378 210379 210380 210381 210384 210387 210391 210392 210393 210394 210395 210396 210397 210398 210401 210402 210403 210404 210405 210406 210408 210409 210413 210414 210415 210416 210417 210418 210419 210420 210423 210424 210425 210426 210427 210429 210430 210431 210434 210435 210436 210437 210438 210439 210440 210441 210449 210453 210454 210455 210459 210459 210460 210462 210465 210466 210467 210468 210469 210470 210471 210472 210476 210476 210478 210479 210480 210481 210482 210483 210487 210488 210489 210490 210491 210491 210491 210491 210491 210491 210491 210492 210493 210494 210495 210496 210505 210506 210507 210508 210510 210511 210513 210515 210518 210520 210521 210522 210523 210524 210525 210526 210529 210530 210532 210533 210534 210542 210543 210544 210552 210555 210556 210557 210558 210559 210560 210561 210564 210565 210566 210568 210569 210572 210575 210576 210579 210582 210583 210585 210586 210587 210589 210590 210593 210594 210595 210596 210597 210598 210599 210600 210603 210604 210605 210607 210608 210609 210610 210611 210614 210615 210616 210617 210618 210619 210620 210621 210624 210625 210628 210634 210635 210636 210637 210638 210641 210642 210643 210645 210647 210648 210649 210650 210655 210658 210659 210661 210662 210663 210664 210665 210670 210672 210673 210674 210675 210676 210677 210678 210683 210685 210687 210690 210691 210692 210693 210694 210697 210700 210701 210702 210703 210704 210705 210706 210709 210710 210711 210712 210713 210714 210715 210716 210719 210720 210721 210722 210723 210724 210725 210726 210729 210730 210732 210739 210740 210741 210742 210744 210747 210748 210749 210750 210751 210752 210753 210754 210757 210758 210759 210761 210762 210763 210765 210766 210769 210771 210772 210773 210774 210775 210777 210779 210783 210785 210786 210787 210789 210791 210792 210794 210797 210798 210799 210806 210807 210808 210810 210811 210814 210815 210816 210819 210820 210821 210822 210823 210827 210828 210829 210830 210831 210832 210833 210834 210837 210838 210839 210840 210853 210857 210858 210859 210862 210863 210864 210865 210866 210867 210868 210869 210872 210873 210874 210875 210876 210877 210878 210879 210882 210883 210884 210885 210886 210887 210888 210889 210892 210893 210894 210895 210897 210898 210899 210900 210903 210904 210905 210906 210907 210908 210909 210910 210913 210914 210915 210916 210917 210917 210918 210919 210922 210923 210926 210928 210929 210930 210931 210934 210937 210938 210939 210940 210941 210942 210944 210945 210951 210953 210954 210955 210956 210957 210958 210959 210963 210964 210965 210967 210969 210971 210980 210981 210986 210987 210988 210989 210990 210991 210992 210994 210998 210999 211001 211003 211004 211005 211006 211007 211010 211011 211012 211013 211014 211016 211017 211018 211036 211037 211038 211039 211040 211041 211042 211051 211069 211070 211071 211072 211077 211091 211100 211117 211131 211151 211153 211154 211156 211169 211170 211171 211174 211175 211181 211185 211191 211195 211196 211198 211208 211209 211211 211212 211213 211214 211215 211216 211220 211221 211222 211228 211229 211233 211234 211235 211238 211239 211243 211246 211251 211256 211257 211258 211261 211262 211263 211264 211265 211266 211277 211283 211287 211288 211289 211290 211291 211292 211299 211305 211310 211311 211312 211313 211315 211316 211317 211318 211321 211322 211325 211326 211327 211328 211329 211330 211350 211351 211352 211362 211363 211364 211365 211366 211369 211370 211371 211373 211374 211375 211376 211377 211380 211381 211382 211383 211384 211385 211386 211387 211390 211391 211392 211393 211395 211396 211397 211400 211408 211410 211411 211412 211413 211414 211415 211416 211419 211420 211421 211422 211423 211426 211427 211429 211432 211433 211436 211437 211438 211439 211440 211441 211444 211445 211446 211447 211448 211449 211450 211451 211456 211458 211459 211475 211476 211477 211478 211479 211482 211483 211488 211492 211493 211494 211496 211497 211501 211502 211503 211504 211505 211508 211509 211510 211513 211514 211515 211516 211517 211518 211519 211520 211523 211524 211525 211526 211527 211545 211557 211559 211609 211637 211639 211642 211651 211652 211653 211654 211657 211658 211667 211677 211679 211680 211682 211683 211686 211687 211688 211689 211690 211691 211692 211693 211696 211697 211698 211700 211701 211702 211703 211704 211707 211708 211709 211710 211711 211712 211713 211714 211717 211718 211719 211720 211721 211722 211723 211724 211728 211729 211730 211731 211732 211733 211734 211735 211738 211739 211740 211741 211742 211743 211744 211745 211748 211749 211750 211751 211752 211753 211754 211755 211758 211759 211760 211761 211762 211763 211764 211765 211768 211769 211770 211771 211772 211773 211774 211775 211778 211779 211780 211781 211782 211783 211784 211785 211789 211791 211792 211793 211794 211795 211796 211797 211809 211810 211811 211812 211813 211814 211816 211824 211827 211829 211831 211832 211833 211835 211836 211838 211841 211856 211874 211875 211876 211877 211878 211902 211907 211908 211909 211910 211912 211913 211924 211925 211931 211932 211941 211942 211945 211947 211948 211949 211964 211968 211969 211971 211973 211974 211975 211976 211986 211988 212002 212003 212021 212023 212027 212029 212037 212040 212041 212042 212043 212053 212065 212067 212070 212071 212072 212073 212091 212092 212093 212113 212126 212158 212159 212160 212161 212162 212195 212196 212199 212201 212202 212206 212207 212261 212331 212361 212402 212404 212408 212409 212410 212411 212412 212413 212442 212444 212449 212460 212461 212462 212463 212464 212467 212469 212470 212471 212472 212473 212475 212476 212482 212483 212484 212485 212486 212487 212488 212489 212500 212503 212504 212519 212523 212524 212529 212532 212551 212552 212553 212559 212561 212566 212567 212569 212575 212577 212579 212580 212589 212590 212591 212592 212595 212596 212597 212598 212599 212600 212601 212602 212605 212606 212607 212608 212609 212610 212611 212612 212615 212616 212617 212618 212619 212620 212621 212622 212625 212626 212635 212638 212639 212640 212641 212642 212646 212648 212649 212650 212653 212657 212664 212665 212670 212672 212673 212674 212676 212678 212679 212693 212699 212702 212707 212710 212712 212714 212717 212722 212738 212750 212751 212753 212754 212757 212759 212761 212769 212770 212772 212778 212781 212782 212797 212798 212804 212806 212810 212811 212813 212814 212815 212816 212819 212820 212821 212822 212824 212829 212831 212832 212835 212836 212837 212838 212839 212840 212841 212843 212848 212849 212850 212851 212852 212853 212854 212855 212860 212861 212862 212863 212864 212865 212866 212867 212870 212871 212872 212873 212874 212875 212878 212882 212886 212891 212892 212893 212894 212895 212896 212897 212900 212901 212904 212906 212907 212908 212909 212911 212914 212915 212916 212917 212918 212919 212920 212921 212924 212925 212926 212927 212928 212929 212930 212931 212934 212935 212936 212937 212938 212939 212940 212941 212944 212945 212946 212947 212948 212949 212950 212951 212958 212959 212961 212964 212965 212971 212978 212979 212983 212984 212985 212986 212987 212988 212989 212990 212993 212994 212997 212998 212999 213000 213001 213003 213006 213008 213015 213020 213021 213022 213024 213025 213028 213029 213030 213031 213032 213033 213034 213035 213046 213047 213048 213049 213050 213051 213053 213054 213060 213061 213062 213063 213064 213065 213066 213067 213073 213074 213076 213077 213080 213082 213087 213089 213094 213095 213096 213097 213098 213099 213100 213101 213104 213105 213110 213111 213112 213113 213114 213115 213118 213119 213120 213127 213129 213130 213131 213134 213139 213142 213144 213145 213146 213149 213152 213153 213156 213157 213158 213159 213168 213169 213170 213171 213174 213175 213177 213178 213179 213180 213181 213182 213185 213187 213189 213190 213191 213200 213201 213202 213207 213208 213209 213210 213211 213212 213213 213214 213217 213218 213219 213220 213221 213222 213223 213224 213227 213228 213229 213230 213231 213232 213233 213234 213237 213238 213239 213240 213241 213242 213243 213247 213250 213251 213252 213253 213254 213255 213256 213257 213260 213261 213262 213263 213264 213265 213266 213267 213270 213271 213272 213273 213274 213275 213276 213277 213280 213281 213282 213284 213285 213286 213287 213288 213291 213292 213293 213294 213295 213296 213297 213299 213303 213304 213306 213308 213310 213311 213312 213313 213316 213317 213318 213319 213320 213321 213322 213323 213326 213327 213328 213339 213340 213342 213343 213344 213347 213348 213349 213350 213351 213352 213353 213354 213359 213360 213361 213362 213363 213364 213365 213366 213369 213370 213374 213375 213386 213387 213388 213389 213392 213393 213394 213395 213396 213397 213398 213400 213414 213415 213417 213418 213420 213429 213430 213431 213434 213435 213436 213439 213440 213441 213446 213450 213461 213462 213463 213464 213465 213466 213467 213468 213479 213484 213486 213487 213488 213489 213493 213494 213498 213499 213500 213501 213504 213505 213508 213509 213512 213513 213514 213515 213516 213521 213531 213538 213566 213567 213568 213570 213572 213573 213578 213580 213594 213595 213597 213598 213601 213602 213604 213605 213608 213609 213610 213611 213612 213613 213618 213631 213638 213639 213640 213641 213642 213643 213644 213645 213648 213649 213650 213651 213652 213653 213654 213655 213658 213659 213660 213662 213663 213668 213669 213671 213674 213675 213676 213677 213678 213679 213680 213681 213684 213685 213686 213687 213696 213697 213699 213700 213703 213704 213705 213706 213707 213708 213710 213711 213715 213717 213722 213723 213724 213725 213726 213728 213731 213732 213733 213734 213735 213736 213737 213738 213741 213742 213743 213744 213746 213748 213750 213751 213755 213756 213758 213762 213764 213766 213768 213769 213772 213773 213774 213775 213776 213777 213778 213779 213787 213796 213797 213798 213799 213800 213801 213802 213805 213806 213807 213808 213809 213810 213811 213812 213816 213817 213819 213820 213821 213822 213823 213824 213827 213828 213829 213830 213831 213832 213833 213834 213837 213838 213839 213840 213842 213845 213846 213853 213866 213868 213870 213872 213874 213880 213881 213882 213887 213888 213890 213891 213892 213893 213894 213895 213898 213902 213905 213906 213907 213908 213909 213910 213913 213914 213915 213917 213918 213919 213920 213922 213925 213926 213927 213928 213929 213930 213937 213938 213941 213945 213946 213953 213955 213962 213964 213965 213968 213970 213972 213973 213974 213976 213977 213978 213981 213982 213983 213984 213985 213986 213987 213988 213991 213992 213993 213994 213997 214001 214003 214004 214007 214008 214009 214010 214011 214012 214013 214014 214017 214018 214019 214020 214021 214022 214023 214024 214027 214028 214029 214030 214031 214032 214033 214034 214037 214038 214039 214040 214041 214042 214043 214044 214047 214048 214049 214050 214051 214052 214053 214054 214057 214058 214059 214060 214061 214062 214063 214064 214067 214068 214069 214070 214071 214072 214073 214074 214077 214078 214079 214080 214081 214082 214083 214084 214088 214089 214090 214091 214092 214093 214094 214095 214098 214099 214100 214101 214102 214103 214105 214106 214109 214110 214111 214112 214113 214114 214115 214116 214119 214120 214121 214122 214123 214124 214125 214126 214129 214130 214131 214132 214133 214134 214135 214137 214140 214141 214142 214143 214144 214145 214148 214149 214152 214154 214156 214157 214158 214159 214160 214161 214164 214165 214166 214167 214169 214170 214171 214174 214177 214178 214179 214180 214181 214182 214183 214184 214188 214189 214190 214191 214192 214193 214194 214195 214199 214200 214201 214202 214203 214204 214205 214206 214210 214211 214219 214220 214224 214227 214228 214229 214232 214233 214238 214242 214244 214245 214246 214261 214268 214280 214297 214314 214315 214318 214319 214321 214326 214341 214367 214368 214369 214370 214371 214372 214376 214377 214378 214379 214381 214383 214385 214386 214389 214390 214392 214393 214394 214395 214396 214397 214400 214401 214402 214403 214404 214405 214406 214408 214411 214412 214413 214414 214415 214416 214417 214418 214422 214423 214424 214425 214426 214427 214436 214437 214457 214458 214458 214458 214458 214458 214459 214460 214463 214464 214465 214466 214467 214468 214469 214470 214473 214474 214475 214476 214477 214478 214479 214480 214484 214485 214487 214488 214489 214490 214491 214492 214495 214496 214497 214498 214499 214500 214501 214502 214505 214506 214507 214508 214509 214510 214511 214512 214515 214516 214517 214518 214519 214520 214521 214522 214525 214526 214528 214529 214530 214532 214533 214535 214539 214541 214542 214543 214546 214547 214550 214551 214554 214555 214556 214557 214558 214559 214560 214561 214566 214580 214582 214583 214584 214592 214593 214594 214597 214598 214599 214600 214601 214602 214606 214607 214610 214611 214612 214613 214614 214615 214616 214618 214622 214623 214624 214625 214626 214627 214628 214629 214632 214633 214635 214636 214637 214638 214639 214640 214643 214644 214645 214646 214647 214648 214649 214650 214653 214654 214655 214656 214657 214658 214659 214660 214663 214664 214665 214666 214667 214668 214669 214670 214673 214674 214675 214676 214677 214678 214679 214680 214683 214684 214685 214686 214687 214688 214689 214690 214693 214694 214695 214696 214697 214698 214702 214706 214709 214710 214718 214721 214724 214728 214740 214741 214762 214763 214764 214765 214767 214768 214769 214770 214773 214774 214775 214776 214777 214778 214785 214786 214789 214790 214791 214792 214793 214794 214795 214796 214800 214801 214802 214803 214804 214805 214811 214816 214820 214821 214823 214824 214825 214826 214827 214828 214831 214832 214833 214834 214836 214837 214838 214839 214842 214843 214844 214845 214846 214848 214849 214850 214853 214854 214855 214856 214857 214858 214859 214860 214863 214864 214865 214866 214867 214868 214869 214870 214874 214875 214876 214877 214878 214879 214880 214881 214884 214885 214886 214887 214888 214889 214890 214891 214894 214895 214896 214897 214898 214899 214900 214901 214904 214905 214906 214907 214908 214909 214912 214913 214916 214917 214918 214919 214920 214921 214922 214923 214928 214929 214930 214931 214932 214933 214935 214936 214939 214940 214941 214942 214943 214944 214945 214946 214949 214950 214951 214952 214953 214954 214955 214956 214959 214960 214961 214962 214963 214964 214965 214966 214969 214970 214971 214972 214973 214974 214975 214976 214979 214980 214981 214982 214983 214984 214985 214986 214989 214990 214992 214993 214994 214995 214996 214997 215000 215001 215002 215003 215004 215005 215006 215026 215115 215117 215123 215124 215125 215126 215127 215134 215137 215138 215139 215140 215141 215142 215143 215144 215147 215148 215149 215150 215151 215152 215178 215179 215183 215184 215190 215191 215192 215193 215194 215195 215198 215199 215200 215201 215202 215203 215204 215205 215208 215209 215210 215211 215212 215213 215214 215215 215218 215219 215220 215221 215222 215223 215224 215225 215228 215229 215230 215231 215232 215233 215234 215235 215238 215239 215240 215241 215242 215243 215244 215245 215248 215249 215250 215251 215252 215253 215254 215255 215258 215259 215260 215261 215262 215263 215264 215265 215268 215269 215270 215271 215272 215273 215274 215275 215279 215280 215281 215282 215283 215284 215285 215286 215289 215290 215291 215292 215293 215294 215295 215296 215299 215300 215301 215302 215303 215304 215305 215306 215309 215310 215311 215312 215313 215314 215315 215316 215319 215320 215321 215322 215323 215324 215325 215326 215329 215330 215331 215332 215333 215334 215335 215336 215339 215340 215341 215342 215343 215344 215345 215346 215349 215350 215351 215352 215353 215354 215356 215357 215361 215362 215363 215364 215365 215366 215367 215391 215406 215407 215408 215409 215410 215411 215412 215413 215421 215422 215424 215425 215426 215427 215428 215429 215432 215433 215434 215435 215436 215437 215438 215439 215442 215443 215444 215446 215448 215449 215450 215451 215454 215455 215456 215457 215458 215459 215460 215461 215464 215465 215466 215467 215468 215469 215470 215471 215475 215476 215477 215478 215481 215482 215483 215484 215488 215489 215490 215491 215492 215493 215494 215495 215498 215499 215500 215501 215503 215504 215505 215506 215509 215510 215512 215513 215514 215515 215516 215517 215520 215521 215522 215524 215525 215526 215527 215528 215531 215532 215534 215535 215536 215537 215538 215539 215543 215544 215547 215550 215551 215552 215553 215554 215557 215569 215570 215579 215580 215581 215582 215583 215587 215589 215592 215594 215595 215596 215597 215598 215601 215602 215603 215604 215605 215606 215607 215608 215611 215612 215613 215614 215615 215616 215617 215618 215622 215623 215624 215625 215626 215627 215628 215629 215632 215633 215634 215635 215636 215663 215666 215670 215714 215717 215719 215742 215754 215770 215773 215792 215842 215848 215852 215855 215860 215861 215862 215863 215901 215922 215931 215980 216008 216009 216024 216026 216081 216090 216091 216092 216117 216118 216126 216137 216142 216143 216146 216167 216217 216226 216231 216244 216251 216267 216272 216283 216284 216285 216286 216287 216290 216291 216292 216293 216294 216295 216296 216298 216301 216302 216303 216305 216306 216307 216308 216309 216312 216314 216315 216316 216317 216318 216319 216320 216323 216324 216326 216327 216328 216354 216365 216367 216407 216436 216438 216483 216484 216485 216486 216524 216560 216561 216562 216565 216566 216624 216635 216636 216640 216641 216642 216643 216644 216645 216646 216647 216650 216651 216652 216653 216654 216655 216656 216657 216660 216661 216662 216663 216664 216665 216666 216667 216674 216675 216676 216677 216678 216679 216680 216681 216685 216686 216687 216688 216689 216690 216691 216692 216696 216697 216698 216699 216700 216701 216702 216703 216707 216708 216709 216710 216711 216712 216713 216714 216717 216718 216719 216721 216722 216723 216724 216725 216728 216729 216730 216731 216732 216733 216735 216736 216740 216741 216742 216744 216746 216747 216748 216750 216753 216754 216755 216756 216757 216758 216759 216760 216764 216765 216766 216767 216768 216769 216770 216771 216775 216776 216777 216778 216779 216780 216781 216782 216785 216786 216787 216788 216790 216791 216792 216793 216863 216864 216865 216866 216867 216868 216873 216874 216939 217013 217014 217015 217016 217017 217018 217019 217025 217026 217027 217028 217046 217047 217067 217075 217172 217184 217189 217203 217225 217227 217231 217234 217247 217248 217249 217251 217252 217253 217254 217255 217258 217259 217260 217262 217263 217264 217265 217266 217269 217270 217271 217273 217274 217275 217276 217278 217283 217284 217286 217288 217289 217292 217293 217295 217300 217301 217302 217303 217305 217306 217307 217308 217314 217315 217316 217317 217319 217320 217321 217322 217328 217329 217330 217331 217332 217334 217335 217337 217343 217344 217345 217346 217347 217348 217349 217350 217353 217354 217355 217356 217357 217358 217359 217360 217363 217364 217365 217366 217367 217368 217369 217370 217373 217374 217375 217376 217377 217378 217379 217380 217383 217384 217385 217386 217387 217388 217389 217390 217393 217395 217396 217397 217398 217399 217400 217401 217404 217405 217406 217407 217408 217409 217410 217411 217414 217415 217416 217417 217443 217444 217445 217446 217449 217450 217451 217452 217453 217454 217455 217456 217459 217460 217461 217462 217463 217464 217465 217466 217469 217470 217471 217472 217473 217474 217475 217476 217479 217480 217481 217483 217484 217485 217486 217487 217490 217491 217492 217493 217494 217495 217496 217497 217500 217501 217502 217503 217504 217505 217506 217507 217510 217511 217512 217513 217514 217515 217516 217517 217520 217521 217522 217523 217524 217525 217526 217527 217530 217531 217532 217533 217534 217535 217537 217538 217541 217542 217543 217544 217545 217546 217547 217548 217551 217552 217553 217554 217555 217556 217557 217558 217561 217562 217563 217564 217566 217567 217568 217569 217572 217573 217574 217575 217576 217577 217578 217579 217582 217583 217584 217585 217586 217587 217588 217590 217594 217595 217596 217597 217598 217599 217600 217601 217605 217608 217609 217610 217611 217612 217613 217614 217618 217619 217620 217621 217622 217623 217626 217627 217633 217634 217636 217637 217638 217642 217644 217645 217648 217649 217651 217652 217653 217654 217655 217656 217659 217660 217661 217662 217663 217664 217665 217666 217670 217671 217672 217673 217674 217675 217676 217677 217680 217681 217682 217683 217684 217685 217686 217688 217691 217692 217693 217694 217695 217696 217697 217698 217701 217702 217703 217704 217705 217706 217707 217708 217711 217712 217713 217714 217715 217716 217717 217718 217721 217722 217723 217724 217725 217726 217727 217728 217731 217732 217733 217734 217735 217736 217737 217738 217741 217742 217743 217744 217745 217746 217747 217748 217751 217752 217753 217754 217755 217756 217758 217759 217762 217763 217764 217765 217766 217767 217768 217769 217772 217773 217774 217775 217776 217777 217778 217779 217782 217783 217784 217785 217786 217787 217788 217789 217792 217793 217794 217795 217796 217798 217799 217800 217803 217804 217805 217806 217807 217808 217809 217810 217813 217814 217815 217816 217817 217818 217819 217820 217824 217825 217826 217827 217828 217829 217831 217832 217835 217837 217838 217839 217841 217842 217843 217844 217847 217848 217849 217850 217851 217852 217853 217854 217857 217859 217860 217861 217862 217863 217864 217865 217868 217869 217870 217871 217872 217873 217874 217875 217878 217879 217880 217881 217882 217883 217884 217885 217888 217889 217890 217891 217892 217893 217894 217895 217898 217899 217900 217901 217902 217903 217904 217905 217908 217909 217910 217911 217912 217913 217914 217915 217918 217919 217920 217921 217922 217923 217924 217925 217928 217929 217930 217931 217932 217933 217934 217935 217938 217939 217940 217941 217942 217943 217944 217945 217948 217949 217950 217951 217952 217953 217954 217955 217958 217959 217960 217961 217962 217963 217964 217965 217969 217970 217971 217972 217973 217974 217975 217976 217980 217981 217982 217983 217984 217985 217986 217987 217990 217991 217992 217993 217994 217995 217996 217997 218000 218001 218002 218003 218004 218005 218006 218007 218010 218011 218012 218013 218014 218015 218016 218017 218020 218021 218022 218023 218024 218025 218026 218027 218030 218031 218032 218033 218034 218035 218036 218037 218040 218041 218042 218043 218044 218045 218046 218047 218050 218051 218052 218053 218054 218055 218056 218057 218060 218061 218062 218063 218064 218066 218067 218068 218071 218072 218073 218074 218075 218076 218077 218078 218081 218082 218083 218084 218085 218086 218087 218088 218091 218092 218093 218094 218095 218096 218097 218098 218101 218102 218103 218104 218105 218106 218107 218108 218111 218112 218113 218114 218115 218116 218117 218118 218121 218122 218123 218124 218125 218126 218127 218129 218132 218133 218134 218135 218136 218137 218138 218139 218142 218143 218144 218145 218146 218147 218148 218150 218153 218154 218155 218156 218157 218158 218159 218160 218163 218164 218165 218166 218167 218168 218169 218170 218173 218174 218175 218176 218177 218178 218179 218180 218183 218184 218185 218186 218187 218188 218189 218190 218193 218194 218195 218196 218197 218198 218199 218200 218203 218204 218205 218206 218207 218208 218209 218210 218213 218214 218215 218216 218217 218218 218219 218220 218223 218224 218225 218226 218227 218228 218229 218230 218233 218234 218235 218236 218237 218238 218239 218240 218243 218244 218245 218246 218247 218248 218249 218250 218253 218254 218255 218256 218257 218258 218259 218260 218264 218265 218266 218267 218268 218269 218270 218271 218274 218275 218276 218277 218278 218279 218280 218281 218284 218285 218286 218287 218288 218289 218290 218291 218294 218295 218296 218297 218298 218299 218300 218301 218304 218305 218306 218307 218308 218310 218311 218312 218318 218319 218320 218321 218322 218323 218324 218325 218328 218329 218330 218331 218332 218333 218334 218335 218338 218339 218340 218341 218342 218343 218344 218345 218348 218349 218350 218351 218352 218353 218354 218355 218358 218359 218360 218361 218362 218363 218365 218366 218369 218370 218371 218372 218373 218374 218375 218377 218380 218381 218382 218383 218384 218385 218386 218387 218391 218392 218393 218394 218395 218396 218397 218398 218401 218402 218403 218404 218405 218406 218407 218408 218411 218412 218414 218415 218416 218417 218418 218419 218422 218423 218424 218425 218426 218427 218428 218429 218457 218458 218459 218460 218461 218467 218468 218469 218472 218477 218478 218479 218480 218481 218482 218483 218486 218487 218488 218489 218490 218491 218492 218493 218496 218497 218498 218499 218502 218503 218504 218505 218508 218509 218510 218511 218532 218533 218534 218535 218538 218539 218540 218541 218542 218543 218544 218545 218548 218549 218550 218551 218552 218553 218554 218555 218558 218559 218560 218561 218562 218563 218564 218565 218568 218569 218570 218571 218572 218573 218574 218575 218578 218579 218580 218581 218582 218583 218584 218585 218589 218590 218591 218592 218593 218594 218608 218633 218673 218675 218711 218714 218718 218719 218720 218721 218850 218855 218872 218877 218878 218880 218881 218882 218885 218887 218888 218889 218890 218891 218892 218893 218896 218897 218898 218899 218900 218901 218902 218903 218906 218907 218908 218909 218910 218911 218912 218913 218916 218917 218918 218919 218920 218921 218922 218923 218926 218927 218928 218929 218930 218931 218932 218933 218936 218937 218938 218939 218940 218941 218942 218943 218946 218947 218948 218949 218950 218951 218952 218953 218956 218957 218958 218959 218960 218961 218962 218963 218966 218967 218968 218969 218970 218971 218972 218973 218976 218977 218978 218979 218980 218981 218982 218983 218986 218987 218988 218989 218990 218991 218992 218993 218996 218997 218998 218999 219000 219001 219003 219004 219007 219008 219009 219010 219011 219012 219013 219014 219017 219018 219019 219020 219021 219022 219023 219024 219027 219028 219029 219030 219031 219032 219033 219034 219037 219038 219039 219040 219041 219042 219043 219044 219047 219048 219049 219050 219051 219052 219053 219054 219057 219058 219059 219060 219061 219062 219063 219064 219068 219069 219070 219071 219072 219073 219074 219075 219078 219079 219080 219081 219082 219083 219084 219085 219088 219089 219090 219091 219092 219093 219094 219095 219098 219099 219100 219101 219102 219103 219104 219105 219108 219109 219110 219111 219112 219113 219114 219115 219118 219119 219120 219121 219122 219123 219124 219125 219128 219129 219130 219131 219132 219133 219134 219135 219138 219139 219140 219141 219142 219143 219144 219145 219148 219149 219150 219151 219152 219153 219154 219155 219158 219159 219160 219161 219162 219163 219164 219165 219168 219169 219170 219171 219172 219173 219174 219175 219198 219202 219236 219263 219296 219312 219331 219351 219365 219366 219367 219369 219371 219372 219373 219374 219380 219381 219383 219384 219385 219386 219387 219388 219391 219392 219394 219395 219396 219397 219398 219399 219402 219403 219404 219405 219406 219407 219408 219409 219412 219413 219414 219415 219417 219418 219419 219420 219423 219424 219425 219426 219427 219428 219429 219430 219433 219486 219487 219488 219489 219490 219491 219492 219495 219496 219497 219498 219501 219502 219503 219504 219507 219508 219509 219510 219511 219512 219514 219515 219518 219519 219520 219521 219522 219523 219524 219526 219529 219530 219534 219535 219536 219538 219540 219541 219544 219545 219546 219547 219548 219549 219550 219551 219555 219556 219557 219558 219559 219560 219561 219562 219565 219566 219567 219569 219570 219571 219572 219573 219578 219579 219580 219582 219583 219584 219585 219586 219589 219590 219591 219592 219593 219594 219595 219596 219599 219600 219601 219602 219603 219604 219605 219606 219609 219610 219611 219612 219613 219614 219615 219616 219619 219620 219621 219622 219623 219624 219625 219626 219629 219630 219631 219632 219633 219634 219635 219636 219640 219641 219642 219643 219644 219645 219646 219647 219650 219651 219652 219653 219654 219655 219656 219657 219660 219661 219662 219712 219713 219715 219717 219717 219720 219721 219722 219723 219724 219725 219725 219726 219730 219731 219732 219733 219735 219736 219737 219738 219741 219742 219743 219744 219745 219746 219748 219749 219752 219754 219755 219757 219758 219759 219760 219761 219765 219766 219767 219768 219769 219770 219771 219772 219775 219777 219778 219779 219780 219781 219794 219795 219798 219799 219800 219801 219803 219804 219805 219806 219810 219811 219812 219813 219814 219815 219818 219820 219823 219830 219832 219833 219845 219846 219847 219848 219853 219905 219906 219908 219909 219910 219911 219912 219915 219916 219917 219918 219919 219953 219955 219962 219986 219987 219988 220001 220002 220003 220005 220006 220009 220010 220011 220012 220013 220014 220015 220016 220020 220021 220022 220023 220024 220025 220026 220027 220030 220031 220032 220034 220038 220039 220040 220042 220046 220047 220048 220049 220050 220051 220052 220053 220057 220058 220059 220060 220061 220063 220064 220066 220070 220071 220074 220075 220076 220077 220081 220082 220089 220090 220092 220093 220094 220099 220100 220101 220107 220108 220109 220110 220128 220129 220174 220175 220178 220179 220180 220181 220182 220183 220184 220185 220188 220189 220190 220191 220192 220193 220194 220195 220198 220199 220200 220201 220202 220203 220204 220205 220232 220255 220256 220257 220258 220260 220261 220262 220265 220266 220267 220268 220269 220271 220272 220273 220276 220277 220279 220280 220281 220282 220283 220284 220287 220288 220289 220291 220292 220293 220294 220295 220298 220299 220300 220301 220302 220303 220305 220306 220309 220310 220311 220355 220356 220358 220359 220360 220385 220395 220396 220397 220398 220399 220401 220402 220432 220433 220434 220435 220436 220437 220451 220452 220455 220456 220457 220458 220459 220460 220461 220462 220465 220466 220467 220468 220469 220470 220471 220472 220475 220476 220516 220517 220518 220519 220520 220521 220525 220526 220527 220528 220529 220530 220532 220533 220536 220537 220538 220539 220540 220541 220542 220594 220797 220852 220853 220854 220855 220856 220857 220858 220864 220865 220866 220867 220937 220939 220940 220941 220944 220945 220946 220947 220948 220949 220950 220951 220954 220955 220956 220957 220958 220959 220960 220961 220964 220966 221012 221013 221014 221015 221016 221017 221020 221021 221022 221023 221024 221025 221026 221027 221030 221031 221032 221033 221035 221036 221037 221038 221041 221042 221043 221044 221045 221046 221054 221055 221059 221060 221061 221062 221063 221064 221065 221066 221069 221070 221131 221132 221133 221135 221136 221137 221142 221154 221155 221156 221157 221158 221370 221371 221374 221376 221377 221378 221379 221390 221391 221392 221395 221396 221397 221399 221400 221401 221402 221403 221406 221407 221408 221409 221410 221411 221413 221414 221419 221420 221421 221422 221423 221424 221426 221427 221430 221431 221432 221496 221497 221498 221687 221747 221750 221751 221752 221753 221754 221755 221756 221758 221761 221762 221763 221765 221766 221767 221768 221769 221778 221779 221780 221782 221783 221784 221785 221787 221791 221792 221793 221794 221795 221796 221797 221798 221802 221803 221804 221805 221806 221807 221809 221810 221851 221852 221853 221854 221855 221856 221929 221929 221931 221931 221932 221932 221933 221933 221934 221935 221946 221946 221948 221949 221950 221951 221952 221953 221979 221980 221981 221983 221984 221985 221986 221987 221990 221991 221992 221993 221994 221995 221996 222001 222065 222066 222067 222068 222069 222070 222071 222072 222075 222076 222077 222078 222080 222084 222085 222086 222089 222090 222091 222092 222093 222094 222095 222096 222099 222100 222101 222102 222103 222104 222105 222106 222109 222111 222112 222113 222115 222116 222117 222118 222126 222127 222171 222385 222386 222387 222388 222389 222393 222394 222395 222396 222397 222398 222399 222400 222467 222468 222469 222470 222471 222472 222646 222647 222693 222703 222710 222712 222713 222715 222716 222717 222721 222722 222723 222724 222725 222726 222730 222735 222738 222739 222740 222741 222742 222743 222744 222746 222749 222750 222751 222752 222753 222754 222755 222756 222759 222766 222767 |
Recalling Firm/ Manufacturer |
Zevex International, Inc. 4314 Zevex Park Ln Salt Lake City UT 84123
|
For Additional Information Contact |
801-264-1001 Ext. 144
|
Manufacturer Reason for Recall |
Infusion pumps may exhibit false error codes which render the device inoperable until a recovery process is performed.
|
FDA Determined Cause 2 |
Software design |
Action |
The firm, MOOG, sent an "URGENT DEVICE RECALL NOTIFICATION" letter dated February 28, 2011 to its customers. The letter described the product, problem and actions to be taken. The Customers were instructed to do the following:
1) Identify the affected Pumps:
a) If they have an affected pump, fill out and return the enclosed Business Reply Card within five (5) working days.
b) If their pump is not affected by this recall, the pump can be used for its intended purposes.
2) Removal of affected Product:
a) remove affected products from service and use non-affected pump
b) If you do not have an alternate pump available for use, MOOG recommended that they weigh the risks and benefits for delivering medications using the affected pump versus the risk of not using any pump for each patient
WARNING: Affected pumps should not be used to deliver critical medications where a delay in therapy could result in serious adverse health effects, injury or death; and affected pumps should not be used with the neonatal population.
3)Should they Have an Affected pump which exhibits Error Code 45 or is used in therapeutic situations specified in the warning, please contact MOOG Customer Service at (800) 970-2337, Monday thru Friday, 7:00am to 5:00pm MST to arrange for a loaner pump and the return of your affected pump.
If you have any additional questions regarding this recall, please contact MOOG's Customers Service Department at (800) 970-2337. |
Quantity in Commerce |
11,902 units |
Distribution |
Worldwide distribution: USA and countries including: Australia, Canada, Costa Rica, Croatia, Dubai, France, Germany, Ireland, Italy, Japan, Jordan, Kuwait, Malaysia, New Zealand, Qatar, Saudi Arabia, Singapore, South Africa, Taiwan, Turkey, UAE, United Kingdom. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = FRN and Original Applicant = BECTON DICKINSON CURLIN, LLC.
|
|
|
|