Scope History (select icon to view/hide Scope History)
|
Action Type |
ASCA Status |
Effective Start Date
|
Biocompatibility Test Method
|
Recognition Number
|
Designation Number |
Exclusions |
| Accredit |
Accreditation Granted |
02/10/2022 |
|
19-36
|
60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION
|
Excluding Figure 3, "nursing homes" as an example of the HOME HEALTHCARE ENVIRONMENT, Subclause 8.9, Table 8 on Page 39: The citation of Note k) under "Conducted disturbances induced by RF fields" (4th Row). Accreditation also excludes clause 8.9, table 6 Electrical transient conduction along supply lines/ISO7637-2, and clause 7.2, table 2 Note d
|
| Accredit |
Accreditation Granted |
02/10/2022 |
|
19-8
|
60601-1-2 Edition 4.0 2014-02
|
Excluding Figure 3, "nursing homes" as an example of the HOME HEALTHCARE ENVIRONMENT, Subclause 8.9, Table 8 on Page 39: The citation of Note b) under "Conducted disturbances induced by RF fields" (4th Row). Accreditation also excludes clause 8.9, table 6 Electrical transient conduction along supply lines/ISO7637-2, and clause 7.2, table 2 Note d.
|
|