• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Recognized Consensus Standards: Medical Devices

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

New Search Back To Search Results
Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 050 Date of Entry 09/17/2018 
FR Recognition Number 14-517
Standard
ASTM  F3293-18
Standard Guide for Application of Test Soils for the Validation of Cleaning Methods for Reusable Medical Devices
Scope/Abstract
1.1 This guide provides methods and considerations for simulated soiling of reusable medical devices for the purpose of validating cleaning instructions. Techniques for application of soil, as well as incorporation of soil by various means (e.g., actuation of devices) will be described in order to assure worst-case contamination of the surface geometry of medical devices.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Relevant FDA Guidance and/or Supportive Publications*
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contacts
 Steven Turtil
  FDA/OC/CDRH/OPEQ/OHTVI/DHTVIB/
  301-796-6305
  steven.turtil@fda.hhs.gov
 Shanil Haugen
  FDA/OC/CDRH/OPEQ/OHTIII/DHTIIIA/
  301-796-0301
  shanil.haugen@fda.hhs.gov
Standards Development Organization
ASTM ASTM International http://www.astm.org/
FDA Specialty Task Group (STG)
Sterility
*These are provided as examples and others may be applicable.
-
-