• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Recognized Consensus Standards: Medical Devices

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

New Search Back To Search Results
Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 056 Date of Entry 06/07/2021 
FR Recognition Number 12-338
Standard(Included in ASCA)
IEC  60601-2-1 Edition 4.0 2020-10
Medical electrical equipment - Part 2-1: Particular requirements for the basic safety and essential performance of electron accelerators in the range 1 MeV to 50 MeV
Scope/Abstract
This part of IEC 60601 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of ELECTRON ACCELERATORS, hereafter referred to as ME EQUIPMENT, in the range 1 MeV to 50 MeV, used for TREATMENT of PATIENTS.

This particular standard, with the inclusion of TYPE TESTS and SITE TESTS, applies to the manufacture and some installation aspects of ELECTRON ACCELERATORS and their included equipment used to increase the precision, accuracy and volumetric targeting of the TREATMENT delivery
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number Device Name Device Class Product Code
§892.5050 Cyclotron, Medical Class 2 IWK
§892.5050 Synchrotron, Medical Class 2 IWM
§892.5050 Accelerator, Linear, Medical Class 2 IYE
§892.5050 Betatron, Medical Class 2 IYG
§892.5050 Microtron, Medical Class 2 JAE
§892.5050 System, Radiation Therapy, Charged-Particle, Medical Class 2 LHN
§892.5050 System, Planning, Radiation Therapy Treatment Class 2 MUJ
§892.5050 Accessory - Film Dosimetry System Class 2 MWW
§892.5050 Dosimeter, Ionizing Radiation, Implanted Class 2 NZT
Relevant FDA Guidance and/or Supportive Publications*
AAMI CR500:2019 Basic Introduction to the IEC 60601 Series.

Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contact
 Jin Zhang
  CDRH/OPEQ/OIDRH/DRH/NMRTB/
  301-796-5938
  Jin.Zhang@fda.hhs.gov
Standards Development Organization
IEC International Electrotechnical Commission http://www.iec.ch/
FDA Specialty Task Group (STG)
Radiology
*These are provided as examples and others may be applicable.
-
-